journal article·2012-09

Virtual evaluation of stent graft deployment: A validated modeling and simulation study

S. De Bock, F. Iannaccone, G. De Santis, M. De Beule, D. Van Loo, D. Devos, F. Vermassen, P. Segers, B. Verhegghe

Publisher page

Abstract

The presented study details the virtual deployment of a bifurcated stent graft (Medtronic Talent) in an Abdominal Aortic Aneurysm model, using the finite element method. The entire deployment procedure is modeled, with the stent graft being crimped and bent according to the vessel geometry, and subsequently released. The finite element results are validated in vitro with placement of the device in a silicone mock aneurysm, using high resolution CT scans to evaluate the result. The presented work confirms the capability of finite element computer simulations to predict the deformed configuration after endovascular aneurysm repair (EVAR). These simulations can be used to quantify mechanical parameters, such as neck dilations, radial forces and stresses in the device, that are difficult or impossible to obtain from medical imaging.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Aortic aneurysm repair treatments · Aortic Disease and Treatment Approaches · Cardiac, Anesthesia and Surgical Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. Dynamic geometry and wall thickness of the aortic neck of abdominal aortic aneurysms with intravascular ultrasonography10.1016/j.jvs.2007.06.030 ↗
2. Shape-memory alloys: macromodeling and numerical simulations of the superelastic behavior10.1016/S0045-7825(96)01232-7 ↗
3. Aneurysm neck diameter after endovascular repair of abdominal aortic aneurysms10.1016/S1051-0443(07)61771-0 ↗
4. Intraluminal thrombus and risk of rupture in patient specific abdominal aortic aneurysm—FSI modelling10.1080/10255840802176396 ↗
5. Influence of the textile structure on the degradation of explanted aortic endoprostheses10.1016/S1078-5884(03)00341-1 ↗
6. A review of the in vivo and in vitro biomechanical behavior and performance of postoperative abdominal aortic aneurysms and implanted stent-grafts10.1583/08-2370.1 ↗
7. Our capricious vessels: the influence of stent design and vessel geometry on the mechanics of intracranial aneurysm stent deployment10.1016/j.jbiomech.2012.03.012 ↗
8. Long-term outcome of open or endovascular repair of abdominal aortic aneurysm10.1056/NEJMoa0909499 ↗
9. Full-hexahedral structured meshing for image-based computational vascular modeling10.1016/j.medengphy.2011.06.007 ↗
10. Computational comparison of the bending behavior of aortic stent-grafts10.1016/j.jmbbm.2011.09.006 ↗
11. Identification of rupture locations in patient-specific abdominal aortic aneurysms using experimental and computational techniques10.1016/j.jbiomech.2009.09.057 ↗
12. Experimental modelling of aortic aneurysms: novel applications of silicone rubbers10.1016/j.medengphy.2009.06.002 ↗
13. Effect of curvature on displacement forces acting on aortic endografts: a 3-dimensional computational analysis10.1583/08-2667.1 ↗
14. Quantification of aortic distensibility in abdominal aortic aneurysm using ECG-gated multi-detector computed tomography10.1007/s00330-007-0833-5 ↗
15. Comparison of endovascular aneurysm repair with open repair in patients with abdominal aortic aneurysm (EVAR trial 1), 30-day operative mortality results: randomised controlled trial10.1016/S0140-6736(04)16979-1 ↗
16. The cardiac cycle is a major contributor to variability in size measurements of abdominal aortic aneurysms by ultrasound10.1016/j.ejvs.2011.09.025 ↗
17. Comparison of element technologies for modeling stent expansion10.1115/1.2264382 ↗
18. Computational fluid dynamics within bifurcated abdominal aortic stent-grafts10.1177/152660280701400204 ↗
19. The endovascular repair of abdominal aortic aneurysm: an update analysis of structural failure modes of endovascular stent grafts10.1016/S0895-7967(03)00006-1 ↗
20. Dynamics of the aorta before and after endovascular aneurysm repair: a systematic review10.1016/j.ejvs.2009.06.018 ↗
21. Computational mechanics of Nitinol stent grafts10.1016/j.jbiomech.2008.05.032 ↗
22. Update on open repair of abdominal aortic aneurysms: the challenges for endovascular repair10.1016/j.jamcollsurg.2004.07.028 ↗
23. Analysis of biomechanical factors affecting stent-graft migration in an abdominal aortic aneurysm model10.1016/j.jbiomech.2005.07.010 ↗
24. Computational analysis of biomechanical contributors to possible endovascular graft failure10.1007/s10237-005-0003-0 ↗
25. Analytical modeling and numerical simulation of forces in an endoluminal graft10.1177/152660280100800405 ↗
26. Martino, E.S.D., Bohra, A., Scotti, C., Finol, E., Vorp, D.A., 2004. Wall stresses before and after endovascular repair of abdominal aortic aneurysms. In: ASME Conference Proceedings 2004, pp. 325–326.10.1115/IMECE2004-61556 ↗
27. Fluid-structure interaction of a patient-specific abdominal aortic aneurysm treated with an endovascular stent-graft10.1186/1475-925X-8-24 ↗
28. Automated generation of a finite element stent model10.1007/s11517-008-0410-3 ↗
29. Changes in the proximal neck of abdominal aortic aneurysms early after endovascular treatment10.1007/s10016-001-0405-0 ↗
30. Powell · 1994
31. Oversizing of aortic stent grafts for abdominal aneurysm repair: a systematic review of the benefits and risks10.1016/j.ejvs.2009.03.025 ↗
32. Wall stress distribution on three-dimensionally reconstructed models of human abdominal aortic aneurysm10.1067/mva.2000.103971 ↗
33. Abdominal aortic aneurysm risk of rupture: patient-specific FSI simulations using anisotropic model10.1115/1.3005200 ↗
34. Aortic neck dilation after endovascular abdominal aortic aneurysm repair: should oversizing be blamed?10.1007/s10016-006-9067-2 ↗
35. Endovascular vs. open repair of abdominal aortic aneurysms in the Medicare population10.1056/NEJMoa0707348 ↗
36. Aneurysm rupture after EVAR: can the ultimate failure be predicted?10.1016/j.ejvs.2008.10.011 ↗
37. Biomechanical properties of abdominal aortic aneurysms assessed by simultaneously measured pressure and volume changes in humans10.1016/j.jvs.2008.06.060 ↗
38. Use of regional mechanical properties of abdominal aortic aneurysms to advance finite element modeling of rupture risk10.1583/11-3456.1 ↗
39. The effects of anisotropy on the stress analyses of patient-specific abdominal aortic aneurysms10.1007/s10439-008-9490-3 ↗
40. Explant analysis of AneuRx stent grafts: relationship between structural findings and clinical outcome10.1016/j.jvs.2004.03.008 ↗
41. Endovascular device design in the future: transformation from trial and error to computational design