Implant failure due to biofilm formation is a substantial problem in the field of dental prosthetics. A solution has been proposed in the form of implants with a built-in drug reservoir, but combining sufficient strength and longevity with controlled release capability has proven difficult. This work investigates the feasibility of using laser powder bed fusion to create Ti6Al4V structures with open pore channels while maintaining their mechanical stability. These interconnected pore channels are generated by increasing the distance between consecutive melt pools, denominated as oversized hatch spacing. The impact of varying hatch spacing, laser power and scan speed on the degree of porosity was examined, with both an increase in hatch spacing and a decrease in energy density leading to higher porosity. The pore channels were found to be fully interconnected at total porosity values of 14% or more. The compressive modulus, yield strength and ultimate compressive strength are shown to be strongly related to the density of the structure. Based on the minimal strength and full interconnectivity requirements, the optimal additive manufacturing building conditions were determined. The fatigue properties of the resulting samples were investigated under uniaxial and under inclined compression-compression testing according to ISO 14801, which indicated an endurance limit of 217 MPa in the heat treated state. The results indicate that the use of an oversized hatch spacing is suitable for engineering open porous networks.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Additive Manufacturing Materials and Processes · Bone Tissue Engineering Materials · Additive Manufacturing and 3D Printing Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (46)

Show source-provided references
1. Fatigue behavior of porous biomaterials manufactured using selective laser melting10.1016/j.msec.2013.08.006 ↗
2. Ashby · 2000
3. Predicting the output dimensions, porosity and elastic modulus of additive manufactured biomaterial structures targeting orthopedic implants10.1016/j.jmbbm.2019.07.023 ↗
4. The effects of design parameters on mechanical failure of Ti-6Al-4V implants using finite element analysis10.1016/j.engfailanal.2020.104445 ↗
5. Low- and high-cycle fatigue resistance of Ti-6Al-4V ELI additively manufactured via selective laser melting: Mean stress and defect sensitivity10.1016/j.ijfatigue.2017.10.021 ↗
6. On the limitations of volumetric energy density as a design parameter for selective laser melting10.1016/j.matdes.2016.10.037 ↗
7. Microstructure and mechanical properties of Ti-6Al-4V manufactured by SLM
8. Novel anti-infective implant substrates: Controlled release of antibiofilm compounds from mesoporous silica-containing macroporous titanium10.1016/j.colsurfb.2014.12.054 ↗
9. Antifouling coatings for dental implants: Polyethylene glycol-like coatings on titanium by plasma polymerization10.1116/1.4913376 ↗
10. Adjunctive systemic and local antimicrobial therapy in the surgical treatment of peri-implantitis: A randomized controlled clinical trial10.1177/0022034515601961 ↗
11. Coen · 2020
12. Adding functionality with additive manufacturing: Fabrication of titanium-based antibiotic eluting implants10.1016/j.msec.2016.04.006 ↗
13. Effect of heat treatments on fatigue properties of Ti–6Al–4V and 316L produced by laser powder bed fusion in as-built surface condition
14. Controlled release of chlorhexidine from a mesoporous silica-containing macroporous titanium dental implant prevents microbial biofilm formation10.22203/eCM.v033a02 ↗
15. Additive manufacturing of metallic components – Process, structure and properties10.1016/j.pmatsci.2017.10.001 ↗
16. Peri-implant health and disease. A systematic review of current epidemiology10.1111/jcpe.12334 ↗
17. Clinical and microbiological aspects of biofilm-associated surgical site infections
18. A controlled antibiotic release system to prevent orthopedic-implant associated infections: An in vitro study10.1016/j.ejpb.2015.08.007 ↗
19. 2016
20. Three-dimensional nonlinear finite element analysis and microcomputed tomography evaluation of microgap formation in a dental implant under oblique loading10.11607/jomi.4179 ↗
21. State of the art of bioimplants manufacturing: part I10.1007/s40436-017-0207-4 ↗
22. Peri-implant diseases10.1111/eos.12529 ↗
23. On the mechanical behaviour of titanium alloy TiAl6V4 manufactured by selective laser melting: Fatigue resistance and crack growth performance10.1016/j.ijfatigue.2012.11.011 ↗
24. Editorial: peri-implantitis: the disease of the future
25. Critical assessment of the fatigue performance of additively manufactured Ti–6Al–4V and perspective for future research10.1016/j.ijfatigue.2015.12.003 ↗
26. Surface treatments and functional coatings for biocompatibility improvement and bacterial adhesion reduction in dental implantology10.3390/coatings6010007 ↗
27. 2008
28. Frequency and stress ratio effects in high cycle fatigue of Ti-6Al-4V10.1016/S0142-1123(99)00030-4 ↗
29. Effect of surface roughness on fatigue strength of Ti-6Al-4V alloy manufactured by additive manufacturing10.1016/j.prostr.2019.12.032 ↗
30. Biomaterial and antibiotic strategies for peri-implantitis: A review10.1002/jbm.b.31152 ↗
31. Fatigue behavior of highly porous titanium produced by powder metallurgy with temporary space holders10.1016/j.msec.2015.11.050 ↗
32. Surface roughness and morphology customization of additive manufactured open porous Ti6Al4V structures10.3390/ma6104737 ↗
33. Fabrication methods of porous metals for use in orthopaedic applications10.1016/j.biomaterials.2005.12.002 ↗
34. Drug-releasing implants: Current progress, challenges and perspectives10.1039/C4TB00548A ↗
35. Schijve · 2008
36. The development of a scanning strategy for the manufacture of porous biomaterials by selective laser melting10.1007/s10856-009-3763-8 ↗
37. Compression fatigue of a cellular Al alloy10.1016/S0921-5093(99)00147-1 ↗
38. Metallic powder-bed based 3D printing of cellular scaffolds for orthopaedic implants: A state-of-the-art review on manufacturing, topological design, mechanical properties and biocompatibility10.1016/j.msec.2017.02.094 ↗
39. A study of the microstructural evolution during selective laser melting of Ti-6Al-4V10.1016/j.actamat.2010.02.004 ↗
40. Improving the fatigue performance of porous metallic biomaterials produced by Selective Laser Melting10.1016/j.actbio.2016.10.005 ↗
41. Implant functionalization with mesoporous silica: A promising antibac terial strategy, but does such an implant osseointegrate?
42. Vanmunster · 2020
43. Heat treatment of Ti6Al4V produced by Selective Laser Melting: Microstructure and mechanical properties10.1016/j.jallcom.2012.07.022 ↗
44. Rapid prototyping: Porous titanium alloy scaffolds produced by selective laser melting for bone tissue engineering10.1089/ten.tec.2008.0288 ↗
45. Specific yielding of selective laser-melted Ti6Al4V open-porous scaffolds as a function of unit cell design and dimensions10.3390/met6070166 ↗
46. Effects of scan line spacing on pore characteristics and mechanical properties of porous Ti6Al4V implants fabricated by selective laser melting