journal article·2014-12

Multi-scale mechanical investigation of stainless steel and cobalt–chromium stents

Konstantinos Kapnisis, Georgios Constantinides, Harry Georgiou, Daniel Cristea, Camelia Gabor, Daniel Munteanu, Brigitta Brott, Peter Anderson, Jack Lemons, Andreas Anayiotos

Publisher page

Abstract

In-stent restenosis (ISR) remains a significant limitation despite the considerable previous clinical and investigative emphasis on the problem. Complications arising from the interaction of stent materials with the surrounding vessel wall as well as from the mechanical forces developing after implantation, play an important role in the development of ISR. To investigate the relation between mechanical factors and stent structural integrity, and to identify any structural weakness points on the geometry of commercially available Stainless Steel and Cobalt-Chromium stents, accelerated pulsatile durability tests were carried out in a simulated physiological environment. Potential spatial variations in the mechanical properties on stent struts and their role in the observed premature failures of the stent devices during operation were also examined. Fretting wear and fatigue-induced fractures were found on stent surfaces after exposure to cyclic loading similar to that arising in vivo. Nanoindentation studies performed on various locations along the stent struts have shown that the hardness of specific stent locations significantly increases after mechanical expansion. The increase in hardness was associated with a reduction of the material's ability to dissipate energy in plastic deformations, therefore an increased vulnerability to fracture and fatigue. We conclude that the locations of fatigue fractures in stent struts are controlled not only by the geometrically-driven stress concentrations developing during cyclic loading but also by the local material mechanical changes that are imparted on various parts of the stent during the deployment process.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hydrogen embrittlement and corrosion behaviors in metals · Metal and Thin Film Mechanics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (62)

