Tribological evaluation of biomedical polycarbonate urethanes against articular cartilage
Source abstract
Abstract
This research investigated the in-vitro wear and friction performance of polycarbonate urethane (PCU) 80A as they interact with articular cartilage, using a customised multidirectional pin-on-plate tester. Condyles were articulated against PCU 80A discs (Bionate® I and Bionate® II) (configuration 1) and the results arising from these tests were compared to those recorded during the sliding of PCU pins against cartilage plates (configuration 2). Configuration 1 produced steadily increasing coefficient of friction (COF) (up to 0.64 ± 0.05) and had the same trend as the cartilage-on-stainless steel articulation (positive control). When synovial fluid rather than bovine calf serum was used as lubricant, average COF significantly decreased from 0.50 ± 0.02-0.38 ± 0.06 for condyle-on-Bionate® I (80AI) and from 0.41 ± 0.02-0.24 ± 0.04 for condyle-on-Bionate® II (80AII) test configurations (p < 0.05). After 15 h testing, the cartilage-on-cartilage articulation (negative control) tests showed no cartilage degeneration. However, different levels of cartilage volume loss were found on the condyles from the positive control (12.5 ± 4.2 mm3) and the PCUs (20.1 ± 3.6 mm3 for 80 AI and 19.0 ± 2.3 mm3 for 80AII) (p > 0.05). A good correlation (R2 =0.84) was found between the levels of average COF and the volume of cartilage lost during testing; increasing wear was found at higher levels of COF. Configuration 2 showed low and constant COF values (0.04 ± 0.01), which were closer to the negative control (0.03 ± 0.01) and significantly lower than configuration 1 (p < 0.05). The investigation showed that PCU is a good candidate for use in hemiarthroplasty components, where only one of the two articulating surfaces is replaced, as long as the synthetic material is implanted in a region where migrating cartilage contact is achieved. Bionate® II showed better tribological performance, which suggests it is more favourable for use in hemiarthroplasty design.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Osteoarthritis Treatment and Mechanisms
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (57)
Show source-provided references
1. Experimental and numerical investigation of the behaviour of articular cartilage under shear loading—Interstitial fluid pressurisation and lubrication mechanisms10.1016/j.triboint.2010.09.009 ↗
2. Effects of fiber orientation on the frictional properties and damage of regenerative articular cartilage surfaces10.1089/ten.tea.2012.0580 ↗
3. ASTM(D7490–13), 2013. Standard Test Method for Measurement of the Surface Tension of Solid Coatings, Substrates and Pigments using Contact Angle Measurements. Am Soc Test Mater, PA, USA.
4. The role of interstitial fluid pressurization in articular cartilage lubrication10.1016/j.jbiomech.2009.04.040 ↗
5. Influence of hyaluronic acid on the time-dependent friction response of articular cartilage under different conditions10.1243/095441105X69060 ↗
6. A comparative tribological study of the wear of composite cushion cups in a physiological hip joint simulator10.1016/S0021-9290(98)00034-7 ↗
7. Bionate® Thermoplastic Polycarbonate Polyurethane (PCU) Product Sheet, 2012. DSM Biomedical Inc., NL.
8. Effects of sustained interstitial fluid pressurization under migrating contact area, and boundary lubrication by synovial fluid, on cartilage friction10.1016/j.joca.2008.02.020 ↗
9. Aging performance of a compliant layer bearing acetabular prosthesis in an ovine hip arthroplasty model10.1016/j.arth.2005.07.023 ↗
10. Biodynamic performance of hyaluronic acid versus synovial fluid of the knee in osteoarthritis10.1016/j.ymeth.2015.03.019 ↗
11. Design considerations for cushion form bearings in artificial hip joints10.1243/PIME_PROC_1991_205_269_02 ↗
12. Wear rate evaluation of a novel polycarbonate-urethane cushion form bearing for artificial hip joints10.1016/j.actbio.2010.07.011 ↗
13. Long-term evaluation of a compliant cushion form acetabular bearing for hip joint replacement: a 20 million cycles wear simulation10.1002/jor.21471 ↗
14. The influence of continuous sliding and subsequent surface wear on the friction of articular cartilage10.1243/0954411991535167 ↗
15. Polyethylene particles of a ‘critical size’ are necessary for the induction of cytokines by macrophages in vitro10.1016/S0142-9612(98)00140-9 ↗
16. Patellofemoral contact pressures. The influence of q-angle and tendofemoral contact10.2106/00004623-198466050-00010 ↗
17. In vivo deformation, surface damage, and biostability of retrieved Dynesys systems10.1097/BRS.0b013e3181d6f84f ↗
18. ISO(14243-4), 2013. Implants for Surgery — Wear of Total Knee Prostheses Part 4: Wear of the Patella-femoral Joint — Loading and Displacement Parameters for Wear-testing Machines and Corresponding Environmental Conditions for Test.
