journal article·2012-05

Development of thermo-mechanical processing for fabricating highly durable β -type Ti–Nb–Ta–Zr rod for use in spinal fixation devices

Kengo Narita, Mitsuo Niinomi, Masaaki Nakai, Junko Hieda, Kazuya Oribe

Publisher page

Abstract

The mechanical strength of a beta titanium alloy such as Ti-Nb-Ta-Zr alloy (TNTZ) can be improved significantly by thermo-mechanical treatment. In this study, TNTZ was subjected to solution treatment, cold caliber rolling, and cold swaging before aging treatment to form a rod for spinal fixation. The {110}(β) are aligned parallel to the cross-section with two strong peaks approximately 180° apart, facing one another, in the TNTZ rods subjected to cold caliber rolling and six strong peaks at approximately 60° intervals, facing one another, in the TNTZ rods subjected to cold swaging. Therefore, the TNTZ rods subjected to cold swaging have a more uniform structure than those subjected to cold caliber rolling. The orientation relationship between the α and β phases is different. A [110](β)//[121](α), (112)(β)//(210)(α) orientation relationship is observed in the TNTZ rods subjected to aging treatment at 723 K after solution treatment and cold caliber rolling. On the other hand, a [110](β)//[001](α), (112)(β)//(200)(α) orientation relationship is observed in TNTZ rod subjected to aging treatment at 723 K after cold swaging. A high 0.2% proof stress of about 1200 MPa, high elongation of 18%, and high fatigue strength of 950 MPa indicate that aging treatment at 723 K after cold swaging is the optimal thermo-mechanical process for a TNTZ rod.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Titanium Alloys Microstructure and Properties · Orthopaedic implants and arthroplasty · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (18)

Show source-provided references
1. Improvement in fatigue characteristics of newly developed beta type titanium alloy for biomedical applications by thermo-mechanical treatments
2. Notch fatigue properties of a Ti–29Nb–13Ta–4.6Zr alloy for biomedical applications10.2464/jilm.55.575 ↗
3. Omega phase embrittlement in aged Ti–15%Mo10.1016/0036-9748(71)90258-4 ↗
4. Microstructures and properties of biomedical TiNbZrFe β-titanium alloy under ageing conditions10.1016/j.msea.2009.08.057 ↗
5. A comparison of radiographic findings in fusion and nonfusion patients ten or more years following lumbar disc surgery10.1097/00007632-197909000-00008 ↗
6. Treatment of Scoliosis: correction and internal fixation by spine instrumentation
7. Effects of microstructural modification on fatigue crack growth resistance of Ti–15V–3Al–3Sn–3Cr10.1016/S0142-1123(99)00064-X ↗
8. Biomechanical effect of spinal fusion and dynamic stabilization on the adjacent segments
9. Design and mechanical properties of new [beta] type titanium alloys for implant materials10.1016/S0921-5093(97)00808-3 ↗
10. ω-assisted nucleation and growth of [alpha] precipitates in the Ti–5Al–5Mo–5V–3Cr–0.5Fe [beta] titanium alloy10.1016/j.actamat.2009.01.007 ↗
11. Heterogeneous a phase precipitation and peculiar ageing strengthening in biomedical β-type Ti–Nb–Ta–Zr alloy having vortical structure
12. Fatigue performance and cyto-toxicity of low rigidity titanium alloy, Ti–29Nb–13Ta–4.6Zr10.1016/S0142-9612(03)00069-3 ↗
13. Mechanical biocompatibilities of titanium alloys for biomedical applications10.1016/j.jmbbm.2007.07.001 ↗
14. Development of low rigidity b-type titanium alloy for biomedical applications10.2320/matertrans.43.2970 ↗
15. The biological response to titanium and titanium–aluminium–vanadium alloy particles: I. Tissue culture studies10.1016/0142-9612(86)90085-2 ↗
16. The biological response to titanium and titanium–aluminium–vanadium alloy particles: II. Long-term animal studies10.1016/0142-9612(86)90086-4 ↗
17. Microstructure and texture evolution during the drawing of tungsten wires10.1016/j.engfracmech.2009.02.012 ↗
18. Experiment study on fracture fixation with low rigidity titanium alloy