Collagen fibre orientation and dispersion govern ultimate tensile strength, stiffness and the fatigue performance of bovine pericardium

A. Whelan, J. Duffy, R.T. Gaul, D. O’Reilly, D.R. Nolan, P. Gunning, C. Lally, B.P. Murphy

Abstract

The durability of bovine pericardium leaflets employed in bioprosthetic heart valves (BHVs) can significantly limit the longevity of heart valve prostheses. Collagen fibres are the dominant load bearing component of bovine pericardium, however fibre architecture within leaflet geometries is not explicitly controlled in the manufacture of commercial devices. Thus, the purpose of this study was to ascertain the influence of pre-determined collagen fibre orientation and dispersion on the mechanical performance of bovine pericardium. Three tissue groups were tested in uniaxial tension: cross-fibre tissue (XD); highly dispersed fibre-orientations (HD); or preferred-fibre tissue (PD). Both the XD and PD tissue were tested under cyclic loading at 1.5 Hz and a stress range of 2.7 MPa. The results of the static tensile experiments illustrated that collagen fibre orientation and degree of alignment significantly influenced the material's response, whereby, there was a statistically significant decrease in material properties between the XD groups and both the PD and HD groups for ultimate tensile strength and stiffness (p < 0.01). Furthermore, HD tissue had a stiffness of approximately 58% of the PD group, and XD tissue had a stiffness of approximately 18% of the PD group. The dynamic behaviour of the XD and PD groups was extremely distinct; for example a Weibull analysis indicated that the 50% probability of failure in specimens with fibres orientated perpendicular (XD) to the loading direction occurred at 375 cycles. Due to this failure, XD specimens survived on average less than 20% of the cycles completed by those in which fibres were aligned along the loading direction (PD). The results from this study indicate that fibre architecture is a significant factor in determining static strength and fatigue life in bovine pericardium, and thus must be incorporated in the design process to improve future device durability.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Cardiac Valve Diseases and Treatments · Orthopaedic implants and arthroplasty · Infective Endocarditis Diagnosis and Management

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (45)

