Wear resistance and flexural properties of low force SLA- and DLP-printed splint materials in different printing orientations: An in vitro study

Philipp Simeon, Alexey Unkovskiy, Bardia Saadat Sarmadi, Robert Nicic, Petra Julia Koch, Florian Beuer, Franziska Schmidt

Abstract

OBJECTIVES: To investigate the influence of material and printing orientation on wear resistance and flexural properties of one low force SLA- and two DLP-printed splint materials and to compare these 3D-printed splints to a subtractively manufactured splint material. METHODS: Two DLP-printed (V-Print splint, LuxaPrint Ortho Plus) and one low force SLA-printed (Dental LT Clear) material, where specimens were printed in three printing orientations (0°, 45°, 90°), were investigated. In addition, one milled splint material (Zirlux Splint Transparent) was examined. A total of 160 specimens were produced for both test series. The two-body wear test was performed in a chewing simulator (80'000 cycles at 50 N with 5-55 °C thermocycling). Steatite balls were used as antagonists. The wear pattern was analyzed with a 3D digital microscope in terms of maximum vertical intrusion depth (mm) and total volume loss (mm³). The flexural properties were investigated by three-point bending in accordance with ISO 20795-1: 2013 (denture base polymers). The flexural strength (MPa) and the flexural modulus (MPa) were measured. Two-way ANOVA was performed to investigate the effects of the two independent variables material and printing orientation for the three 3D-printed materials. The comparison of the printing orientations within one material was carried out with one-way ANOVA with post-hoc Tukey tests. RESULTS: Two-way ANOVA revealed that wear and flexural properties are highly dependent on the 3D-printed material (p < 0.001). Across groups, a significant effect was observed for wear depth (p = 0.031) and wear volume (p = 0.044) with regard to printing orientation but this was not found for flexural strength (p = 0.080) and flexural modulus (p = 0.136). One-way ANOVA showed that both DLP-printed groups showed no significant differences within the printing orientations in terms of wear and flexural properties. Dental LT Clear showed that 90° oriented specimens had higher flexural strength than 0° oriented ones (p < 0.001) and 45° oriented specimens also showed higher values than 0° ones (p = 0.038). No significant differences were observed within the printing orientations for flexural modulus and wear behaviour within this group. T-tests showed that the milled splints exhibited statistically higher wear resistance and flexural properties compared to all three 3D-printed splint materials (p < 0.001) and that highly significant differences were found between the 3D-printed splint materials for both test series. CONCLUSION: Within the limitations of this in vitro study, it can be stated that wear behaviour and flexural properties are highly dependent on the 3D-printed material itself. Currently, milled splints exhibit higher wear resistance and flexural properties compared to 3D-printed splint materials. The printing orientation has a minor influence on the properties investigated. Nevertheless, two-way ANOVA also showed a significant influence of printing orientation in the wear test across groups and one-way ANOVA detected significant effects for SLA material in terms of flexural strength, with printing in 90° showing the highest flexural strength. Therefore, anisotropy was found in SLA material, but it can be limited with the employed printing parameters. Both DLP-printed materials showed no significant difference within the printing orientation.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Dental materials and restorations · Orthopaedic implants and arthroplasty · Dental Trauma and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (60)

