Effect of two-year degradation on mechanical interaction between a bioresorbable scaffold and blood vessel
Source abstract
Abstract
This paper aims to evaluate the mechanical behaviour of a bioresorbable polymeric coronary scaffold using finite element method, focusing on scaffold-artery interaction during degradation and vessel remodelling. A series of nonlinear stress-strain responses was constructed to match the experimental measurement of radial stiffness and strength for polymeric scaffolds over 2-year in-vitro degradation times. Degradation process was modelled by incorporating the change of material property as a function of time. Vessel remodelling was realised by changing the size of artery-plaque system manually, according to the clinical data in literature. Over degradation times, stress on the scaffold tended to increase firstly and then decreased gradually, corresponding to the changing yield stress of the scaffold material; whereas the stress on the plaque and arterial layers showed a continuous decrease. In addition, stress reduction was also observed for scaffold, plaque and artery in the simulations with the consideration of vessel remodelling. For the first time, the work offered insights into mechanical interaction between a bioresorbable scaffold and blood vessel during two-year in-vitro degradation, which has significance in assisting with further development of bioresorbable implants for treating cardiovascular diseases.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Coronary Interventions and Diagnostics · Elasticity and Material Modeling · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (42)
Show source-provided references
1. Fatigue life prediction of cardiovascular stent using finite element method10.1080/10255842.2012.713675 ↗
2. In-stent stenosis: pathology and implications for the development of drug eluting stents10.1136/heart.89.2.218 ↗
3. Mathematical modeling of degradation for bulk-erosive polymers: applications in tissue engineering scaffolds and drug delivery systems10.1016/j.actbio.2010.09.038 ↗
4. Restenosis delineating the numerous causes of drug-eluting stent restenosis
5. Mechanisms of post-intervention arterial remodelling10.1093/cvr/cvs276 ↗
6. Mechanical properties of a biodegradable balloon-expandable stent from poly (L-lactide) for peripheral vascular applications10.1115/1.2355683 ↗
7. On the effects of different strategies in modelling balloon-expandable stenting by means of finite element method10.1016/j.jbiomech.2008.01.027 ↗
8. In vitro and in vivo degradability and cytocompatibility of poly (l-lactic acid) scaffold fabricated by a gelatin particle leaching method10.1016/j.actbio.2006.12.008 ↗
9. Determination of layer-specific mechanical properties of human coronary arteries with nonatherosclerotic intimal thickening and related constitutive modeling10.1152/ajpheart.00934.2004 ↗
10. Analysis of the stent expansion in a stenosed artery using finite element method: application to stent versus stent study10.1177/0954411914556788 ↗
11. On the finite element modelling of balloon-expandable stents10.1016/j.jmbbm.2007.07.002 ↗
12. Cardiovascular stent design and vessel stresses: a finite element analysis10.1016/j.jbiomech.2004.07.022 ↗
13. Lumen gain and restoration of pulsatility after implantation of a bioresorbable vascular scaffold in porcine coronary arteries
14. Fatigue life analysis and experimental verification of coronary stent10.1007/s00380-009-1203-9 ↗
15. Tailoring crystallization: towards High‐Performance Poly (lactic acid)10.1002/adma.201305413 ↗
16. Static circumferential tangential modulus of human atherosclerotic tissue10.1016/0021-9290(94)90209-7 ↗
17. Degradation model of bioabsorbable cardiovascular stents10.1371/journal.pone.0110278 ↗
18. Degradability of polymers for implantable biomedical devices10.3390/ijms10094033 ↗
19. Computational analysis of the radial mechanical performance of PLLA coronary artery stents10.1016/j.medengphy.2014.09.014 ↗
20. Large deformation isotropic elasticity-on the correlation of theory and experiment for incompressible rubberlike solids10.1098/rspa.1972.0026 ↗
21. First serial assessment at 6 months and 2 years of the second generation of absorb everolimus-eluting bioresorbable vascular scaffold
22. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial10.1016/S0140-6736(08)60415-8 ↗
23. Effect of molecular weight and crystallinity on poly (lactic acid) mechanical properties10.1002/(SICI)1097-4628(19960103)59:1<37::AID-APP6>3.0.CO;2-N ↗
24. Modeling of degradation and drug release from a biodegradable stent coating10.1002/jbm.a.31053 ↗
25. Qiu · 2017
26. Manufacture and characterisation of porous PLA Scaffolds10.1016/j.procir.2015.07.025 ↗
27. Endovascular stent design dictates experimental restenosis and thrombosis10.1161/01.CIR.91.12.2995 ↗
28. Computational analysis of mechanical stress–strain interaction of a bioresorbable scaffold with blood vessel10.1016/j.jbiomech.2016.05.035 ↗
29. Crimping and deployment of metallic and polymeric stents-finite element modelling
30. Dynamics of vessel wall changes following the implantation of the Absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study at 6, 12, 24 and 36 months10.4244/EIJV9I11A217 ↗
31. Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes10.1016/j.jacc.2011.05.050 ↗
32. Evaluation of the second generation of a bioresorbable everolimus drug-eluting vascular scaffold for treatment of De Novo coronary artery stenosisclinical perspective10.1161/CIRCULATIONAHA.110.970772 ↗
33. A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods10.1016/S0140-6736(09)60325-1 ↗
34. A polylactide bioresorbable scaffold eluting everolimus for treatment of coronary stenosis10.1016/j.jacc.2015.11.060 ↗
35. Constitutive framework for biodegradable polymers with applications to biodegradable stents10.1097/MAT.0b013e31816ba55a ↗
36. Modeling of deformation-accelerated breakdown of polylactic acid biodegradable stents10.1115/1.4002759 ↗
37. Effect of endovascular stent strut geometry on vascular injury, myointimal hyperplasia, and restenosis10.1067/mva.2002.122878 ↗
38. Mechanical study of PLA–PCL fibers during in vitro degradation10.1016/j.jmbbm.2010.12.006 ↗
39. Shear-stress and wall-stress regulation of vascular remodeling after balloon angioplasty effect of matrix metalloproteinase inhibition10.1161/01.CIR.104.1.91 ↗
40. Determination of the influence of stent strut thickness using the finite element method: implications for vascular injury and in-stent restenosis10.1007/s11517-009-0432-5 ↗
41. The biocompatibility of rapidly degrading polymeric stents in porcine carotid arteries10.1016/j.biomaterials.2010.06.057 ↗
42. Modeling and analysis of drug-elu dfRkting stents with biodegradable PLGA coating: consequences on intravascular drug delivery10.1115/1.4028135 ↗