Show source-provided references
1. Characterization of three-dimensional anisotropic heart valve tissue mechanical properties using inverse finite element analysis10.1016/j.jmbbm.2016.04.031 ↗
2. Mechanical testing of pericardium for manufacturing prosthetic heart valves10.1093/icvts/ivv282 ↗
3. Modeling collagen remodeling10.1016/j.jbiomech.2009.09.022 ↗
4. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp – part II: a structural constitutive model10.1115/1.429624 ↗
5. Evaluation of transcatheter heart valve biomaterials: biomechanical characterization of bovine and porcine pericardium10.1016/j.jmbbm.2017.08.013 ↗
6. Reconstructive surgery of mitral valve incompetence: ten-year appraisal10.1016/S0022-5223(19)37942-5 ↗
7. Effect of applied uniaxial stress on rate and mechanical effects of cross-linking in tissue-derived biomaterials10.1016/0142-9612(95)00305-3 ↗
8. Early bioprosthetic valve failure: a pictorial review of rare causes10.1016/j.jcmg.2014.06.025 ↗
9. The tensile strength of natural and chemically modified bovine pericardium10.1002/jbm.820220202 ↗
10. Effect of dynamic glutaraldehyde fixation on the viscoelastic properties of bovine pericardial tissue10.1016/S0142-9612(97)00215-9 ↗
11. Mechanical effects of increases in the load applied in uniaxial and biaxial tensile testing: part I. Calf pericardium10.1023/A:1014388618649 ↗
12. Collagen fibre characterisation in arterial tissue under load using SALS10.1016/j.jmbbm.2017.07.036 ↗
13. A comparative study of bovine and porcine pericardium to highlight their potential advantages to manufacture percutaneous cardiovascular implants10.1177/0885328212465482 ↗
14. Effects of mechanical fatigue on the bending properties of porcine bioprosthetic heart valve10.1097/00002480-199901000-00014 ↗
15. Pure cusp tear of trifecta bioprosthesis 2 years after aortic valve replacement10.5761/atcs.cr.16-00183 ↗
16. “Microstructural and biomechanical alterations of the human aorta as a function of age and location10.1007/s10237-010-0209-7 ↗
17. Optimal bovine pericardial tissue selection sites. II. Cartographic analysis10.1002/(SICI)1097-4636(199802)39:2<215::AID-JBM7>3.0.CO;2-S ↗
18. Biological heart valves
19. Transplantation material bovine pericardium: biomechanical and immunogenic characteristics after decellularization vs. glutaraldehyde-fixing10.1111/j.1399-3089.2012.00719.x ↗
20. Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress10.1002/cnm.2814 ↗
21. Mädler · 2015
22. Martin, C., Sun, W., Modeling of tissue fatigue damage in bio-prosthetic heart valve, in: Proceedings of the ASME 2011 Summer Bioengineering Conference, Parts A and B, 2011, p. 1217.10.1115/SBC2011-53888 ↗
23. Dynamic mechanical characteristics of intact and structurally modified bovine pericardial tissues10.1016/j.jbiomech.2004.05.019 ↗
24. Effects of cyclic flexural fatigue on porcine bioprosthetic heart valve heterograft biomaterials10.1002/jbm.a.32659 ↗
25. The effects of collagen fiber orientation on the flexural properties of pericardial heterograft biomaterials10.1016/j.biomaterials.2004.03.004 ↗
26. Burden of valvular heart diseases: a population-based study10.1016/S0140-6736(06)69208-8 ↗
27. Biomechanical characterization of decellularized and cross-linked bovine pericardium
28. “Timing of aortic valve surgery10.1136/heart.84.2.211 ↗
29. Bovine pericardium for heart valve bioprostheses: in vitro and in vivo characterization of new chemical treatments10.1007/BF01172023 ↗
30. “Aortic bioprosthetic valve durability: incidence, mechanisms, predictors, and management of surgical and transcatheter valve degeneration10.1016/j.jacc.2017.07.715 ↗
31. Orthotropic mechanical properties of chemically treated bovine pericardium10.1114/1.135 ↗
32. Collagen fiber disruption occurs independent of calcification in clinically explanted bioprosthetic heart valves10.1002/jbm.10293 ↗
33. “Collagen fiber architecture of bovine pericardium10.1097/00002480-199407000-00075 ↗
34. A small angle light scattering device for planar connective tissue microstructural analysis10.1007/BF02684845 ↗
35. On the biomechanics of heart valve function10.1016/j.jbiomech.2009.05.015 ↗
36. A novel fibre-ensemble level constitutive model for exogenous cross-linked collagenous tissues10.1098/rsfs.2015.0090 ↗
37. The micromechanical behavior of lyophilized glutaraldehyde-treated bovine pericardium under uniaxial tension10.1016/j.jmbbm.2010.06.010 ↗
38. Sellaro · 2003 · 1
39. Effects of collagen fiber orientation on the response of biologically derived soft tissue biomaterials to cyclic loading10.1002/jbm.a.30871 ↗
40. Bioprosthetic heart valves: modes of failure10.1111/j.1365-2559.2008.03190.x ↗
41. Response of heterograft heart valve biomaterials to moderate cyclic loading10.1002/jbm.a.30031 ↗
42. The mechanical response of glutaraldehyde-fixed bovine pericardium to uniaxial load10.1007/BF00555905 ↗
43. The evolution of bioprosthetic heart valve design and its impact on durability10.1016/S1054-8807(03)00075-9 ↗
44. Mechanisms of bioprosthetic heart valve failure: fatigue causes collagen denaturation and glycosaminoglycan loss10.1002/(SICI)1097-4636(199907)46:1<44::AID-JBM5>3.0.CO;2-D ↗
45. Mechanical and optical anisotropy of bovine pericardium10.1007/BF02446197 ↗