A statistical approach to understand the role of inclusions on the fatigue resistance of superelastic Nitinol wire and tubing
Source abstract
Abstract
Superelastic wires and diamond-shaped stent surrogates were manufactured from Nitinol rods and tubing, respectively, from five different mill product suppliers - Standard VAR, Standard VIM, Standard VIM+VAR, Process-Optimized VIM+VAR, and High-Purity VAR. High-cycle fatigue tests up to 10(7) cycles were conducted under tension-tension conditions for wires and bending conditions for diamonds. These materials were compared under both testing methods at 37°C with 6% prestrain and 3% mean strain (unloading plateau) with a range of alternating strains. The High-Purity VAR material outperformed all alloys tested with a measured 10(7)-fatigue alternating strain limit of 0.32% for wire and 1.75% for diamonds. Process-Optimized VIM+VAR material was only slightly inferior to the High Purity VAR with a diamond alternating bending strain limit of 1.5%. These two "second generation" Nitinol alloys demonstrated approximately a 2× increase in 10(7)-cycle fatigue strain limit compared to all of the Standard-grade Nitinol alloys (VAR, VIM, and VIM+VAR) that demonstrated virtually indistinguishable fatigue performance. This statistically-significant increase in fatigue resistance in the contemporary alloys is ascribed to smaller inclusions in the Process-Optimized VIM+VAR material, and both smaller and fewer inclusions in the High-Purity VAR Nitinol.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Shape Memory Alloy Transformations · Orthopaedic implants and arthroplasty · Metal and Thin Film Mechanics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (23)
Show source-provided references
1. ASTM F136-13, Standard specification for wrought titanium–6aluminum–4vanadium ELI (Extra Low Interstitial) alloy for surgical implant applications (UNS R56401), Book of Standards Volume 13.01, Subcommittee F04.12, 2013.
2. ASTM F2004-05, Standard test method for transformation temperature of nickel–titanium alloys by thermal analysis, Book of Standards Volume 13.01, Subcommittee F04.15, 2005.
3. ASTM F2063-12, Standard specification for wrought nickel–titanium shape memory alloys for medical devices and surgical implants. ASTM Book of Standards Volume 13.01, Subcommittee F04.12, 2012.
4. ASTM F2082-06, Standard test method for determination of transformation temperature of nickel–titanium shape memory alloys by bend and free recovery, Book of Standards Volume 13.01, Subcommittee F04.15, 2006.
5. ASTM F2516-07e2, Standard test method for tension testing of nickel–titanium superelastic materials, ASTM Book of Standards Volume 13.2, Subcommmittee F04.15, 2007.
6. Effect of microstructure on the fatigue of hot-rolled and cold-drawn NiTi shape memory alloys10.1016/j.msea.2007.11.033 ↗
7. Hosmer · 1989
8. Bard recovery filter: evaluation and management of vena cava limb perforation, fracture, and migration10.1016/j.jvir.2008.09.032 ↗
9. Influence of microstructural purity on the bending fatigue behavior of VAR-melted superelastic Nitinol10.1016/j.jmbbm.2014.02.008 ↗
10. Nitinol fatigue investigation on Stent-finish specimens using tension–tension method10.1007/s11665-010-9792-0 ↗
11. Nitinol fatigue life for variable strain amplitude fatigue10.1007/s11665-012-0387-9 ↗
12. Macherauch · 2002
13. Nitinol fatigue: a review of microstructures and mechanisms
14. Fatigue and durability of Nitinol stents10.1016/j.jmbbm.2007.08.001 ↗
15. Rotary-bending fatigue characteristics of medical-grade Nitinol wire10.1016/j.jmbbm.2013.06.003 ↗
16. Development of the Z specimen for tensile–tensile, tensile–compression, compression–compression wire testing10.1007/s11665-010-9766-2 ↗
17. Impurity levels and fatigue lives of pseudoelastic NiTi shape memory alloys10.1016/j.actamat.2013.02.054 ↗
18. A fracture-mechanics-based approach to fracture control in biomedical devices manufactured from superelastic Nitinol tube10.1002/jbm.b.30840 ↗
19. Mechanical fatigue and fracture of Nitinol10.1179/1743280411Y.0000000009 ↗
20. Fatigue performance of Nitinol round wire with varying cold work reductions10.1007/s11665-009-9363-4 ↗
21. Prevalence and clinical impact of stent fractures after femoropopliteal stenting10.1016/j.jacc.2004.11.026 ↗
22. Tolomeo, D., Davidson, S., Santinoranout, M., 2000. Cyclic properties of superelastic Nitinol: design implications. In: Proceedings of the International Conference on Shape Memory and Superelastic Technologies. SMST. Pacific Grove, CA, USA, pp. 409–417.
23. Wick, A., et al., 2004. Bending fatigue characteristics of Nitinol. In: Proceedings of the ASM Materials & Processes for Medical Devices Conference. Minneapolis, MN, USA. ASM International. pp. 15–20.