A large taper mismatch is one of the key factors behind high wear rates and failure at the taper junction of total hip replacements: A finite element wear analysis
Source abstract
Abstract
Total hip replacement (THR) is one of the most successful orthopaedic surgeries; however, failures can occur due to adverse reactions to wear debris. Recently, a large number of failures linked to the release of metal particles from the taper junction between femoral head and femoral stem have been reported. One possible reason for this may be design variations such as taper mismatches associated with the taper and trunnion angles. Could a large taper mismatch lead to inappropriate contact mechanics and increase relative micromotion and thus wear? In this study, 3D finite element (FE) models of a commercial THR from a perfectly matched interface to large taper mismatches and a wear algorithm were used to investigate the extent of wear that could occur at this junction and identify the optimum tolerances in order to reduce the wear. A co-ordinate measuring machine (CMM) was used to analyse the wear depth and volumetric wear rate of the tapers of 54 explanted 36mm diameter Cobalt Chromium femoral heads, which had been in service for 5.1 years in average, to validate the FE analyses. It was found that a large taper mismatch (e.g. 9.12´) results in a high wear rate (2.960mm3 per million load cycles). Such wear rates can have a major negative effect on the clinical outcomes of these implants. It was also found that even a slight reduction in mismatch significantly reduced the magnitude of the wear rates (0.069mm3 per million load cycles on average for 6´ taper mismatch). It is recommended that the cone angles of femoral head and femoral trunnion should be manufactured to produce a taper mismatch of less than 6´ at the taper junction.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (38)
Show source-provided references
1. 2016. National joint registry, 13th annual report for England, Wales, Northern Ireland and the Isle of Man 13th annual report for England, Wales, Northern Ireland and the Isle of Man.
2. Metal-on-metal bearing in hip prosthesis generates 100-fold less wear debris than metal-on-polyethylene10.3109/17453679908997845 ↗
3. Corrosion-induced fracture of a double-modular hip prosthesis10.2106/JBJS.I.00980 ↗
4. Hip contact forces and gait patterns from routine activities10.1016/S0021-9290(01)00040-9 ↗
5. Wear patterns of taper connections in retrieved large diameter metal‐on‐metal bearings10.1002/jor.22326 ↗
6. Surface analysis of the taper junctions of retrieved and in vitro tested modular hip prostheses
7. Bone, M.C., Sidaginamale, R.P., Lord, J.K., Scholes, S.C., Joyce, T.J., Nargol, A.V., Langton, D.J., 2015. Determining material loss from the femoral stem trunnion in hip arthroplasty using a coordinate measuring machine. Proc. Inst. Mech. Eng. Part H: J. Eng. Med. 229, 69–76.10.1177/0954411914567552 ↗
8. Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal‐on‐metal total hip arthroplasty system10.1002/jor.22970 ↗
9. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty10.2106/JBJS.K.01352 ↗
10. A computational approach to fretting wear prediction at the head–stem taper junction of total hip replacements10.1016/j.wear.2015.06.016 ↗
11. The effect of different assembly loads on taper junction fretting wear in total hip replacements10.1016/j.triboint.2015.11.025 ↗
12. Fessler, H., Fricker, D., 1989. Friction in femoral prosthesis and photoelastic model cone taper joints. Proc. Inst. Mech. Eng. Part H: J. Eng. Med. 203, 1–14.10.1243/PIME_PROC_1989_203_003_01 ↗
13. An energy description of wear mechanisms and its applications to oscillating sliding contacts10.1016/S0043-1648(03)00117-0 ↗
14. Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacements10.1016/S0736-0266(03)00186-4 ↗
15. Fretting corrosion mechanisms at modular implant interfaces, Modularity of Orthopedic Implants
16. Joyce, T.J., Langton, D.J., Jameson, S.S., Nargol, A.V.F., 2009. Tribological analysis of failed resurfacing hip prostheses and comparison with clinical data. Proc. Inst. Mech. Eng. Part H: J. Eng. Tribol. 223, 317–323.10.1243/13506501JET484 ↗
17. Taper junction failure in large-diameter metal-on-metal bearings10.1302/2046-3758.14.2000047 ↗
18. Accelerating failure rate of the ASR total hip replacement10.1302/0301-620X.93B8.26040 ↗
19. Reducing metal ion release following hip resurfacing arthroplasty10.1016/j.ocl.2011.01.006 ↗
20. Investigation of taper failure in a contemporary metal on metal hip arthroplasty system through examination of explanted prosthesis
21. Practical considerations for volumetric wear analysis of explanted hip arthroplasties10.1302/2046-3758.33.2000249 ↗
22. Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis10.1302/0301-620X.93B10.27645 ↗
23. Development of a friction energy capacity approach to predict the surface coating endurance under complex oscillating sliding conditions10.1016/j.triboint.2004.06.002 ↗
24. Volumetric wear assessment of failed metal-on-metal hip resurfacing prostheses10.1016/j.wear.2011.07.009 ↗
25. Wear characterization of Ti–6Al–4V under fretting–reciprocating sliding conditions10.1016/j.wear.2007.08.004 ↗
26. Pseudotumor in a well-fixed metal-on-polyethylene uncemented hip arthroplasty10.1016/j.arth.2011.07.015 ↗
27. Modularity of orthopedic implants
28. Material loss at the taper junction of retrieved large head metal‐on‐metal total hip replacements10.1002/jor.22431 ↗
29. Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty10.1302/0301-620X.77B6.7593096 ↗
30. Why does titanium alloy wear cobalt chrome alloy despite lower bulk hardness: a nanoindentation study?10.1016/j.tsf.2013.06.020 ↗
31. Variations in the trunnion surface topography between different commercially available hip replacement stems10.1002/jor.22741 ↗
32. Taper design affects failure of large-head metal-on-metal total hip replacements10.1007/s11999-013-3115-3 ↗
33. Corrosion at the head-neck interface of current designs of modular femoral components10.1302/0301-620X.98B5.35592 ↗
34. Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topography10.1002/jor.22461 ↗
35. Quantitative assessment of walking activity after total hip or knee replacement*10.2106/00004623-199801000-00010 ↗
36. Assessing Morse taper function: the relationship between impaction force, disassembly force, and design variables
37. Effect of manufacturing tolerances on the micromotion at the Morse taper interface in modular hip implants using the finite element technique10.1016/0142-9612(96)88713-8 ↗
38. A finite element methodology for wear-fatigue analysis for modular hip implants10.1016/j.triboint.2013.02.016 ↗