journal article·2016-01

On the correct interpretation of measured force and calculation of material stress in biaxial tests

D.R. Nolan, J.P. McGarry

Publisher page

Abstract

Biaxial tests are commonly used to investigate the mechanical behaviour of soft biological tissues and polymers. In the current paper we uncover a fundamental problem associated with the calculation of material stress from measured force in standard biaxial tests. In addition to measured forces, localised unmeasured shear forces also occur at the clamps and the inability to quantify such forces has significant implications for the calculation of material stress from simplified force-equilibrium relationships. Unmeasured shear forces are shown to arise due to two distinct competing contributions: (1) negative shear force due to stretching of the orthogonal clamp, and (2) positive shear force as a result of material Poisson-effect. The clamp shear force is highly dependent on the specimen geometry and the clamp displacement ratio, as consequently, is the measured force-stress relationship. Additionally in this study we demonstrate that commonly accepted formulae for the estimation of material stress in the central region of a cruciform specimen are highly inaccurate. A reliable empirical correction factor for the general case of isotropic materials must be a function of specimen geometry and the biaxial clamp displacement ratio. Finally we demonstrate that a correction factor for the general case of non-linear anisotropic materials is not feasible and we suggest the use of inverse finite element analysis as a practical means of interpreting experimental data for such complex materials.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Elasticity and Material Modeling · Mechanical Engineering and Vibrations Research · Orthopaedic implants and arthroplasty

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (24)

Show source-provided references
1. Abaqus/Standard Users Manual, Ver. 6.13-2, 2013. Dassault Systèmes Simulia Corporation, Pawtucket, RI, USA.
2. Biaxial mechanical properties of the natural and glutaraldehyde treated aortic valve cusp. Part I10.1115/1.429624 ↗
3. Biaxial mechanical testing of posterior sclera using high-resolution ultrasound speckle tracking for strain measurements10.1016/j.jbiomech.2013.12.009 ↗
4. Strain uniformity in biaxial specimens is highly sensitive to attachment details10.1115/1.3148467 ↗
5. Directional differences in the biaxial material properties of fascia lata and the implications for fascia function10.1007/s10439-014-0999-3 ↗
6. A new constitutive framework for arterial wall mechanics and a comparative study of material models10.1023/A:1010835316564 ↗
7. Determination of a constitutive relation for passive myocardium10.1115/1.2891193 ↗
8. Biaxial mechanical behavior of excised ventricular epicardium
9. Biaxial tension of fibrous tissue10.1115/1.4023503 ↗
10. Mechanical characterisation of polyurethane elastomer for biomedical applications10.1016/j.jmbbm.2009.03.005 ↗
11. Mixed numerical-experimental technique for orthotropic parameter identification using biaxial tensile tests on cruciform specimens10.1016/j.ijsolstr.2006.06.050 ↗
12. Mechanical characterisation of porcine rectus sheath under uniaxial and biaxial tension10.1016/j.jbiomech.2014.03.009 ↗
13. A robust anisotropic hyperelastic formulation for the modelling of soft tissue10.1016/j.jmbbm.2014.06.016 ↗
14. Nolan, D.N., McGarry, J.P., 2015. On the compressibility of arterial tissue: an experimental and computational investigation. Ann Biomed Eng., 10.1007/s10439-015-1417-110.1007/s10439-015-1417-1 ↗
15. A constitutive description of the anisotropic response of the fascia lata10.1016/j.jmbbm.2013.12.002 ↗
16. Planar biaxial behavior of fibrin-based tissue-engineered heart valve leaflets10.1089/ten.tea.2008.0426 ↗
17. Biaxial mechanical evaluation of planar biological materials
18. Comparison of different biaxial tests for the inverse identification of sheet steel material parameters10.1111/str.12080 ↗
19. Unraveling the effect of boundary conditions and strain monitoring on estimation of the constitutive parameters of elastic membranes by biaxial tests10.1016/j.mechrescom.2014.01.009 ↗
20. Design of a cruciform specimen for biaxial testing of fibre reinforced composite laminates10.1016/j.compscitech.2005.08.011 ↗
21. Effects of boundary conditions on the estimation of the planar biaxial mechanical properties of soft tissues10.1115/1.1933931 ↗
22. Boundary conditions during biaxial testing of planar connective tissues. Part 1
23. Boundary conditions during biaxial testing of planar connective tissues. Part II10.1023/A:1016576603938 ↗
24. Effect of sample geometry on the apparent biaxial mechanical behaviour of planar connective tissues10.1016/j.biomaterials.2005.05.056 ↗