Strength and reliability of alumina ceramic femoral heads: Review of design, testing, and retrieval analysis
Source abstract
Abstract
Ceramic components are used increasingly in total hip arthroplasty (THA). Compared with metallic components, ceramic femoral heads for THA have the potential advantage of lower wear rates in articulations with acetabular liners. However, the use of ceramic components is also associated with unique risks, including sudden fracture and intolerable bearing noise or squeaking. This review paper summarizes the published literature regarding alumina ceramic femoral heads and tries to identify areas where uncertainties remain. We will discuss the following topics: (1) the fracture mechanics of ceramic materials; (2) design-related stresses acting on the femoral head, especially at the interface between the stem trunnion and the head; (3) gradual loss of strength in service by fatigue or slow crack growth and simulation using a proof test; and (4) information that can be gathered from examination of fractured explants.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (38)
Show source-provided references
1. Stress concentration regions in ceramic ball heads for total hip replacement considering trauma-like loading10.4028/www.scientific.net/MSF.486-487.185 ↗
2. Alumina and its use in surgery of the hip. (Experimental study)
3. Total arthroplasty of the hip by sintered aluminum prosthesis. Experimental study and 1st clinical applications
4. Femoral head technologies to reduce polyethylene wear in total hip arthroplasty
5. Crack growth resistance of alumina, zirconia and zirconia toughened alumina ceramics for joint prostheses10.1016/S0142-9612(01)00206-X ↗
6. Do ceramic components of hip joints maintain their strength in human bodies
7. Slow crack growth behavior of alumina ceramics10.1557/JMR.2000.0024 ↗
8. Subcritical crack growth in dense alumina exposed to physiological media10.1111/j.1151-2916.1980.tb10205.x ↗
9. Fréchette · 1990
10. Ceramic femoral head fractures in total hip arthroplasty10.1097/00003086-199607000-00021 ↗
11. Ceramic component fracture: trends and recommendations with modern components based on improved reporting methods10.1007/3-7985-1540-9_25 ↗
12. Fractures of ceramic bearings: history and present status10.1097/01.blo.0000096806.78689.50 ↗
13. Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup10.1016/j.arth.2006.05.020 ↗
14. In-vitro testing and validation of zirconia toughened alumina (ZTA)
15. Isolated fracture of the ceramic head after third-generation alumina-on-alumina total hip arthroplasty10.2106/JBJS.F.01489 ↗
16. The wear of ultra-high molecular weight polyethylene sliding on metallic and ceramic counterfaces representative of current femoral surfaces in joint replacement10.1243/0954411971534647 ↗
17. Influence of the ball diameter, material and taper design on transferred stresses of ceramic THR10.4028/www.scientific.net/MSF.544-545.375 ↗
18. Development of a new zirconia-toughened alumina: promising mechanical properties and absence of in vitro carcinogenicity10.1177/039463200902200323 ↗
19. Delayed fracture of the ceramic femoral head after trauma10.1054/arth.2002.32185 ↗
20. Clinical results of alumina-on-alumina couple in total hip replacement
21. Spontaneous fracture of the ceramic ball in a ceramic-polyethylene total hip arthroplasty10.1016/S0883-5403(05)80089-0 ↗
22. Fractography of ceramic femoral heads
23. Relevance of the insertion force for the taper lock reliability of a hip stem and a ceramic femoral head10.1515/BMT.2006.018 ↗
24. Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing
25. Quinn, G.D., 2007. Fractography of ceramics and glasses. Special Publication 960-16. National Institute of Standards and Technology.
26. Fracture of an alumina femoral head used in ceramic-on-ceramic total hip arthroplasty10.1016/j.arth.2007.09.017 ↗
27. Proof-testing of ceramic hip joint heads
28. Fractography of bioceramics10.4028/www.scientific.net/KEM.223.157 ↗
29. Improving the reliability of the ceramic-on-ceramic wear couple in THR
30. Characterization of mechanical properties of alumina based hip joint prostheses
31. Shikata, T., Oonishi, H., Hashimato, Y., et al. 1977. Wear resistance of irradiated UHMW polyethylenes to Al2O3 ceramics in total hip prostheses. Transactions of the 3rd Annual Meeting of the Society for Biomaterials, 118.
32. Ceramic total hip arthroplasty in the United States: safety and risk issues revisited
33. Mechanical strength of used ceramic femoral heads10.1177/112070000301300208 ↗
34. Fracture of a ceramic femoral head after ceramic-on-ceramic total hip arthroplasty10.1016/j.arth.2005.03.045 ↗
35. Squeaking hips10.2106/JBJS.H.00867 ↗
36. Failure analysis of in vivo fractured ceramic femoral heads10.1016/j.engfailanal.2008.07.015 ↗
37. Contemporary alumina-on-alumina total hip arthroplasty performed in patients younger than forty years: a 5-year minimum follow-up study10.1002/jbm.b.30457 ↗
38. Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study10.2106/00004623-200503000-00008 ↗