Mechanical degradation of biological heart valve tissue induced by low diameter crimping: An early assessment
Source abstract
Abstract
Transcatheter aortic valve implantation (TAVI) has become today an increasingly attractive procedure to relieve patients from aortic valve disease. However, the procedure requires crimping biological tissue within a metallic stent for low diameter catheter insertion purpose. This step induces specific stress in the leaflets especially when the crimping diameter is small. One concern about crimping is the potential degradations undergone by the biological tissue, which may limit the durability of the valve once implanted. The purpose of the present work is to study the effect of low diameter crimping on the mechanical performances of pericardium valve prototypes. The prototypes were compressed to a diameter of 1mm within braided stents for 20 min. SEM observations performed on crimped material show that crimped leaflets undergo degradations characterized by apparent surface defects. Moreover mechanical extension tests were performed on pericardium strips before and after crimping. The strips (15 mm long, 5mm wide) were taken from both crimped and native leaflets considering 2 different valve diameters, 19 and 21 mm. In order to prevent the premature drying of the pericardium tissue during the procedure, the biological tissue was kept in contact with a formaldehyde solution. Results show that the ultimate strength value decreases nearly by up to 50%. The modifications observed in the material may jeopardize the long term durability of the device. However, further tests are necessary with a larger amount of samples to confirm these early results.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Cardiac Valve Diseases and Treatments · Infective Endocarditis Diagnosis and Management · Cardiac Structural Anomalies and Repair
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (11)
Show source-provided references
1. The effects of transcatheter valve crimping on pericardial leaflets10.1016/j.athoracsur.2013.11.009 ↗
2. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis, first human case description10.1161/01.CIR.0000047200.36165.B8 ↗
3. Percutaneous therapies for valvular heart disease10.1016/j.carpath.2006.02.004 ↗
4. Khoffi F, Heim F, Chakfe N, Lee JT, 2014. Transcatheter fiber heart valve: effect of crimping on material performances. J. Biomed. Mater. Res. B. DOI: 10.1002/jbm.b.33330 (epub ahead of print)10.1002/jbm.b.33330 ↗
5. Crimping may affect the durability of transcatheter valves: an experimental analysis10.1016/j.athoracsur.2011.03.020 ↗
6. Percutaneous pulmonary valve replacement: autologous tissue-engineered valved stents10.1093/cvr/cvq212 ↗
7. Long-term outcomes after transcatheter aortic valve implantation in high-risk patients with severe aortic stenosis: the U.K. TAVI (United Kingdom Transcatheter Aortic Valve Implantation) registry10.1016/j.jacc.2011.08.050 ↗
8. Long-term outcomes after transcatheter aortic valve implantation10.1016/j.jacc.2012.08.960 ↗
9. The effect of rolling and crimping on cell-seeded synthetic scaffolds for the development of transcatheter valves by tissue engineering
10. 5-Year outcome after transcatheter aortic valve implantation10.1016/j.jacc.2012.11.010 ↗
11. Is it reasonable to treat all calcified stenotic aortic valves with a valved stent10.1016/j.jacc.2007.10.023 ↗