journal article·2021-03Open access

A finite element investigation on design parameters of bare and polymer-covered self-expanding wire braided stents

Ciara G. McKenna, Ted J. Vaughan

Abstract

Self-expanding covered braided stents are routinely used across a diverse range of clinical applications, but few computational studies have attempted to replicate their complex behaviour. In this study, a computational framework was developed to predict the functional performance of bare and covered self-expanding wire braided stents, with a systematic evaluation on the effect of various braid and cover parameters presented. Simulated radial force and kink deformation tests show good agreement to experimental data for covered braided stents across a range of braid angles and cover thicknesses. Our results demonstrate that braid angle is a key governing parameter that dictates the radial and kink performance of both bare-metal and covered wire braided stents. It was also demonstrated that addition of a polymeric cover to a wire braided stent causes a stiffer radial response across all braid angles, particularly when thicker and/or stiffer covering systems were considered. This study represents the first experimentally-validated computational model for covered wire braided stent systems and has excellent potential to be used in future design of these devices for a range of applications.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Aortic Disease and Treatment Approaches · Aortic aneurysm repair treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. A numerical framework for the mechanical analysis of dual-layer stents in intracranial aneurysm treatment10.1016/j.jbiomech.2016.02.026 ↗
2. Shape-memory alloys: modelling and numerical simulations of the finite-strain superelastic behavior10.1016/S0045-7825(96)01147-4 ↗
3. Supera self-expanding stents for endovascular treatment of femoropopliteal disease: a review of the clinical evidence
4. Virtual optimization of self-expandable braided wire stents10.1016/j.medengphy.2008.11.008 ↗
5. Filling the void: a coalescent numerical and experimental technique to determine aortic stent graft mechanics10.1016/j.jbiomech.2013.07.010 ↗
6. Computational comparison of the bending behavior of aortic stent-grafts10.1016/j.jmbbm.2011.09.006 ↗
7. Numerical investigations of the mechanical properties of braided vascular stents
8. Three-year results of the VIBRANT trial of VIABAHN endoprosthesis versus bare nitinol stent implantation for complex superficial femoral artery occlusive disease10.1016/j.jvs.2013.01.050 ↗
9. Finite element analysis of covered microstents10.1016/j.jbiomech.2004.06.008 ↗
10. Buckling of thin-walled long steel cylinders subjected to bending10.1115/1.4002902 ↗
11. A study of the geometrical and mechanical properties of a self-expanding metallic stent—theory and experiment10.1002/jab.770040111 ↗
12. Engineering Comparison of computational modelling techniques for braided stent analysis
13. Mechanical modeling of self-expandable stent fabricated using braiding technology10.1016/j.jbiomech.2008.08.005 ↗
14. Covered self-expandable metal stents with an anti-migration system improve patency duration without increased complications compared with uncovered stents for distal biliary obstruction caused by pancreatic carcinoma: a randomized multicenter trial10.1038/ajg.2013.305 ↗
15. Computational mechanics of Nitinol stent grafts10.1016/j.jbiomech.2008.05.032 ↗
16. Heparin-bonded covered stents versus bare-metal stents for complex femoropopliteal artery lesions the randomized VIASTAR trial (viabahn endoprosthesis with PROPATEN bioactive surface [VIA] versus bare nitinol stent in the treatment of long lesions in supe10.1016/j.jacc.2013.05.079 ↗
17. Liu · 2017
18. Computer modeling of deployment and mechanical expansion of neurovascular flow diverter in patient-specific intracranial aneurysms10.1016/j.jbiomech.2012.06.013 ↗
19. Biliary and pancreatic stenting: devices and insertion techniques in therapeutic endoscopic retrograde cholangiopancreatography and endoscopic ultrasonography10.4253/wjge.v8.i3.143 ↗
20. Evaluation of cover effects on bare stent mechanical response10.1016/j.jmbbm.2016.04.023 ↗
21. An experimental evaluation of the mechanics of bare and polymer-covered self-expanding wire braided stents10.1016/j.jmbbm.2019.103549 ↗
22. Analytical models for predicting mechanical properties of mesh-type self-expandable metal stents with cover membrane10.1016/j.cap.2007.12.004 ↗
23. A comparison of covered vs bare expandable stents for the treatment of aortoiliac occlusive disease10.1016/j.jvs.2011.06.097 ↗
24. Ni · 2015 · vol. 18 · 1117
25. Influence of design parameters on the mechanical behavior and porosity of braided fibrous stents10.1016/j.matdes.2015.07.051 ↗
26. Management of inoperable malignant oesophageal strictures with fully covered WallFlex® stent: a multicentre prospective study10.1016/j.dld.2014.08.037 ↗
27. Mechanical properties of stent–graft materials
28. Partially versus fully covered self-expanding metal stents for benign and malignant esophageal conditions: a single center experience10.1007/s00464-012-2738-x ↗
29. Looped ends versus open ends braided stent: a comparison of the mechanical behaviour using analytical and numerical methods10.1016/j.jmbbm.2017.08.025 ↗
30. Self-expanding nitinol stents: material and design considerations10.1007/s00330-003-2022-5 ↗
31. Selection of helical braided flow diverter stents based on hemodynamic performance and mechanical properties10.1136/neurintsurg-2016-012561 ↗
32. Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device10.3174/ajnr.A2023 ↗
33. Fully-covered esophageal stent migration rates in benign and malignant disease: a multicenter retrospective study
34. A multicenter prospective study of the short-term outcome of a newly developed partially covered self-expandable metallic biliary stent (WallFlex®)10.1007/s10620-011-1581-6 ↗
35. Verhegghe · 2019
36. Wahl · 1963
37. Deformation mechanisms of prototype composite braided stent-grafts in bending fatigue for peripheral artery application10.1016/j.jmbbm.2017.10.037 ↗
38. Modeling of braided stents: comparison of geometry reconstruction and contact strategies10.1016/j.jbiomech.2020.109841 ↗
39. Mechanical performance of cup-spherical-shaped and straight form braided esophageal stent
40. Mechanical characterization of braided self-expanding stents: impact of design parameters10.1142/S0219519419500386 ↗
41. Mechanical characteristics of novel polyester/NiTi wires braided composite stent for the medical application10.1016/j.rinp.2016.07.007 ↗