Improved fatigue life of acrylic bone cements reinforced with zirconia fibers
Source abstract
Abstract
Poly(methyl methacrylate) (PMMA) bone cements have a long and successful history of use for implant fixation, but suffer from a relatively low fracture and fatigue resistance which can result in failure of the cement and the implant. Fiber or particulate reinforcement has been used to improve mechanical properties, but typically at the expense of the pre-cured cement viscosity, which is critical for successful integration with peri-implant bone tissue. Therefore, the objective of this study was to investigate the effects of zirconia fiber reinforcement on the fatigue life of acrylic bone cements while maintaining a relatively low pre-cured cement viscosity. Sintered straight or variable diameter fibers (VDFs) were added to a PMMA cement and tested in fully reversed uniaxial fatigue until failure. The mean fatigue life of cements reinforced with 15 and 20 vol% straight zirconia fibers was significantly increased by approximately 40-fold, on average, compared to a commercial benchmark (Osteobond) and cements reinforced with 0-10 vol% straight zirconia fibers. The mean fatigue life of a cement reinforced with 10 vol% VDFs was an order of magnitude greater than the same cement reinforced with 10 vol% straight fibers. The time-dependent viscosity of cements reinforced with 10 and 15 vol% straight fibers was comparable to the commercial benchmark during curing. Therefore, the addition of relatively small amounts of straight and variable diameter zirconia fibers was able to substantially improve the fatigue resistance of acrylic bone cement while exhibiting similar handling characteristics compared to current commercial products.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (36)
Show source-provided references
1. ASTM Standard F2118-03, 2003. Standard test method for constant amplitude of force controlled fatigue testing of acrylic bone cement materials. American Society for Testing and Materials, West Conshohocken, PA.
2. The pathology of total joint arthroplasty. II. Mechanisms of implant failure10.1007/s002560050552 ↗
3. Stereological measurement and error estimates for three-dimensional fiber orientation10.1002/pen.760320404 ↗
4. Stiffness and thermal expansion predictions for hybrid short fiber composites10.1002/pc.750110406 ↗
5. Innovative ceramic-fiber technology energizes advanced cerametrics
6. Cass, R.B., Loh, R.R., Allen, T.C., 1998. Method for producing refractory filaments. US Patent No. 5,827,797.
7. Anchorage of the femoral head prosthesis to the shaft of the femur10.1302/0301-620X.42B1.28 ↗
8. Comparison of the mechanical properties of Simplex P, Zimmer Regular, and LVC bone cements10.1002/jbm.820210604 ↗
9. Self-reinforced composite poly(methyl methacrylate): static and fatigue properties10.1016/0142-9612(95)98900-Y ↗
10. Flexural and fatigue properties of a bone cement based upon polyethylmethacrylate and hydroxyapatite10.1007/BF00134320 ↗
11. Tensile characteristics of ten commercial acrylic bone cements10.1002/1097-4636(200009)53:5<605::AID-JBM22>3.0.CO;2-5 ↗
12. Mechanical properties of hydroxyapatite reinforced poly(ethylmethacrylate) bone cement after immersion in a physiological solution: influence of a silane coupling agent10.1023/A:1013057724268 ↗
13. Quantitative analysis of the effect of porosity on the fatigue strength of bone cement10.1016/j.actbio.2008.08.024 ↗
14. Statistical distribution of the fatigue strength of porous bone cement10.1016/j.biomaterials.2009.07.053 ↗
15. Adhesion enhancement of steel fibers to acrylic bone cement through a silane coupling agent10.1002/jbm.a.30543 ↗
16. Improved mechanical properties of acrylic bone cement with short titanium fiber reinforcement10.1007/s10856-006-0616-6 ↗
17. Fracture toughness of steel-fiber-reinforced bone cement10.1002/jbm.a.30107 ↗
18. Reinforcement of bone cement using zirconia fibers with and without acrylic coating10.1002/jbm.a.31783 ↗
19. Evaluation of cement stresses in finite element analyses of cemented orthopaedic implants10.1115/1.1412452 ↗
20. Properties of acrylic bone cement: state of the art review10.1002/(SICI)1097-4636(199722)38:2<155::AID-JBM10>3.0.CO;2-C ↗
21. Fatigue testing and performance of acrylic bone–cement materials: state-of-the-art review10.1002/jbm.b.10018 ↗
22. Alternative acrylic bone cement formulations for cemented arthroplasties: present status, key issues, and future prospects10.1002/jbm.b.30873 ↗
23. Influence of powder-to-liquid monomer ratio on properties of an injectable iodine-containing acrylic bone cement for vertebroplasty and balloon kyphoplasty10.1002/jbm.b.31427 ↗
24. Direct numerical predictions for the elastic and thermoelastic properties of short fibre composites10.1016/S0266-3538(02)00106-9 ↗
25. Axial fatigue strength of carbon fibre reinforced bone cement
26. New processing of apatite fiber and some application for medical use10.1016/j.compscitech.2003.09.006 ↗
27. Which primary total hip replacement?10.1302/0301-620X.77B4.7615593 ↗
28. Carbon fiber-reinforced bone cement in orthopedic surgery10.1002/jbm.820100608 ↗
29. Mechanical properties of poly(methyl methacrylate) bone cements10.1002/jbm.820150208 ↗
30. Direct observation and measurement of fiber architecture in short fiber-polymer composite foam through micro-CT imaging10.1016/j.compscitech.2004.03.003 ↗
31. Microstructural characterization of injection moulded composites using image analysis10.1016/0010-4361(91)90006-3 ↗
32. The effects of centrifugation and titanium fiber reinforcement on fatigue mechanisms in poly(methyl methacrylate) bone cement10.1002/jbm.820290304 ↗
33. Micromechanical model for hydroxyapatite whisker reinforced polymer biocomposites10.1557/jmr.2006.0263 ↗
34. Shape optimization of randomly oriented short fibers for bone cement reinforcements10.1016/j.msea.2004.11.008 ↗
35. Static and fatigue mechanical characterizations of variable diameter fibers reinforced bone cement10.1007/s10856-008-3551-x ↗
36. Bone-shaped short fiber composites–an overview10.1016/S0921-5093(01)01486-1 ↗