Nitinol stent design – understanding axial buckling
Source abstract
Abstract
Nitinol׳s superelastic properties permit self-expanding stents to be crimped without plastic deformation, but its nonlinear properties can contribute towards stent buckling. This study investigates the axial buckling of a prototype tracheobronchial nitinol stent design during crimping, with the objective of eliminating buckling from the design. To capture the stent buckling mechanism a computational model of a radial force test is simulated, where small geometric defects are introduced to remove symmetry and allow buckling to occur. With the buckling mechanism ascertained, a sensitivity study is carried out to examine the effect that the transitional plateau region of the nitinol loading curve has on stent stability. Results of this analysis are then used to redesign the stent and remove buckling. It is found that the transitional plateau region can have a significant effect on the stability of a stent during crimping, and by reducing the amount of transitional material within the stent hinges during loading the stability of a nitinol stent can be increased.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Shape Memory Alloy Transformations · Ferroelectric and Piezoelectric Materials · Ear Surgery and Otitis Media
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (32)
Show source-provided references
1. Statistical finite element analysis of the buckling behavior of honeycomb structures10.1016/j.compstruct.2013.05.014 ↗
2. Shape-memory alloys: macromodelling and numerical simulations of the superelastic behavior10.1016/S0045-7825(96)01232-7 ↗
3. Deployment of a self-expanding stent inside an artery: a finite element analysis10.1016/j.matdes.2012.05.019 ↗
4. Optimization based simulation of self-expanding nitinol stent10.1016/j.matdes.2013.03.012 ↗
5. An airway wallstent for the treatment of tracheobronchial malignancies10.1136/thx.51.11.1127 ↗
6. Therapeutic bronchoscopy with immediate effect: laser, electrocautery, argon plasma coagulation and stents10.1183/09031936.06.00013906 ↗
7. Numerical analysis of the influence of stent parameters on the fatigue properties10.5539/mas.v6n4p23 ↗
8. Airway stents10.1016/j.athoracsur.2007.11.051 ↗
9. Chung, W.J., Cho, J.W., 1998. On the dynamic effects of explicit FEM in sheet metal forming analysis, vol. 15, pp. 750–776.10.1108/02644409810231880 ↗
10. Finite element comparison of performance related characteristics of balloon expandable stents10.1080/10255840601086234 ↗
11. Performance of fully covered self-expanding metallic stents in benign airway strictures10.1159/000203364 ↗
12. An overview of superelastic stent design10.1080/13645700009169654 ↗
13. Mechanical behaviour modelling of balloon-expandable stents10.1016/S0021-9290(00)00098-1 ↗
14. Treatment of airway complications following lung transplantation10.1016/j.arbres.2010.10.011 ↗
15. Airway stents
16. The role of airway stent placement in the management of tracheobronchial stenosis caused by inoperable advanced lung cancer10.1007/s00595-008-4058-2 ↗
17. Influence of geometrical parameters on radial force during self-expanding stent deployment. Application for a variable radial stiffness stent10.1016/j.jmbbm.2012.02.006 ↗
18. Gong, X., Pelton, A., 2004. Finite element analysis and experimental evaluation of superelastic Nitinol stent. In: SMST 2003 International Conference on Shape Memory and Superelastic Technologies, International Organization on SMST, Pacific Grove, CA, pp. 453–462.
19. Computational mechanics of nitinol stent grafts10.1016/j.jbiomech.2008.05.032 ↗
20. Optimum design of shell structures with random geometric, material and thickness imperfections10.1016/j.ijsolstr.2006.02.019 ↗
21. Medium-term follow-up after deployment of ultraflex expandable metallic stents to manage endobronchial pathology10.1016/j.athoracsur.2004.05.062 ↗
22. Analysis of the mechanical performance of a cardiovascular stent design based on micromechanical modelling10.1016/j.commatsci.2004.05.001 ↗
23. A comparison of different nitinol material data sources for finite element analysis10.1007/s11665-011-9910-7 ↗
24. Self-expanding stent modelling and radial force accuracy
25. Physical metallurgy of Ti–Ni-based shape memory alloys10.1016/j.pmatsci.2004.10.001 ↗
26. Finite element analysis for the design of nitinol medical devices10.3109/13645700009063053 ↗
27. Radial force of coronary stents: a comparative analysis
28. Mechanical fatigue and fracture of Nitinol10.1179/1743280411Y.0000000009 ↗
29. Ryu, Y.J., Kim, H., Yu, C., Choi, J.C., Kwon, Y.S., Kim, J., 2006. Comparison of Natural and Dumon airway stents for the management of benign tracheobronchial stenoses, pp. 748–754. 10.1111/j.1400-1843.2006.00955.x.10.1111/j.1440-1843.2006.00955.x ↗
30. Airway stenting10.1007/s00595-004-2942-y ↗
31. Shen, X., Yi, H., Ni, Z., 2008. Effects of stent design parameters on radial force of stent. In: 2008 2nd International Conference on Bioinformatics and Biomedical Engineering, vol. 1, pp. 1712–1716. 10.1109/ICBBE.2008.756.10.1109/ICBBE.2008.756 ↗
32. Self-expanding nitinol stents: material and design considerations10.1007/s00330-003-2022-5 ↗