19. Tribological properties of PVA/PVP blend hydrogels against articular cartilage10.1016/j.jmbbm.2017.10.027 ↗
20. Effect of load variation on the friction properties of articular cartilage
21. Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part I: materials selection and evaluation10.1016/j.biomaterials.2004.02.065 ↗
22. Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part II: in vivo investigation10.1016/j.biomaterials.2004.02.064 ↗
23. Accelerated aging, natural aging, and small punch testing of gamma-air sterilized polycarbonate urethane acetabular components10.1002/jbm.b.31601 ↗
24. Lizhang · 2010
25. The effect of contact stress on cartilage friction, deformation and wear10.1177/2041303310392626 ↗
26. Effect of clearance on cartilage tribology in hip hemi-arthroplasty10.1177/0954411913502156 ↗
27. Polyurethane as a potential knee hemiarthroplasty biomaterial: an in-vitro simulation of its tribological performance10.1243/09544119JEIM657 ↗
28. Nanoporous substrate-infiltrated hydrogels: a bioinspired regenerable surface for high load bearing and tunable friction10.1002/adfm.201503681 ↗
29. Both hyaluronan and collagen type II keep proteoglycan 4 (lubricin) at the cartilage surface in a condition that provides low friction during boundary lubrication10.1021/la504345c ↗
30. Role of hydrophobicity on the adsorption of synovial fluid proteins and biolubrication of polycarbonate urethanes: materials for permanent meniscus implants10.1016/j.matdes.2015.06.075 ↗
31. Finite element analysis of the meniscectomised tibio-femoral joint: implementation of advanced articular cartilage models10.1080/10255842.2012.758253 ↗
32. Tribological testing of articular cartilage of the medial compartment of the knee using a friction simulator10.1016/j.triboint.2008.03.012 ↗
33. An investigation of the effect of conformity of knee hemiarthroplasty designs on contact stress, friction and degeneration of articular cartilage: a tribological study10.1016/j.jbiomech.2009.03.028 ↗
34. Hydrophilic polyurethane elastomers for hemiarthroplasty: a preliminary in vitro wear study
35. Mow · 1997
36. Low-friction hydrophilic surface for medical devices10.1016/0142-9612(90)90098-B ↗
37. In vivo compatibility of Dynesys((R)) spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants10.1007/s00586-014-3705-0 ↗
38. Determination of the mechanical properties of Bionate 80A and Bionate 75D for the stress analysis of cushion form bearings10.1243/09544119JEIM372 ↗
39. Investigation of the friction and surface degradation of innovative chondroplasty materials against articular cartilage10.1243/09544119JEIM178 ↗
40. Wear and damage of articular cartilage with friction against orthopedic implant materials10.1016/j.jbiomech.2015.04.008 ↗
41. Compliant layer acetabular cups: friction testing of a range of materials and designs for a new generation of prosthesis that mimics the natural joint10.1243/09544119H06404 ↗
42. Design aspects of compliant, soft layer bearings for an experimental hip prosthesis10.1243/095441105X9318 ↗
43. Polyurethane unicondylar knee prostheses: simulator wear tests and lubrication studies10.1088/0031-9155/52/1/013 ↗
44. Shi · 2016
45. Retrieval analysis of a polycarbonate-urethane acetabular cup: a case report10.1615/JLongTermEffMedImplants.v18.i1.570 ↗
46. In vitro macrophage response to polyethylene and polycarbonate-urethane particles10.1002/jbm.a.32529 ↗
47. In vivo response to cross-linked polyethylene and polycarbonate-urethane particles10.1002/jbm.a.32531 ↗
48. A tribological study of UHMWPE acetabular cups and polyurethane compliant layer acetabular cups10.1002/1097-4636(2000)53:6<710::AID-JBM14>3.0.CO;2-R ↗
49. Evaluation of the wear performance of a polycarbonate-urethane acetabular component in a hip joint simulator and comparison with UHMWPE and cross-linked UHMWPE10.1177/0885328210394471 ↗
50. Mechanical properties of a novel PVA hydrogel in shear and unconfined compression10.1016/S0142-9612(00)00242-8 ↗
51. Combinatorial scaffold morphologies for zonal articular cartilage engineering10.1016/j.actbio.2013.12.030 ↗
52. Polyurethane elastomer biostability10.1177/088532829500900402 ↗
53. A review of protein adsorption on bioceramics10.1098/rsfs.2012.0012 ↗
54. Ward · 2011
55. Explantation and analysis of the first retrieved human acetabular cup made of polycarbonate urethane: a case report10.1615/JLongTermEffMedImplants.v18.i1.580 ↗
56. Zeng · 2013
57. Chondroprotective effects of a polycarbonate-urethane meniscal implant: histopathological results in a sheep model10.1007/s00167-010-1210-5 ↗