journal article·2008-10

Development of a new niobium-based alloy for vascular stent applications

Barry O’Brien, Jon Stinson, William Carroll

Publisher page

Abstract

This study was performed in order to develop a new stent material that would provide reduced MR image artifact compared to current stent materials. Alloy design rationale is initially presented and following this the development of a Nb-28 Ta-3.5 W-1.3 Zr alloy is described, including the manufacture of stent tubing. Tensile testing of this new alloy showed that it had approximately twice the yield strength of current Nb-1 Zr material with a 25% higher elastic modulus. The new alloy was also confirmed to have suitably low magnetic susceptibility. Mechanical testing of demonstration coronary stents made from the new alloy were shown to have acceptable compression strength and elastic recoil performance. It is concluded that this new Nb-28 Ta-3.5 W-1.3 Zr alloy is a practical candidate stent material for both coronary applications and peripheral uses such as carotid or intracranial stenting, where reduced MR image artifact would be beneficial.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Hydrogen embrittlement and corrosion behaviors in metals · Titanium Alloys Microstructure and Properties · High Temperature Alloys and Creep

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (43)

Show source-provided references
1. ASM Handbook, Tenth ed. Vol. 2. Properties and Selection: Nonferrous Alloys and Special-Purpose Materials. ASM International 1990. ISBN 0-87170-378-5
2. Elastic recoil of coronary stents: A comparative analysis10.1002/(SICI)1522-726X(200005)50:1<112::AID-CCD25>3.0.CO;2-X ↗
3. MR imaging of vascular stents: Effects of susceptibility, flow and radiofrequency eddy currents10.1016/S1051-0443(07)61918-6 ↗
4. First in-human randomized comparison of an anodized niobium stent versus a standard stainless steel stent10.1007/s00392-006-0406-7 ↗
5. Artifact-free in-stent lumen visualization by standard magnetic resonance angiography using a new metallic magnetic resonance imaging stent10.1161/01.CIR.0000015852.60212.DB ↗
6. New metallic MR stents for artifact-free coronary MR angiography: Feasibility study in a Swine model10.1097/01.rli.0000116894.48384.c2 ↗
7. 1994
8. Oxidation of refractory metals in air and low pressure oxygen gas10.1016/S0263-4368(00)00026-3 ↗
9. Biocompatibility of β-stabilizing elements of titanium alloys
10. Diagnostic accuracy of 64-slice computed tomography for detecting angiographically significant coronary artery stenosis in an unselected consecutive patient population10.1253/circj.70.564 ↗
11. Contrast-enhanced MR angiography at 1.5T after implantation of platinum stents: In vitro and in vivo comparison with conventional stent designs10.2214/ajr.184.1.01840288 ↗
12. Hubbell, J.H., Seltzer, S.M., Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients (version 1.4). [Online] Available: http://physics.nist.gov/xaamdi [2007, August 28]. National Institute of Standards and Technology, Gaithersburg, MD, 2004
13. Coronary arterial stents: Safety and artifacts during MR imaging10.1148/radiology.216.3.r00se03781 ↗
14. A removal torque and histomorphometric study of commercially pure niobium and titanium implants in rabbit bone10.1034/j.1600-0501.1991.020103.x ↗
15. Stent appearance at contrast-enhanced MR angiography: In vitro examination with 14 stents10.1148/radiology.217.1.r00se28173 ↗
16. Strecker stent implantation in iliac arteries: Patency and predictive factors for long-term success10.1148/radiology.194.3.7862972 ↗
17. Coronary stents: A materials perspective10.1016/j.biomaterials.2006.11.042 ↗
18. Biocompatibility and osteogenesis of refractory metal implant, titanium, hafnium, niobium, tantalum and rhenium10.1016/S0142-9612(00)00275-1 ↗
19. The influence of niobium and vanadium on passivity of titanium-based implants in physiological solution10.1016/S0142-9612(03)00252-7 ↗
20. MR evaluation of stent patency: In vitro tests of 22 metallic stents and the possibility of determining their patency by MR angiography10.1097/00004424-200012000-00008 ↗
21. Mechanical biocompatibilities of titanium alloys for biomedical applications10.1016/j.jmbbm.2007.07.001 ↗
22. MRI artifacts of metallic stents derived from imaging sequencing and the ferromagnetic nature of materials
23. Technology insight: An overview of research in drug-eluting stents10.1038/ncpcardio0378 ↗
24. Degradation of tungsten coils implanted into the subclavian artery of New Zealand white rabbits is not associated with local or systemic toxicity10.1016/S0142-9612(02)00352-6 ↗
25. Biocompatibility of corroding tungsten coils: In vitro assessment of degradation kinetics and cytotoxicity on human cells10.1016/S0142-9612(03)00274-6 ↗
26. Stent tubing: Understanding the desired attributes
27. The NUGGET study: NIR ultra gold-gilded equivalency trial10.1002/ccd.20026 ↗
28. MR angiography in the evaluation of atherosclerotic peripheral vascular disease10.1148/radiology.214.2.r00fe42325 ↗
29. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds10.1118/1.597854 ↗
30. Non-invasive imaging of coronary arteries: Current and future role of multi-detector row CT10.1148/radiol.2321021803 ↗
31. Solid solution strengthening of high purity niobium alloys10.1007/BF02644520 ↗
32. Metallic renal artery MR imaging stent: Artifact-free Lumen visualization with projection and standard renal MR angiography10.1148/radiol.2273020354 ↗
33. Artifact-Free coronary magnetic resonance angiography and coronary vessel wall imaging in the presence of a new, metallic, coronary resonance imaging stent10.1161/01.CIR.0000156462.97532.8F ↗
34. Self-expanding nitinol stents: Material and design considerations10.1007/s00330-003-2022-5 ↗
35. MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents and coils10.1148/radiology.166.3.3340759 ↗
36. Clinical experience with the Wiktor stent in native coronary arteries and coronary bypass grafts10.1136/hrt.72.3.288 ↗
37. A precious metal alloy for construction of MR imaging-compatible balloon expandable vascular stents10.1148/radiology.219.1.r01ap05284 ↗
38. Aneurysm clips made of titanium: Magnetic characteristics and artifacts in MR
39. Radiopacity of current endovascular stents: Evaluation in a multiple reader phantom study10.1097/01.RVI.0000136985.06722.AD ↗
40. High temperature niobium alloys
41. Coronary MR angiography
42. Effects of Ta content on Young’s modulus and tensile properties of binary Ti–Ta alloys for biomedical applications10.1016/j.msea.2003.12.011 ↗
43. The electrochemical behavior of metallic implant materials as an indicator of their biocompatibility10.1002/jbm.820210705 ↗