“Peroperative estimation of bone quality and primary dental implant stability”
Source abstract
Abstract
Objectives Dental implants are widely used to restore function and appearance. It may be essential to choose the appropriate drilling protocol and implant design in order to optimise primary stability. This could be achieved based on an assessment of the implantation site with respect to bone quality and objective biomechanical descriptors such as stiffness and strength of the bone-implant system. The aim of this ex vivo study is to relate these descriptors with bone quality, with a pre-implantation indicator of implant stability: pilot-hole drilling force (Fdrilling), and with two post-implantation indicators: maximal implantation torque (Timplantation) and resonance frequency analysis (RFA). Methods Eighty trabecular bone specimens were cored from human vertebrae and bovine tibiae. Bone volume fraction (BV/TV), a representative for bone quality, was obtained through micro-computed tomography scans. Implants were kept in controlled laboratory conditions following standard surgical procedures. Forces and torques were recorded and RFA was assessed after implantation. Off-axis compression tests were conducted on the implants until failure. Implant stability was identified by stiffness and ultimate force (Fultimate). The relationships between BV/TV, Stiffness, Fultimate and Fdrilling, Timplantation, RFA were established. Results Fdrilling correlated well with BV/TV of the implantation site (r2 = 0.81), stiffness (r2 = 0.75) and Fultimate (r2 = 0.80). Timplantation correlated better with stiffness (r2 = 0.86) and Fultimate (r2 = 0.94) than RFA (r2 = 0.77 and r2 = 0.74, respectively). Conclusion Our results indicate that BV/TV and bone-implant stability can be directly estimated by the force needed for the pilot drilling that occurs during the site preparation before implantation. Moreover, implantation torque outperforms RFA for evaluating the mechanical competence of the bone-implant system.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Dental Radiography and Imaging · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (54)
Show source-provided references
1. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw10.1016/S0300-9785(81)80077-4 ↗
2. Differences in implant stability associated with various methods of preparation of the implant bed: an in vitro study10.1016/S0022-3913(12)60092-4 ↗
3. Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis10.1111/j.1600-0501.2006.01265.x ↗
4. Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test10.5051/jpis.2013.43.1.30 ↗
5. Correlation of insertion torques with bone mineral density from dental quantitative CT in the mandible10.1034/j.1600-0501.2003.00932.x ↗
6. Influence of bone volume fraction and architecture on computed large-deformation failure mechanisms in human trabecular bone10.1016/j.bone.2006.06.016 ↗
7. Biocare, Nobel, 2013. Nobel Active Procedures and Products. Göteborg, Sweden.
8. Early loading of nonsubmerged titanium implants with a chemically modified sand-blasted and acid-etched surface: 6-month results of a prospective case series study in the posterior mandible focusing on peri-implant crestal bone changes and implant stabili10.1111/j.1708-8208.2009.00148.x ↗
9. The compressive behavior porous of bone structure as a two-phase10.2106/00004623-197759070-00021 ↗
10. Bone density and its importance in orthodontics10.1016/j.jobcr.2013.01.001 ↗
11. A calibration methodology of QCT BMD for human vertebral body with registered micro-CT images10.1118/1.3582946 ↗
12. DensiProbe spine: a novel instrument for intraoperative measurement of bone density in transpedicular screw fixation10.1097/BRS.0b013e3181bc9470 ↗
13. The effects of superficial roughness and design on the primary stability of dental implants10.1111/j.1708-8208.2009.00202.x ↗
14. Three-dimensional morphometric analysis of human cadaver bone: microstructural data from maxilla and mandible10.1111/j.1600-0501.2004.00969.x ↗
15. Relation between bone density and primary implant stability10.4317/medoral.16.e62 ↗
16. The effect of implant design on insertion torque and immediate micromotion10.1111/j.1600-0501.2010.02142.x ↗
17. Consensus statements and clinical recommendations for implant loading protocols10.11607/jomi.2013.g4 ↗
18. Critical review of immediate implant loading10.1034/j.1600-0501.2003.00950.x ↗
