Tribolayer formation in a metal-on-metal (MoM) hip joint: An electrochemical investigation
Source abstract
Abstract
The demand for total hip replacement (THR) surgery is increasing in the younger population due to faster rehabilitation and more complete restoration of function. Up to 2009, metal-on-metal (MoM) hip joint bearings were a popular choice due to their design flexibility, post-operative stability and relatively low wear rates. The main wear mechanisms that occur along the bearing surface of MoM joints are tribochemical reactions that deposit a mixture of wear debris, metal ions and organic matrix of decomposed proteins known as a tribolayer. No in-depth electrochemical studies have been reported on the structure and characteristics of this tribolayer or about the parameters involved in its formation. In this study, we conducted an electrochemical investigation of different surfaces (bulk-like: control, nano-crystalline: new implant and tribolayer surface: retrieved implant) made out of two commonly used hip CoCrMo alloys (high-carbon and low-carbon). As per ASTM standard, cyclic polarization tests and electrochemical impedance spectroscopy tests were conducted. The results obtained from electrochemical parameters for different surfaces clearly indicated a reduction in corrosion for the tribolayer surface (Icorr: 0.76μA/cm(2)). Further, polarization resistance (Rp:2.39±0.60MΩ/cm(2)) and capacitance (Cdl:15.20±0.75μF/cm(2)) indicated variation in corrosion kinetics for the tribolayer surface, that attributed to its structure and stability in a simulated body environment.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Fuel Cells and Related Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (34)
Show source-provided references
1. Synergistic effect of I− ions on the corrosion inhibition of Al in 1.0;M phosphoric acid solutions by purine10.1016/j.matchemphys.2008.10.057 ↗
2. Corrosion processes of physical vapor deposition-coated metallic implants10.1615/CritRevBiomedEng.v37.i6.10 ↗
3. Influence of fretting regimes on the tribocorrosion behaviour of Ti6Al4V in 0.9wt% sodium chloride solution10.1016/j.wear.2003.11.003 ↗
4. The epidemiology of revision total hip arthroplasty in the United States
5. Nucleation of corrosion pits in Ringer's solution containing bovine serum10.1016/j.corsci.2007.05.018 ↗
6. Subsurface microstructure of metal-on-metal hip joints and its relationship to wear particle generation10.1002/jbm.b.30132 ↗
7. New insights into wear and biological effects of metal-on-metal bearings
8. Inflammatory pseudotumor associated with femoral nerve palsy following metal-on-metal resurfacing of the hip. A case report
9. The tribological difference between biomedical steels and CoCrMo-alloys10.1016/j.jmbbm.2012.01.007 ↗
10. Influence of bovine serum albumin in sulphuric acid aqueous solution on the corrosion and the passivation of an iron–chromium alloy10.1016/j.electacta.2005.02.116 ↗
11. Metal sensitivity in patients with orthopaedic implants
12. Differences in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements10.1016/S0736-0266(03)00186-4 ↗
13. Determination of effective capacitance and film thickened from constant-phase-element parameters10.1016/j.electacta.2009.10.065 ↗
14. Passive and transpassive behaviour of CoCrMo in simulated biological solutions10.1016/j.electacta.2003.12.043 ↗
15. Electrochemical modeling of passivation phenomena in tribocorrosion10.1016/S0043-1648(99)00314-2 ↗
16. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030
17. Graphitic tribological layers in metal-on-metal hip replacements10.1126/science.1213902 ↗
18. Dominant role of molybdenum in the electrochemical deposition of biological macromolecules on metallic surfaces10.1021/la304046q ↗
19. Significance of tribocorrosion in biomedical applications: overview and current status10.1155/2009/250986 ↗
20. Tribocorrosion behavior of CoCrMo alloy for hip prosthesis as a function of loads: a comparison between two testing systems10.1016/j.wear.2011.01.086 ↗
21. Passivation and corrosion behaviours of cobalt and cobalt–chromium–molybdenum alloy10.1016/j.corsci.2007.03.023 ↗
22. Pseudotumours associated with metal-on-metal hip resurfacings
23. Characterization of wear particles generated from CoCrMo alloy under sliding wear conditions10.1016/j.wear.2010.12.045 ↗
24. Transfer, mixing and associated chemical and mechanical processes during the sliding of ductile materials10.1016/S0043-1648(00)00460-9 ↗
25. Do oral biofilms influence the wear and corrosion behavior of titanium?10.1080/08927011003767985 ↗
26. Study of the synergistic effect observed for the corrosion protection of a carbon steel by an association of phosphates10.1016/S0010-938X(02)00013-6 ↗
27. Effect of physico-chemical properties of simulated body fluids on the electrochemical behaviour of CoCrMo alloy10.1016/j.electacta.2011.06.068 ↗
28. Adsorption of bovine serum albumin on CoCrMo surface: effect of temperauture and protein concentration10.1016/j.colsurfb.2010.05.005 ↗
29. Electrochemical studies on the influence of proteins on the corrosion of implant alloys10.1016/0142-9612(88)90119-6 ↗
30. Tribochemical reaction on metal-on-metal hip joint bearings: a comparison between in-vitro and in-vivo results10.1016/S0043-1648(03)00127-3 ↗
31. Wear mechanisms in metal-on-metal bearings: the importance of tribochemical reaction layers
32. A novel dual-impedance-analysis EQCM system—investigation of bovine serum albumin adsorption on gold and platinum electrode surfaces10.1016/S0021-9797(03)00196-6 ↗
33. Understanding the role of corrosion in the degradation of metal-on-metal implants10.1243/095441105X63246 ↗
34. Competitive binding of chromium, cobalt and nickel to serum proteins10.1016/0142-9612(94)90049-3 ↗