Abstract

This study evaluated the tradeoffs amongst fatigue crack propagation resistance, wear resistance, and oxidative stability in a wide variety of clinically-relevant cross-linked ultra-high molecular weight polyethylene. Highly cross-linked re-melted materials showed good oxidation and wear performance, but diminished fatigue crack propagation resistance. Highly cross-linked annealed materials showed good wear and fatigue performance, but poor oxidation resistance. Moderately cross-linked re-melted materials showed good oxidation resistance, but moderate wear and fatigue resistance. Increasing radiation dose increased wear resistance but decreased fatigue crack propagation resistance. Annealing reduced fatigue resistance less than re-melting, but left materials susceptible to oxidation. This appears to occur because annealing below the melting temperature after cross-linking increased the volume fraction and size of lamellae, but failed to neutralize all free radicals. Alternately, re-melting after cross-linking appeared to eliminate free radicals, but, restricted by the network of cross-links, the re-formed lamellae were fewer and smaller in size which resulted in poor fatigue crack propagation resistance. This is the first study to simultaneously evaluate fatigue crack propagation, wear, oxidation, and microstructure in a wide variety of clinically-relevant ultra-high. The tradeoff we have shown in fatigue, wear, and oxidation performance is critical to the material's long-term success in total joint replacements.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Tribology and Wear Analysis · Polymer crystallization and properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (49)

Show source-provided references
1. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene10.1002/jbm.a.10606 ↗
2. Compression and tension fatigue resistance of medical grade ultra high molecular weight polyethylene: the effect of morphology, sterilization, aging and temperature10.1016/S0032-3861(99)00199-8 ↗
3. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement10.2106/00004623-198668070-00010 ↗
4. High pressure crystallized UHMWPEs
5. Tensile and tribological properties of high-crystallinity radiation crosslinked UHMWPE10.1002/jbm.b.31265 ↗
6. The epidemiology of revision total knee arthroplasty in the United States
7. The epidemiology of revision total hip arthroplasty in the United States10.2106/JBJS.H.00155 ↗
8. Comparison of cross-linked polyethylene materials for orthopaedic applications10.1097/01.blo.0000073343.50837.03 ↗
9. Overview of polyethylene as a bearing material: comparison of sterilization methods
10. In vivo UHMWPE biodegradation of retrieved prosthesis10.1016/S0142-9612(98)00013-1 ↗
11. Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide10.1016/S0142-9612(97)00160-9 ↗
12. How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement?10.5435/00124635-200800001-00017 ↗
13. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups10.2106/00004623-200709000-00019 ↗
14. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings10.1016/j.arth.2006.07.006 ↗
15. The basis for a second-generation highly cross-linked UHMWPE10.1097/01.blo.0000238856.61862.7d ↗
16. Degradation of mechanical behavior in UHMWPE after natural and accelerated aging10.1016/S0142-9612(00)00021-1 ↗
17. Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacement. Ultra-high-molecular-weight polyethylene10.1016/S0883-5403(99)90086-4 ↗
18. The effect of sliding velocity on the friction and wear of UHMWPE for use in total artificial joints10.1016/0043-1648(94)90185-6 ↗
19. Structure and order in lamellar phases determined by small-angle scattering10.1107/S0021889804012956 ↗
20. Clinical fracture of cross-linked UHMWPE acetabular liners10.1016/j.biomaterials.2009.07.013 ↗
21. Furmanski, J., Atwood, S.A., Feest, E., Tang, J., Hoang, M., Pruitt, L., 2007. Effect of vitamin E on fatigue of irradiated UHMWPE. In: 53rd Annual Meeting of the Orthopaedic Research Society. San Diego, CA.
22. Peak stress intensity dictates fatigue crack propagation in UHMWPE10.1016/j.polymer.2007.04.006 ↗
23. Fatigue crack propagation resistance of virgin and highly crosslinked, thermally treated ultra-high molecular weight polyethylene10.1016/j.biomaterials.2005.09.010 ↗
24. Glantz · 2005
25. Wear and periprosthetic osteolysis: the problem10.1097/00003086-200112000-00007 ↗
26. Combined chemical and mechanical effects on free radicals in UHMWPE joints during implantation10.1002/jbm.820250807 ↗
27. Tribological properties and microstructure evolution of ultra-high molecular weight polyethylene10.1115/1.2833952 ↗
28. Vitamin-E-blended UHMWPE biomaterials
29. A primer on UHMWPE
30. Lenth, R.V., 2006–2009. Java applets for power and sample size.
31. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements10.1002/jor.1100170203 ↗
32. Comparative fatigue behavior and toughness of remelted and annealed highly crosslinked polyethylenes10.1002/jbm.b.30807 ↗
33. Evaluation of sequentially crosslinked ultra-high molecular weight polyethylene10.1002/jbm.b.31257 ↗
34. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul award10.1054/arth.2001.20540 ↗
35. Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE)10.1016/S0142-9612(99)00039-3 ↗
36. Molecular-conformations in polyethylene after recrystallization or annealing at high-pressures
37. Patten, E.W., 2008. Multi-axial tribo-system: rolling, sliding and rotation of UHMWPE in total joint replacements. University of California at Berkeley. Berkeley, CA.
38. Factors affecting near-threshold fatigue crack propagation behavior of orthopedic grade ultra high molecular weight polyethylene10.1016/S0032-3861(97)00448-5 ↗
39. Effect of cross-linking on the microstructure and mechanical properties of ultra-high molecular weight polyethylene10.1097/01.blo.0000185310.59202.e5 ↗
40. Elastic compliances and stress-intensity factors for side-grooved compact specimens10.1007/BF00034260 ↗
41. The combined effects of crosslinking and high crystallinity on the microstructural and mechanical properties of ultra high molecular weight polyethylene10.1016/j.biomaterials.2005.09.033 ↗
42. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty10.2106/00004623-200710000-00016 ↗
43. A study of the nanostructure and tensile properties of ultra-high molecular weight polyethylene10.1016/j.biomaterials.2003.10.027 ↗
44. Rolling sliding wear of UHMWPE for knee bearing applications10.1016/j.wear.2006.11.038 ↗
45. Structural and chemical changes in ultra-high-molecular-weight polyethylene due to gamma radiation-induced crosslinking and annealing in air
46. A unified theory of wear for ultra-high molecular weight polyethylene in multi-directional sliding10.1016/S0043-1648(00)00522-6 ↗
47. Wear mechanisms of UHMWPE in total joint replacements10.1016/0043-1648(94)07027-X ↗
48. A highly crosslinked UHMWPE for CR and PS total knee arthroplasties10.1016/j.arth.2007.05.007 ↗
49. Wear mechanisms of untreated and gamma irradiated ultra-high molecular weight polyethylene for total joint replacements10.1115/1.1828079 ↗