19. Is micro-computed tomography reliable to determine the microstructure of the maxillary alveolar bone?10.1111/j.1600-0501.2012.02478.x ↗
20. The effect of bone compaction on early fixation of porous-coated implants10.1016/S0883-5403(99)90208-5 ↗
21. Surgical guidelines for dental implant placement10.1038/sj.bdj.4813947 ↗
22. Mathematical relationships between bone density and mechanical properties: a literature review10.1016/j.clinbiomech.2007.08.024 ↗
23. The influence of bone volume fraction and ash fraction on bone strength and modulus10.1016/S8756-3282(01)00467-7 ↗
24. Young's modulus, density and material properties in cancellous bone over a large density range10.1007/BF00705371 ↗
25. Validating cone-beam computed tomography for peri-implant bone morphometric analysis10.1038/boneres.2014.10 ↗
26. A novel osseous densification approach in implant osteotomy preparation to increase biomechanical primary stability, bone mineral density, and bone-to-implant contact10.11607/jomi.4817 ↗
27. Clinical and radiological classification of the jawbone anatomy in endosseous dental implant treatment10.5037/jomr.2013.4202 ↗
28. Predicting trabecular bone stiffness from clinical cone-beam CT and HR-pQCT data; an In vitro study using finite element analysis10.1371/journal.pone.0161101 ↗
29. Bone compaction enhances implant fixation in a canine gap model10.1016/j.orthres.2004.11.009 ↗
30. Modeling and experimentation of bone drilling forces10.1016/j.jbiomech.2011.12.012 ↗
31. Bone quality assessment in routine dental implant treatment among Brazilian and Swedish specialists10.1111/clr.12221 ↗
32. Significance of primary stability for osseointegration of dental implants10.1111/j.1600-0501.2005.01201.x ↗
33. Finite element based nonlinear normalization of human lumbar intervertebral disc stiffness to account for its morphology10.1115/1.4027300 ↗
34. Continuum damage interactions between tension and compression in osteonal bone10.1016/j.jmbbm.2015.05.007 ↗
35. Comparison of biting forces in different age and sex groups: a study of biting efficiency with mobile and non-mobile teeth10.1046/j.1365-2842.1999.00364.x ↗
36. Effects on primary stability of three different techniques for implant site preparation in synthetic bone models of different densities10.1016/j.bjoms.2016.07.004 ↗
37. Not only stiffness, but also yield strength of the trabecular structure determined by non-linear µFE is best predicted by bone volume fraction and fabric tensor10.1016/j.jmbbm.2016.10.004 ↗
38. Python for scientific computing python overview10.1109/MCSE.2007.58 ↗
39. A new device for measuring mastication force10.1016/S0940-9602(02)80063-2 ↗
40. Parfitt, A.M., Drezner, M.K., Glorieux, F.H., Kanis, J.A., Malluche, H., Meunier, P.J., Ott, S.M., Recker, R.R., 1987. Bone Histomorphometry : Standardization of Nomenclature, Symbols, and Units 2.10.1002/jbmr.5650020617 ↗
41. Primary implant stability in the atrophic sinus floor of human cadaver maxillae: impact of residual ridge height, bone density, and implant diameter10.1111/clr.12071 ↗
42. R Core Team, 2015. R: A Language and Environment for Statistical Computing.
43. Multi-axial mechanical properties of human trabecular bone10.1007/s10237-008-0128-z ↗
44. Constant mineralization density distribution in cancellous human bone10.1016/S8756-3282(02)00973-0 ↗
45. A generalized anisotropic quadric yield criterion and its application to bone tissue at multiple length scales10.1007/s10237-013-0472-5 ↗
46. Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications10.1111/j.1600-0757.2008.00267.x ↗
47. Mechanical testing of trabecular bone in knee replacement10.1007/BF00271079 ↗
48. Computational analysis of primary implant stability in trabecular bone10.1016/j.jbiomech.2014.12.008 ↗
49. Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature10.1002/(SICI)1097-4636(199822)43:2<192::AID-JBM14>3.0.CO;2-K ↗
50. Immediate loading of threaded implants at stage 1 surgery in edentulous arches: ten consecutive case reports with 1- to 5-year data
51. Biomechanical aspects of primary implant stability: a human cadaver study10.1111/j.1708-8208.2008.00097.x ↗
52. Effect of bone density on the damping behavior of dental implants: an in vitro method10.1016/j.msec.2005.08.019 ↗
53. Damage accumulation in vertebral trabecular bone depends on loading mode and direction10.1016/j.jbiomech.2011.01.018 ↗
54. A 3D damage model for trabecular bone based on fabric tensors10.1016/S0021-9290(96)80006-6 ↗