Impact of surface porosity and topography on the mechanical behavior of high strength biomedical polymers
Source abstract
Abstract
The ability to control the surface topography of orthopedic implant materials is desired to improve osseointegration but is often at the expense of mechanical performance in load bearing environments. Here we investigate the effects of surface modifications, roughness and porosity, on the mechanical properties of a set of polymers with diverse chemistry and structure. Both roughness and surface porosity resulted in samples with lower strength, failure strain and fatigue life due to stress concentrations at the surface; however, the decrease in ductility and fatigue strength were greater than the decrease in monotonic strength. The fatigue properties of the injection molded polymers did not correlate with yield strength as would be traditionally observed in metals. Rather, the fatigue properties and the capacity to maintain properties with the introduction of surface porosity correlated with the fracture toughness of the polymers. Polymer structure impacted the materials relative capacity to maintain monotonic and cyclic properties in the face of surface texture and porosity. Generally, amorphous polymers with large ratios of upper to lower yield points demonstrated a more significant drop in ductility and fatigue strength with the introduction of porosity compared to crystalline polymers with smaller ratios in their upper to lower yield strength. The latter materials have more effective dissipation mechanisms to minimize the impact of surface porosity on both monotonic and cyclic damage.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials · Dental materials and restorations
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (79)
Show source-provided references
1. Surface roughness of implants: a review
2. Effect of annealing on the properties of poly(ether ether ketone)10.1002/pen.760310809 ↗
3. Effects of titanium surface roughness on mesenchymal stem cell commitment and differentiation signaling
4. Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial10.1302/0301-620X.81B5.0810907 ↗
5. Effect of surface finish on fatigue strength10.1016/0013-7944(94)00297-U ↗
6. Influence of stress concentrations on the mechanical property–porosity correlation in porous materials10.1023/A:1006503524162 ↗
7. The morphology of poly(aryl-ether-ether-ketone)10.1016/0032-3861(83)90144-1 ↗
8. Crystallinity in poly (aryl-ether-ketone) plaques studied by multiple internal reflection spectroscopy10.1007/BF00265483 ↗
9. Fatigue properties of carbon‐and porous‐coated Ti–6Al–4V alloy10.1002/jbm.820180504 ↗
10. The effect of molecular weight and crystallinity on the mechanical properties of injection moulded poly(aryl-ether-ether-ketone) resin10.1016/0032-3861(94)90057-4 ↗
11. Titanium alloys (AoN) and their involvement in osseointegration
12. Effect of surface roughness on osteogenesis in vitro and osseointegration in vivo of carbon fiber-reinforced polyetheretherketone-nanohydroxyapatite composite
13. High-strength, surface-porous polyether-ether-ketone for load-bearing orthopedic implants10.1016/j.actbio.2014.11.030 ↗
14. Implant osseointegration and the role of microroughness and nanostructures: lessons for spine implants10.1016/j.actbio.2014.03.037 ↗
15. Notch sensitivity and fractography of polyolefins10.1002/pen.760230303 ↗
16. A phenomenological model for predicting fatigue life in bovine trabecular bone10.1115/1.1762893 ↗
17. Gibson · 1997
18. Microstructured titanium regulates interleukin production by osteoblasts, an effect modulated by exogenous BMP-210.1016/j.actbio.2012.10.030 ↗
19. Direct three-dimensional morphometric analysis of human cancellous bone: microstructural data from spine, femur, iliac crest and calcaneus10.1359/jbmr.1999.14.7.1167 ↗
20. Effect of flexibility of the femoral stem on bone-remodeling and fixation of the stem in a canine total hip arthroplasty model without cement10.2106/00004623-199901000-00013 ↗
21. Simulation of elastic and plastic response in the glassy polycarbonate of 4,4׳-isopropylidenediphenol10.1021/ma00057a034 ↗
22. Fatigue crack propagation in polymeric materials10.1007/BF00556040 ↗
23. Jones · 2005
24. Surface roughness modulates the local production of growth factors and cytokines by osteoblast-like MG-63 cells10.1002/(SICI)1097-4636(199609)32:1<55::AID-JBM7>3.0.CO;2-O ↗
25. Microrough titanium surface affects biologic response in MG63 osteoblast‐like cells10.1002/jbm.a.31040 ↗
26. BMP‐induced osteogenesis on the surface of hydroxyapatite with geometrically feasible and nonfeasible structures: topology of osteogenesis10.1002/(SICI)1097-4636(199802)39:2<190::AID-JBM4>3.0.CO;2-K ↗