Show source-provided references
1. AAMI Standard 25539-2, 2008. Cardiovascular Implants – Endovascular Devices – Part 2: Vascular Stents. Association for the Advancement of Medical Instrumentation, Arlington, VA. 〈http://www.aami.orgS〉.
2. Current treatment of in-stent restenosis10.1016/j.jacc.2014.02.545 ↗
3. Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation
4. Fatigue life assessment of cardiovascular balloon-expandable stents: a two-scale plasticity–damage model approach10.1016/j.jmbbm.2012.06.011 ↗
5. ASTM Standard F138-13a, 2013. Standard Specification for Wrought 18Chromium–14Nickel–2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673). ASTM International. doi:10.1520/F0138.10.1520/F0138 ↗
6. ASTM Standard F139-12, 2012. Standard Specification for Wrought 18Chromium–14Nickel–2.5Molybdenum Stainless Steel Sheet and Strip for Surgical Implants (UNS S31673). ASTM International. doi:10.1520/F0139-12.10.1520/F0139-12 ↗
7. ASTM Standard F90-09, 2009. Standard Specification for Wrought Cobalt–20Chromium–15Tungsten–10Nickel Alloy for Surgical Implant Applications (UNS R30605). ASTM International. doi:10.1520/F0090-09.10.1520/F0090-09 ↗
8. ASTM Standard F2477-07, 2007. Standard Test Methods for In Vitro Pulsatile Durability Testing of Vascular Stents. ASTM International, West Conshohocken, PA. 〈10.1520/F2477-07〉, /〈http://www.astm.orgS〉.
9. Influences of stress on the measurements of mechanical properties using nanoindentation: Part II. Finite element simulations10.1557/JMR.1996.0092 ↗
10. Brunski · 2004 · 137
11. Enhanced inflammatory response to coronary stenting marks the development of clinically relevant restenosis
12. Coronary stent fracture: the hidden truth of a problem more common than stent thrombosis
13. Meta-analysis of incidence, clinical characteristics and implications of stent fracture10.1016/j.amjcard.2010.06.010 ↗
14. Scaling, dimensional analysis, and indentation measurements10.1016/j.mser.2004.05.001 ↗
15. Grid indentation analysis of composite microstructure and mechanics: principles and validations10.1016/j.msea.2006.05.125 ↗
16. Mechanisms of smooth muscle cell proliferation and endothelial regeneration after vascular injury and stenting: approach to therapy10.1253/circj.CJ-11-0366 ↗
17. The consequences of the mechanical environment of peripheral arteries for nitinol stenting10.1007/s11517-011-0815-2 ↗
18. Fischer-Cripps · 2011
19. Simulation of stent deployment in a realistic human coronary artery
20. The role of vascular calcification in inducing fatigue and fracture of coronary stents10.1002/jbm.b.31911 ↗
21. Clinical device-related article. Surface characterization of explanted endovascular stents: evidence of in vivo corrosion10.1002/jbm.b.31698 ↗
22. In-vivo corrosion and local release of metallic ions from vascular stents into surrounding tissue
23. Materials for metallic stents10.1007/s10047-008-0456-x ↗
24. Detailed in vivo visualization of stent fracture causing focal restenosis using 3D reconstruction software for high-resolution optical coherence tomography images10.1093/ehjci/jet273 ↗
25. Influence of Stent Fracture on the Long-Term Patency in the Femoro-Popliteal Artery10.1016/j.jcin.2009.04.014 ↗
26. Partial strut fracture in everolimus-eluting stents: an intravascular ultrasound study
27. Fracture potentials of five drug-eluting-stent platforms in proximal right coronary artery replica under in vivo simulated cyclically bended environment10.1016/S0735-1097(10)61265-X ↗
28. Fatigue-life assessment and validation techniques for metallic vascular implants10.1016/j.biomaterials.2009.10.028 ↗
29. Stent overlapping and geometric curvature influence the structural integrity and surface characteristics of coronary nitinol stents10.1016/j.jmbbm.2012.11.006 ↗
30. OCT-verified neointimal hyperplasia is increased at fracture site in drug-eluting stents10.1016/j.jcmg.2011.07.014 ↗
31. Guidant multi-link vision stent registry investigators. Usefulness of a cobalt chromium coronary stent alloy10.1016/S0002-9149(03)00669-6 ↗
32. Effect of strut thickness on neointimal atherosclerotic change over an extended follow-up period (≥4 years) after bare-metal stent implantation: intracoronary optical coherence tomography examination10.1016/j.ahj.2012.01.007 ↗
33. Magnetic field interactions of orthodontic wires during magnetic resonance imaging (MRI) at 1.5T10.1007/s00056-005-0505-5 ↗
34. Intravascular ultrasound assessed incomplete stent apposition and stent fracture in stent thrombosis after bare metal versus drug-eluting stent treatment the Nordic Intravascular Ultrasound Study (NIVUS)10.1016/j.ijcard.2012.10.033 ↗
35. Frequency of stent fracture as a cause of coronary restenosis after sirolimus-eluting stent implantation10.1016/j.amjcard.2007.03.073 ↗
36. Thin-strut drug-eluting stents are more favorable for severe calcified lesions after rotational atherectomy than thick-strut drug-eluting stents
37. Materials, fluid dynamics, and solid mechanics aspects of coronary artery stents: a state-of-the-art review10.1002/jbm.b.31028 ↗
38. Fatigue life analysis and experimental verification of coronary stent10.1007/s00380-009-1203-9 ↗
39. Stent fracture: insights on mechanisms, treatments, and outcomes from the food and drug administration manufacturer and user facility device experience database
40. Coronary stents: a materials perspective10.1016/j.biomaterials.2006.11.042 ↗
41. Fatigue and life prediction for cobalt–chromium stents: a fracture mechanics analysis10.1016/j.biomaterials.2005.10.012 ↗
42. In situ fracture of stents implanted for relief of pulmonary arterial stenosis in patients with congenitally malformed hearts10.1017/S1047951108002424 ↗
43. Mechanical behavior of coronary stents investigated through the finite element method10.1016/S0021-9290(02)00033-7 ↗
44. Coronary artery bifurcation biomechanics and implications for interventional strategies
45. Comparison of second-generation stents for application in the superficial femoral artery: an in vitro evaluation focusing on stent design10.1583/10-3069.1 ↗
46. Incidence and predictors of drug-eluting stent fracture in human coronary artery. A pathologic analysis10.1016/j.jacc.2009.05.075 ↗
47. Measurement of hardness and elastic modulus by instrumented indentation: advances in understanding and refinements to methodology10.1557/jmr.2004.19.1.3 ↗
48. Stainless steel ions stimulate increased thrombospondin-1-dependent TGF-beta activation by vascular smooth muscle cells: implications for in-stent restenosis10.1159/000265565 ↗
49. Simulation of longitudinal stent deformation in a patient-specific coronary artery10.1016/j.medengphy.2014.02.004 ↗
50. Development and assessment of 316LVM cardiovascular stents10.1016/S0921-5093(04)00974-8 ↗
51. What can we learn from explanted endovascular devices?10.1053/ejvs.2002.1677 ↗
52. Mechanical behaviours of coronary stents10.4028/www.scientific.net/AST.49.91 ↗
53. Impact of strut thickness on late luminal loss after coronary artery stent placement10.1016/j.amjcard.2003.10.049 ↗
54. Effects of the substrate on the determination of thin film mechanical properties by nanoindentation10.1016/S1359-6454(01)00328-7 ↗
55. In vitro host response assessment of biomaterials for cardiovascular stent manufacture10.1023/B:JMSM.0000021123.51752.11 ↗
56. Restenosis following implantation of bare metal coronary stents: pathophysiology and pathways involved in the vascular response to injury10.1016/j.addr.2006.01.015 ↗
57. Stent fracture and restenosis in the drug- eluting stent era
58. Hardness measurements for the determination of residual stresses
59. Mechanical behavior of fully expanded commercially available endovascular coronary stents
60. Influences of stress on the measurements of mechanical properties using nanoindentation: Part I. Experimental studies in an aluminum alloy10.1557/JMR.1996.0091 ↗
61. Dedicated bifurcation analysis: basic principles10.1007/s10554-010-9795-9 ↗
62. US Food · 2010