journal article·2016-09Open access

Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis

Rafaela Coray, Marco Zeltner, Mutlu Özcan

Abstract

PURPOSE: The use of implants and their respective suprastructures to replace missing teeth has become a common therapeutic option in dentistry. Prior to their clinical application, all implant components have to demonstrate suitable durability in laboratory studies. Fatigue tests utilising cyclic loading typically simulate masticatory function in vitro. The objectives of this systematic review were to assess the loading conditions used for fatigue testing of implant abutments and to compare the fracture strength of different types of implant abutment and abutment-connection types after cyclic loading. MATERIALS AND METHODS: Original scientific papers published in MEDLINE (PubMed) and Embase database in English between 01/01/1970 and 12/31/2014 on cyclic loading on implant abutments were included in this systematic review. The following MeSH terms, search terms and their combinations were used: "in vitro" or "ex vivo" or experimental or laboratory, "dental implants", "implants, experimental", "dental prosthesis, implant-supported", "fatigue", "dental abutments", "cyclic loading", "cyclic fatigue", "mechanical fatigue", "fatigue resistance", "bending moments", and "fracture". Two reviewers performed screening and data abstraction. Only the studies that reported, static fracture values before and after fatigue cycling of implant abutments, were included that allowed comparison of aging effect through cyclic loading. Data (N) were analyzed using a weighted linear regression analysis (α=0.05). RESULTS: The selection process resulted in the final sample of 7 studies. In general, loading conditions of the fatigue tests revealed heterogeneity in the sample but a meta-analysis could be performed for the following parameters: a) abutment material, b) implant-abutment connection, and (c) number of fatigue cycles. Mean fracture strength of titanium (508.9±334.6N) and for zirconia abutments (698.6±452.6N) did not show significant difference after cyclic loading (p>0.05). Internal implant-abutment connections demonstrated significantly higher fracture strength after cyclic loading compared to external ones (internal: 774.0±582.3N; external: 481.2±137.5N; p=0.022). The mean fracture strength of all abutment types decreased significantly when number of loading cycles exceeded 1,000,000 cycles (<1×10(-6): 1047.0±751.3N; >1×10(-6): 556.7±317.6N; p=0.032). CONCLUSION: The results of this meta-analysis, favour the use of internal implant-abutment connections in combination with either titanium or zirconia abutment materials. Number of cycles had a significant impact on the fracture strength after cyclic loading.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Periodontal Regeneration and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (40)

Show source-provided references
1. Post fatigue fracture resistance of modified prefabricated zirconia implant abutments10.1016/j.prosdent.2013.08.023 ↗
2. Effect of cyclic loading on fracture strength and microleakage of a quartz fiber dowel with different adhesive, cement and resin core material combinations
3. Effectiveness of screw surface coating on the stability of zirconia abutments after cyclic loading
4. Fatigue resistance and failure mode of adhesively restored custom metal-composite resin premolar implant abutments10.11607/jomi.2836 ↗
5. Influence of hex geometry and prosthetic table width on static and fatigue strength of dental implants10.1016/S0022-3913(99)70030-2 ↗
6. Influence of three types of abutments on preload values before and after cyclic loading with structural analysis by scanning electron microscopy10.11607/jomi.2481 ↗
7. Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation10.1111/j.1365-2842.2005.01515.x ↗
8. Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading10.1016/j.jpor.2011.02.002 ↗
9. Fracture resistancce of titanium and zirconia abutmets: An in vitro study10.1016/S0022-3913(13)60306-6 ↗
10. Reliability and failure modes of anterior single-unit implant-supported restorations10.1111/j.1600-0501.2011.02269.x ↗
11. Reliability and failure modes of internal conical dental implant connections10.1111/j.1600-0501.2012.02443.x ↗
12. Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis10.1016/j.dental.2012.05.004 ↗
13. Zirconium implant abutments: Fracture strentght and influence of cyclic loading on retaining-screw loosening
14. Influence of surface and heat treatments on the flexural strength of Y-TZP dental ceramic10.1016/j.jdent.2004.07.001 ↗
15. Evaluation of loading coniditons on fatigue-failed implants by fracture surface analysis
16. ISO 14801:2007. Dentistry – Implants – Dynamic fatigue test for endosseous dental implants
17. The slippery slope – critical perspectives on in vitro research methodologies10.1016/j.dental.2011.09.001 ↗
18. Stability of implant-abutment interface with a hexagon-mediate butt joint: failure mode and bending resistance10.1111/j.1708-8208.2005.tb00067.x ↗
19. Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading10.1111/j.1708-8208.2004.tb00223.x ↗
20. Stability of Prototype two-piece Zirconia and titanium implants after artificial Aging: a in vitro Pilot Study10.1111/j.1708-8208.2008.00116.x ↗
21. Use of contact testing in the characterization and design of all-ceramic crownlike layer structures: a review10.1067/mpr.2001.119581 ↗
22. Fatigue resistance and failure mode of CAD/CAM composite resin implant abutments restored with type III composite resin and porcelain veneers10.1111/j.1600-0501.2010.02103.x ↗
23. Fatigue resistance and failure mode of novel-design anterior single-tooth implant restorations: influence of material selection for type III veneers bonded to zirconia abutments10.1111/j.1600-0501.2010.02012.x ↗
24. Damping behavior of implant-supported restorations10.1111/j.1600-0501.2011.02311.x ↗
25. Bending moments of zirconia and titanium implant abutments supporting all-ceramic crowns after aging10.1111/clr.12192 ↗
26. Load fatigue performance of implant-ceramic abutment combinations
27. Fracture behavior of zirconia implant abutments is influenced by superstructure geometry10.1007/s00784-013-1111-3 ↗
28. Fatigue resistance and failure mode adhesively restored custom implant zirconia abutments10.1111/j.1600-0501.2011.02360.x ↗
29. Effect of air-particle abrasion protocols on the biaxial flexural strength, surface characteristics and phase transformation of zirconia after cyclic loading10.1016/j.jmbbm.2013.01.005 ↗
30. Fatigue resistance of ITI implant-abutment connectors. A comparison of the starad cone with a novel internally keyed design10.1034/j.1600-0501.2002.130515.x ↗
31. Comparison of fracture resistance of pressable metal ceramic custom implant abutment with a commercially fabricated CAD/CAM zirconia implant abutment10.1016/j.prosdent.2013.06.004 ↗
32. Load fatigue Performance of a single-tooth implant abutment system: effect of diameter
33. Load fatigue performance of four implant-abutment interface designs: Effect of torque level and implant system
34. In vitro performance of zirconia and titanium implant/abutment systems for anterior application10.1016/j.jdent.2014.03.010 ↗
35. Load fatigue performance of conical implant-abutment connections
36. Implant-abutment interface design affects fatigue and fracture strength of implants10.1111/j.1600-0501.2008.01581.x ↗
37. In vitro fatigue and fracture strenght testig of one-piece Zircoia implant abutments ad ziconia implant abutmets connected to titanium cores10.11607/jomi.2772 ↗
38. Fracture strength and failure mode of five different single-tooth imlant-abutment combinations
39. Bending moments of zirconia ad titanium abutments with internal and external implant-abutmet connections after aging and chewing simulation10.1111/j.1600-0501.2010.02141.x ↗
40. Biomechanical aspects and material properties