Mechanical assessment of grit blasting surface treatments of dental implants
Source abstract
Abstract
This paper investigates the influence of surface preparation treatments of dental implants on their potential (mechanical) fatigue failure, with emphasis on grit-blasting. The investigation includes limited fatigue testing of implants, showing the relationship between fatigue life and surface damage condition. Those observations are corroborated by a detailed failure analysis of retrieved fracture dental implants. In both cases, the negative effect of embedded alumina particles related to the grit-blasting process is identified. The study also comprises a numerical simulation part of the grit blasting process that reveals, for a given implant material and particle size, the existence of a velocity threshold, below which the rough surface is obtained without damage, and beyond which the creation of significant surface damage will severely reduce the fatigue life, thus increasing fracture probability. The main outcome of this work is that the overall performance of dental implants comprises, in addition to the biological considerations, mechanical reliability aspects. Fatigue fracture is a central issue, and this study shows that uncontrolled surface roughening grit-blasting treatments can induce significant surface damage which accelerate fatigue fracture under certain conditions, even if those treatments are beneficial to the osseointegration process.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (39)
Show source-provided references
1. version 612 –2. Abaqus documentation. Abaqus analysis user׳s manual
2. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man10.3109/17453678108991776 ↗
3. Oral implant surfaces: part 1-review focusing on topographic and chemical properties of different surfaces and in vivo responses to them
4. Anderson, C.E., Walker, J.D., Lankford, J. 1995. Investigations of the ballistic response of brittle materials. Technical report. SwRI Project 06-5117/002, Southwest Research Institute, Materials and Structures Division, San Antonio, TX 78228-0510.
5. The effect of sandblasting treatment on endurance properties of titanium alloy hip prostheses10.1046/j.1525-1594.2000.06486.x ↗
6. Roughening of metallic biomaterials by abrasiveless waterjet peening: characterization and viability10.1016/j.wear.2011.01.024 ↗
7. Fatigue behavior of Ti–6Al–4V and 316 LVM blasting with ceramic particles of interest for medical devices10.1016/j.jmbbm.2013.10.013 ↗
8. Stamping formability of pure titanium sheets10.1016/j.jmatprotec.2005.05.004 ↗
9. Analyses of fractured implant fixture after prolonged implantation10.1007/BF03185981 ↗
10. Rough surfaces of titanium and titanium alloys for implants and prostheses10.1016/j.msec.2004.08.004 ↗
11. Biomedical applications of titanium and its alloys10.1007/s11837-008-0031-1 ↗
12. Improving osseointegration of dental implants10.1586/erd.09.74 ↗
13. Surface treatments of titanium dental implants for rapid osseointegration10.1016/j.dental.2006.06.025 ↗
14. Stress singularity in a two-material wedge10.1007/BF00184307 ↗
15. Henley · 1981
16. Constitutive modeling of aluminum nitride for large strain, high-strain rate, and high-pressure applications Int10.1016/S0734-743X(00)00046-4 ↗
17. ISO 14801. 2007. Fatigue test for endosseous dental implants. Geneva: international organization for standardization.
18. ISO B600. 2007. Surface preparation and marking of metallic surgical impalnts. Geneva: International Organization for Standardization.
19. ISO F86. 2007. Surface preparation and marketing of metallic surgical implants. Geneva: International Organization for Standardization.
20. An improved computational constitutive model for brittle materials
21. Modes of Fracture
22. Fatigue and cyclic deformation behavior of surface-modified titanium alloys in simulated physiological media10.1016/j.biomaterials.2005.08.012 ↗
23. Fractures of the fixture component of an osseointegrated implant
24. Interrogations on the sub-surface strain hardening of grit blasted Ti–6Al–4V alloy10.1016/j.surfcoat.2009.02.003 ↗
25. Mechanical properties and deformation mechanisms of commercially pure titanium10.1016/S1359-6454(99)00203-7 ↗
26. The effect of machined topography and integrity on fatigue life10.1016/j.ijmachtools.2003.10.018 ↗
27. Effect of surface treatments on the fatigue life of titanium for biomedical applications10.1016/j.jmbbm.2010.03.006 ↗
28. A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FPDs) after a mean observation period of at least 5 years10.1111/j.1600-0501.2012.02546.x ↗
29. Meta-analysis of oral implant fracture incidence and related determinants10.1155/2014/263925 ↗
30. Thermo-mechanical aspects of adiabatic shear failure of AM50 and Ti6Al4V alloys10.1016/j.mechmat.2008.03.002 ↗
31. Geometrical imperfection and adiabatic shear banding10.1016/j.ijimpeng.2007.07.009 ↗
32. Scanning electron microscopy fractography analysis of fractured hollow implants10.1563/AAID-JOI-D-10-90000 ↗
33. Effect of dental implant diameter on fatigue performance. Part I: mechanical behavior10.1111/j.1708-8208.2012.00477.x ↗
34. Effect of dental implant diameter on fatigue performance. Part II: failure analysis10.1111/j.1708-8208.2012.00476.x ↗
35. Identification of failure mechanisms in retrieved fractured dental implants10.1016/j.engfailanal.2014.01.002 ↗
36. Suresh · 1994
37. Effects of titanium surface topography on bone integration: a systematic review10.1111/j.1600-0501.2009.01775.x ↗
38. On implant surfaces: a review of current knowledge and opinions
39. Fracture mechanisms of retrieved titanium screw thread in dental implant10.1016/S0142-9612(01)00380-5 ↗