27. Porosity of 3D biomaterial scaffolds and osteogenesis10.1016/j.biomaterials.2005.02.002 ↗
28. Kinloch A. Fracture Behaviour of Polymers: Springer Netherlands; 1983.
29. The measurement of the crystallinity of polymers by DSC10.1016/S0032-3861(02)00235-5 ↗
30. Competition between plasticity-controlled and crack-growth controlled failure in static and cyclic fatigue of polymer systems10.1016/j.polymertesting.2016.01.008 ↗
31. A parametric study of the factors affecting the fatigue strength of porous coated Ti–6A1–4V implant alloy10.1002/jbm.820241106 ↗
32. The effect of physical ageing on the properties of amorphous PEEK10.1016/0032-3861(85)90136-3 ↗
33. Microrough implant surface topographies increase osteogenesis by reducing osteoclast formation and activity10.1002/jbm.a.30025 ↗
34. Response of MG63 osteoblast-like cells to titanium and titanium alloy is dependent on surface roughness and composition10.1016/S0142-9612(98)00144-6 ↗
35. Landel · 1993
36. Lohbauer · 2010
37. Lampman S. Fatigue Testing and Behavior. Characterization and Failure Analysis of Plastics: A S M International; 2003.10.31399/asm.tb.cfap.9781627082815 ↗
38. Current strategies to improve the bioactivity of PEEK10.3390/ijms15045426 ↗
39. Threaded versus porous-surfaced designs for implant stabilization in bone-endodontic implant model10.1002/jbm.820200907 ↗
40. Effect of polycaprolactone scaffold permeability on bone regeneration in vivo10.1089/ten.tea.2010.0560 ↗
41. Titanium Castings for Marine Propellers
42. Porosity in Polyaryletheretherketone
43. Effect of porosity on the tensile properties of low ductility aluminum alloys10.1590/S1516-14392004000200002 ↗
44. Fatigue behavior of some filled polymers10.1177/002199837500900205 ↗
45. Poly(aryl-ether-ether-ketone) and its advanced composites: a review10.1002/pc.750080202 ↗
46. The roles of Wnt signaling modulators Dickkopf-1 (Dkk1) and Dickkopf-2 (Dkk2) and cell maturation state in osteogenesis on microstructured titanium surfaces10.1016/j.biomaterials.2009.11.071 ↗
47. Integrin alpha2beta1 plays a critical role in osteoblast response to micron-scale surface structure and surface energy of titanium substrates10.1073/pnas.0805420105 ↗
48. Rough titanium alloys regulate osteoblast production of angiogenic factors10.1016/j.spinee.2013.03.047 ↗
49. Polymeric biomaterials for load-bearing medical devices10.1007/s11837-009-0126-3 ↗
50. Osseointegration: an Update10.1007/s13191-013-0252-z ↗
51. Pruitt · 2011
52. Influence of loading frequency on high cycle fatigue properties of b.c.c. and h.c.p. metals10.1016/S0921-5093(00)01978-X ↗
53. Thermal Plasma Spray Deposition of Titanium and Hydroxyapatite on Polyaryletheretherketone Implants
54. Regulation of angiogenesis during osseointegration by titanium surface microstructure and energy10.1016/j.biomaterials.2010.02.071 ↗
55. Evaluation and extension of physical property-porosity models based on minimum solid area10.1007/BF00355133 ↗
56. Comparison of stress-concentration versus minimum solid area based mechanical property porosity relations10.1007/BF00367582 ↗
57. Limitations of pore-stress concentrations on the mechanical properties of porous materials10.1023/A:1018674713006 ↗
58. Implant biomaterials: a comprehensive review10.12998/wjcc.v3.i1.52 ↗
59. Implant surface roughness and bone healing: a systematic review10.1177/154405910608500603 ↗
60. Effect of micrometer-scale roughness of the surface of Ti6Al4V pedicle screws in vitro and in vivo10.2106/JBJS.G.00499 ↗
61. In vivo performance of a modified CSTi dental implant coating
62. Morphometric and mechanical evaluation of titanium implant integration: comparison of five surface structures10.1002/(SICI)1097-4636(200007)51:1<15::AID-JBM3>3.0.CO;2-9 ↗
63. Suhr · 1986
64. Simske · 1997 · 151
65. Scheirs · 2000
66. Notch sensitivity of PEEK in monotonic tension10.1016/j.biomaterials.2009.08.020 ↗
67. Schaffer · 1999
68. The notch sensitivity of polymeric materials10.1002/app.1976.070200814 ↗
69. Fatigue of biomaterials: a review10.1016/S0142-1123(00)00052-9 ↗
70. Selective laser melting-produced porous titanium scaffolds regenerate bone in critical size cortical bone defects10.1002/jor.22293 ↗
71. Tailoring of mesenchymal stem cells behavior on plasma-modified polytetrafluoroethylene10.1002/adma.201104967 ↗
72. Weiss · 1999
73. Effect of surface-topology on the osseointegration of implant materials in trabecular bone10.1002/jbm.820291213 ↗
74. Wallenberger · 2009
75. Ward · 2012 · 319
76. Ward · 2012 · 379
77. The fatigue strength of porous‐coated Ti–6% Al–4% V implant alloy10.1002/jbm.820180908 ↗
78. Osteoblast‐like cells are sensitive to submicron‐scale surface structure10.1111/j.1600-0501.2005.01195.x ↗
79. Zhan · Numer. Charact. Mech. Prop. Metal. Nanowires · 2013