Show source-provided references
1. Effect of 3D printing direction and water storage on nano-mechanical properties of 3D printed and auto-polymerized polymer with special emphasis on printing layer interface10.1166/mex.2019.1502 ↗
2. 3D printed remendable polylactic acid blends with uniform mechanical strength enabled by a dynamic Diels-Alder reaction10.1039/C7PY00310B ↗
3. Surface quality of 3D-printed models as a function of various printing parameters10.3390/ma12121970 ↗
4. Surface roughness and wear behavior of occlusal splint materials made of contemporary and high-performance polymers10.1007/s10266-019-00463-1 ↗
5. Comparing the mechanical properties of pressed, milled, and 3D-printed resins for occlusal devices10.1016/j.prosdent.2019.10.024 ↗
6. Effect of build orientation on the trueness of occlusal splints fabricated by three-dimensional printing10.2334/josnusd.23-0115 ↗
7. In vitro wear of various orthotic device materials10.1016/S0022-3913(03)00545-6 ↗
8. Oral splints: the crutches for temporomandibular disorders and bruxism?10.1177/10454411980090030701 ↗
9. Mechanical performance of 3-dimensionally printed resins compared with conventional and milled resins for the manufacture of occlusal devices: a systematic review
10. Effect of layer orientation on mechanical properties of rapid prototyped samples10.1080/10426910008912976 ↗
11. Two-body wear and surface hardness of occlusal splint materials10.4012/dmj.2022-100 ↗
12. Evaluation of wear behaviour of various occlusal splint materials and manufacturing processes10.1016/j.jmbbm.2021.105053 ↗
13. Polishability and wear resistance of splint material for oral appliances produced with conventional, subtractive, and additive manufacturing10.1016/j.jmbbm.2017.07.019 ↗
14. Volumetric wear of various orthotic appliance materials10.1179/crn.2013.31.4.004 ↗
15. The influence of antagonist standardization on wear10.1177/00220345990780021201 ↗
16. High-resolution tomographic volumetric additive manufacturing10.1038/s41467-020-14630-4 ↗
17. Fracture resistance and 2-body wear of 3-dimensional-printed occlusal devices10.1016/j.prosdent.2018.04.007 ↗
18. Formlabs Dental LT Clear Safety Data Sheet, https://formlabs-media.formlabs.com/datasheets/Safety_Data_Sheet_EN-EU_-_Dental_LT_Clear.pdf (accessed 27 January 2024).
19. Accuracy of CAD/CAM-fabricated bite splints: milling vs 3D printing10.1007/s00784-020-03329-x ↗
20. Mayra Torres Vasques · Accuracy and Internal Fit of 3D printed Occlusal Splint, according to the printing position. · 2022
21. Anisotropy of photopolymer parts made by digital light processing10.3390/ma10010064 ↗
22. Materials and manufacturing techniques for occlusal splints: a literature review10.1111/joor.13550 ↗
23. Effects of polishing and artificial aging on mechanical properties of dental LT Clear(®) resin10.3390/jfb14060295 ↗
24. In vitro time efficiency, fit, and wear of conventionally- versus digitally-fabricated occlusal splints10.3390/ma15031085 ↗
25. Evaluation of flexible three-dimensionally printed occlusal splint materials: an in vitro study10.1016/j.dental.2023.08.178 ↗
26. Evaluation of the mechanical properties and degree of conversion of 3D printed splint material10.1016/j.jmbbm.2020.104254 ↗
27. Wear performance of dental ceramics after grinding and polishing treatments10.1016/j.jmbbm.2012.03.002 ↗
28. A study of the flexural strength and surface hardness of different materials and technologies for occlusal device fabrication10.1016/j.prosdent.2018.09.022 ↗
29. Mechanical properties of 3D-printed occlusal splint materials10.3390/dj11080199 ↗
30. Effects of environmental conditions, aging, and build orientations on the mechanical properties of ASTM type I specimens manufactured via stereolithography10.1108/13552541211250373 ↗
31. Influence of print orientation and wet-dry storage time on the intaglio accuracy of additively manufactured occlusal devices
32. Influence of printing angulation on the surface roughness of additive manufactured clear silicone indices: an in vitro study10.1016/j.prosdent.2020.02.008 ↗
33. Influences of build angle on the accuracy, printing time, and material consumption of additively manufactured surgical templates10.1016/j.prosdent.2020.09.012 ↗
34. A digital process for additive manufacturing of occlusal splints: a clinical pilot study10.1098/rsif.2013.0203 ↗
35. Two-body wear of occlusal splint materials from subtractive computer-aided manufacturing and three-dimensional printing10.1007/s00784-022-04543-5 ↗
36. 3D printing in digital prosthetic dentistry: an overview of recent developments in additive manufacturing10.3390/jcm10092010 ↗
37. On reducing anisotropy in 3D printed polymers via ionizing radiation10.1016/j.polymer.2014.07.054 ↗
38. Potential countersample materials for in vitro simulation wear testing10.1016/S0109-5641(01)00043-4 ↗
39. LuxaPrint Ortho Plus Safety Data Sheet, https://www.dmg-america.com/fileadmin/DMG_America/SDS/MSDS_LuxaPrintOrthoPlus.pdf (accessed 27 January 2024).
40. Materials in digital dentistry-A review10.1111/jerd.12566 ↗
41. Additive manufactured polymers in dentistry, current state-of-the-art and future perspectives-A review10.3390/polym14173658 ↗
42. Efficacy of stabilization splints for the management of patients with masticatory muscle pain: a qualitative systematic review10.1007/s00784-004-0265-4 ↗
43. Objects build orientation, positioning, and curing influence dimensional accuracy and flexural properties of stereolithographically printed resin10.1016/j.dental.2018.09.011 ↗
44. International Organization of Standardization, 2013. ISO 20795-1:2013: Dentistry - Base Polymers. Part 1: Denture Base Polymers.
45. User Manual Zirlux Splint Transparent, https://www.zirlux.de/assets/sites/2/2017/12/Zirlux-splint-transparent-IFU.pdf (accessed 27 January 2024).
46. The anisotropicity of the flexural properties of an occlusal device material processed by stereolithography10.1016/j.prosdent.2016.03.018 ↗
47. Analysis of compressive strength of occlusal splints manufactured with three liquid resins at three angles of orientation on 3D printer10.33448/rsd-v11i3.26820 ↗
48. VOCO V-Print splint Safety Data Sheet, https://www.voco.dental/en/portaldata/1/resources/products/safety-data-sheets/gb/v-print-splint_sds_gb.pdf (accessed 27 January 2024)..
49. Effect of 3D printing system and post-curing atmosphere on micro- and nano-wear of additive-manufactured occlusal splint materials10.1016/j.jmbbm.2023.105799 ↗
50. Effect of nitrogen gas post-curing and printer type on the mechanical properties of 3D-printed hard occlusal splint material10.3390/polym14193971 ↗
51. Novel three-dimensional-printing strategy based on dynamic urea bonds for isotropy and mechanical robustness of large-scale printed products10.1021/acsami.1c20659 ↗
52. A two-body frictional wear test10.1177/00220345940730091001 ↗
53. Polymers for conventional, subtractive, and additive manufacturing of occlusal devices differ in hardness and flexural properties but not in wear resistance10.1016/j.dental.2020.11.020 ↗
54. Bioinspired ultra-low adhesive energy interface for continuous 3D printing: reducing curing induced adhesion10.1155/2018/4795604 ↗
55. Biaxial flexural strength of 3D-printed splint materials
56. Effect of post-rinsing time on the mechanical strength and cytotoxicity of a 3D printed orthodontic splint material10.1016/j.dental.2021.01.016 ↗
57. Diels-alder reversible thermoset 3D printing: isotropic thermoset polymers via fused filament fabrication10.1002/adfm.201700318 ↗
58. Two-body wear of occlusal splint materials against different antagonists10.1186/s12903-020-01165-9 ↗
59. Effect of occlusal splints for the management of patients with myofascial pain: a randomized, controlled, double-blind study10.3760/cma.j.issn.0366-6999.20130613 ↗
60. Material anisotropy in additively manufactured polymers and polymer composites: a review10.3390/polym13193368 ↗