Adhesive strength of bone-implant interfaces and in-vivo degradation of PHB composites for load-bearing applications
Source abstract
Abstract
Aim of this study was to evaluate the response of bone to novel biodegradable polymeric composite implants in the femora of growing rats. Longitudinal observation of bone reaction at the implant site (BV/TV) as well as resorption of the implanted pins were monitored using in vivo micro-focus computed tomography (µCT). After 12, 24 and 36 weeks femora containing the implants were explanted, scanned with high resolution ex vivo µCT, and the surface roughness of the implants was measured to conclude on the ingrowth capability for bone tissue. Scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDX) were used to observe changes on the surface of Polyhydroxybutyrate (PHB) during degradation and cell ingrowth. Four different composites with zirconium dioxide (ZrO2) and Herafill(®) were compared. After 36 weeks in vivo, none of the implants did show significant degradation. The PHB composite with ZrO2 and a high percentage (30%) of Herafill® as well as the Mg-alloy WZ21 showed the highest values of bone accumulation (increased BV/TV) around the implant. The lowest value was measured in PHB with 3% ZrO2 containing no Herafill®. Roughness measurements as well as EDX and SEM imaging could not reveal any changes on the PHB composites׳ surfaces. Biomechanical parameters, such as the adhesion strength between bone and implant were determined by measuring the shear strength as well as push-out energy of the bone-implant interface. The results showed that improvement of these mechanical properties of the studied PHBs P3Z, P3Z10H and P3Z30H is necessary in order to obtain appropriate load-bearing material. The moduli of elasticity, tensile strength and strain properties of the PHB composites are close to that of bone and thus promising. Compared to clinically used PLGA, PGA and PLA materials, their additional benefit is an unchanged local pH value during degradation, which makes them well tolerated by cells and immune system. They might be used successfully for personalized 3D printed implants or as coatings of rapidly dissolving implants.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty · biodegradable polymer synthesis and properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (59)
Show source-provided references
1. Densification route and mechanical properties of Si3N4-bioglass biocomposites10.1016/S0142-9612(01)00194-6 ↗
2. Qualitative and quantitative study of human osteoblast adhesion on materials with various surface roughnesses10.1002/(SICI)1097-4636(200002)49:2<155::AID-JBM2>3.0.CO;2-J ↗
3. Continuous production of poly([R]-3-hydroxybutyrate) by Cupriavidus necator in a multistage bioreactor cascade10.1007/s00253-011-3260-0 ↗
4. Surface treatments and roughness properties of Ti-based biomaterials10.1023/B:JMSM.0000042679.28493.7f ↗
5. Berzins, A., Sumner, D.R., 2000. Implant pushout and pullout test. In: Mech. Test. Bone Bone-Implant Interface. pp. 463–476.10.1201/9781420073560.ch29 ↗
6. Novel materials for bone and cartilage regeneration10.1016/j.cbpa.2006.09.009 ↗
7. Role of material surfaces in regulating bone and cartilage cell response10.1016/0142-9612(96)85758-9 ↗
8. Biomechanical characterization of osseointegration during healing: an experimental in vivo study in the rat10.1016/S0142-9612(97)00018-5 ↗
9. Bone nanostructure near titanium and porous tantalum implants studied by scanning small angle X-ray scattering10.22203/eCM.v012a10 ↗
10. Bone-implant interface strength and osseointegration: biodegradable magnesium alloy versus standard titanium control10.1016/j.actbio.2010.08.020 ↗
11. PHB, crystalline and amorphous magnesium alloys: promising candidates for bioresorbable osteosynthesis implants?10.1016/j.msec.2012.04.032 ↗
12. Progress and challenges in biomaterials used for bone tissue engineering: bioactive glasses and elastomeric composites10.1186/2194-0517-1-2 ↗
13. A role for surface topography in creating and maintaining bone at titanium endosseous implants10.1067/mpr.2000.111966 ↗
14. In situ nanomechanical characterization of the early stages of swelling and degradation of a biodegradable polymer10.1039/C5NR00265F ↗
15. Degradable polymeric materials for osteosynthesis: tutorial10.22203/eCM.v016a09 ↗
16. In vivo degradation performance of micro-arc-oxidized magnesium implants: a micro-CT study in rats10.1016/j.actbio.2012.09.017 ↗
17. Treatment of a chronic recurrent fistulized tibial osteomyelitis: administration of a novel antibiotic-loaded bone substitute combined with a pedicular muscle flap sealing10.1007/s00590-012-0956-5 ↗
18. Mechanical and histomorphometric evaluations of titanium implants with different surface treatments inserted in sheep cortical bone10.1016/S0142-9612(02)00548-3 ↗
19. Nanofunctionalized zirconia and barium sulfate particles as bone cement additives
20. Mechanisms of polymer degradation and erosion10.1016/0142-9612(96)85755-3 ↗
21. Grünewald, T.A., Rennhofer, H., Meischel, M., Nue, V., Hesse, B., Burghammer, M., Gianoncelli, A., Birkedal, H., Cotte, M., Weinberg, A.M., Stanzl-Tschegg S.E., Lichtenegger, H.C., 2015. Response of rat bone structure and mineralization to a degrading Mg implant. In: 79. Jahrestagung der Deutschen Physikalischen Gesellschaft (DPG), Berlin
22. In vitro corrosion and biocompatibility of binary magnesium alloys10.1016/j.biomaterials.2008.10.021 ↗
23. On the in vitro and in vivo degradation performance and biological response of new biodegradable Mg-Y-Zn alloys10.1016/j.actbio.2009.10.008 ↗
24. Effect of metal alloy surface stresses on the viability of ROS-17/2.8 osteoblastic cells10.1016/S0142-9612(02)00107-2 ↗
25. Effect of surface roughness on biodegradability of Poly (3-hydroxybutyrate)
26. Magnesium alloys for temporary implants in osteosynthesis: in vivo studies of their degradation and interaction with bone10.1016/j.actbio.2011.11.008 ↗
27. Degradation behavior and mechanical properties of magnesium implants in rabbit tibiae10.1007/s10853-009-3936-3 ↗
28. Effect of mechanical stimuli on skeletal regeneration around implants10.1016/j.bone.2006.10.027 ↗
29. Hydrolytic degradation characteristics of aliphatic polyesters derived from lactic and glycolic acids10.1002/(SICI)1097-4636(1999)48:3<342::AID-JBM20>3.0.CO;2-7 ↗
30. Lindtner, R.A., 2009. Biomechanical and Radiological Evaluation of the Bone-Implant Interface of New Biodegradable Implants in Comparison to Findings in a Transcortical Rat Model.
31. Limitations of using micro-computed tomography to predict bone-implant contact and mechanical fixation10.1111/j.1365-2818.2011.03541.x ↗
32. Surface Modification of Biomedical Magnesium Alloys by PHB10.4028/www.scientific.net/AMR.146-147.1170 ↗
33. Quantitative regulation of bone-mimetic, oriented collagen/apatite matrix structure depends on the degree of osteoblast alignment on oriented collagen substrates10.1002/jbm.a.35189 ↗
34. Meischel, M., Tschegg, S., Berger, L., Tschegg, E., Fischer, L., Prohaska, T., Martinelli, E., Weinberg, A., 2013. Biodegradable bone implants. In: Proceedings of the Jahrestagung Der Deutschen Gesellschaft für Biomaterialien in Erlangen. Erlangen.
35. Effect of electrodeposition current density on the microstructure and the degradation of electroformed iron for degradable stents10.1016/j.mseb.2011.02.031 ↗
36. Bone-implant interface shear modulus and ultimate stress in a transcortical rabbit model of open-pore Ti6Al4V implants10.1016/j.jbiomech.2005.05.036 ↗
37. Stress shielding and bone resorption in shoulder arthroplasty10.1067/mse.2003.22 ↗
38. Effects of tissue preservation on murine bone mechanical properties10.1016/j.jbiomech.2008.09.037 ↗
39. Effects of surface treatments on high cycle corrosion fatigue of metallic implant materials10.1016/S0142-1123(00)00057-8 ↗
40. 3D printing of poly(3-hydroxybutyrate) porous structures using selective laser sintering10.1002/masy.201100237 ↗
41. Evoluation of the internal fixation of long bone fractures10.1302/0301-620X.84B8.0841093 ↗
42. Polyhydroxyalkanoates: an overview10.1016/S0960-8524(02)00212-2 ↗
43. In vivo and in vitro degradation of poly(3-hydroxybutyrate) in rat10.1016/0142-9612(91)90039-D ↗
44. Analysis of the fatigue behavior characterized by stiffness and permanent deformation for different distal volar radius compressdion plates10.1016/j.msec.2009.07.013 ↗
45. The bone-implant interface - nanoscale analysis of clinically retrieved dental implants10.1016/j.nano.2014.05.015 ↗
46. Sinn, G., Gindl, M., Reiterer, A., Stanzl-Tschegg, S., 2004. Changes in the surface properties of wood due to sanding. Holzforschung. 10.1515/HF.2004.03810.1515/HF.2004.038 ↗
47. Magnesium and its alloys as orthopedic biomaterials: a review10.1016/j.biomaterials.2005.10.003 ↗
48. Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature10.1002/(SICI)1097-4636(199822)43:2<192::AID-JBM14>3.0.CO;2-K ↗
49. An evaluation of variables influencing implant fixation by direct bone apposition10.1002/jbm.820190802 ↗
50. Mechanical and fracture-mechanical properties of asphalt-concrete interfaces
51. Characterization methods of bone-implant-interfaces of bioresorbable and titanium implants by fracture mechanical means10.1016/j.jmbbm.2010.08.004 ↗
52. Adhesive power measurements of bonds between old and new concrete10.1007/BF01143212 ↗
53. Effect of composition of poly(3-hydroxybutyrate-co-3-hydroxyhexanoate) on growth of fibriblast and osteoblast10.1016/j.biomaterials.2004.03.023 ↗
54. Micromotion-induced strain fields influence early stages of repair at bone-implant interfaces10.1016/j.actbio.2013.01.014 ↗
55. The importance of surface roughness for implant incorporation10.1016/S0890-6955(97)00114-4 ↗
56. In vivo corrosion of four magnesium alloys anf the associated bone response10.1016/j.biomaterials.2004.09.049 ↗
57. Effect of surface topology on the osseointegration of implant materials in trabecular bone10.1002/jbm.820291213 ↗
58. Design, construction and mechanical testing of digital 3D anatomical data-based PCL-HA bone tissue engineering scaffold10.1007/s10856-014-5360-8 ↗
59. In vivo evaluation of biodegradable magnesium alloy bone implant in the first 6 months implantation10.1002/jbm.a.32132 ↗