Abstract

The present study has two major purposes: firstly, to investigate whether the analytical model proposed by Jedwab and Clerc for assessing the mechanical behaviour of an open ends metallic braided stent is applicable to the looped ends stent design and secondly, to compare the response of the two stent designs subjected to radial compression. We use finite element analysis to evaluate the performance of the two braided stents emulating well established designs: WALLSTENT and WallFlex. We validate the WALLSTENT model analytically. We perform a radial crimping simulation and evaluate the radial forces and stresses induced. This study confirms the validity of using the analytical model in the biomechanical analysis of the WALLSTENT design. However, in case of the WallFlex design, a major difference in the results can be observed in the levels of radial forces and wire peak stresses, justifying the decision of using a different alloy in the fabrication of the WallFlex design.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Shape Memory Alloy Transformations · Elasticity and Material Modeling · Mechanical stress and fatigue analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (37)

Show source-provided references
1. Adanur · 1995
2. Braiding · 1987 · 519
3. Assessment of wrought ASTM f1058 cobalt alloy properties for permanent surgical implants10.1002/(SICI)1097-4636(199723)38:3<229::AID-JBM7>3.0.CO;2-R ↗
4. Conti · 2007
5. Corbman · 1983
6. De Beule · 2008
7. De Beule, M., Mortier, P., De Santis, G., Conti, M., Segers, P., Verdonck, P., Verhegghe, B., 2009a. A virtual product development strategy for minimally invasive medical devices. In: Proceedings of the 2009 SIMULIA Cust. Conference, pp. 19–27.
8. Virtual optimization of self-expandable braided wire stents10.1016/j.medengphy.2008.11.008 ↗
9. De Beule, M., Van Impe, R., Verdonck, P., Verhegghe B., 2007. Finite element stent design. In: Proceedings of the Symp. FEA U Gent BE, pp. 4–5.
10. Metallic surface treatment using electrochemical polishing decreases thrombogenicity and neointimal hyperplasia after coronary stent implantation in a porcine model10.1016/S0735-1097(98)81848-2 ↗
11. Yielding criteria for ductile metals
12. Transluminal treatment of arteriosclerotic obstruction: description of a new technic and a preliminary report of its application10.1161/01.CIR.30.5.654 ↗
13. Transluminal expandable nitinol coil stent grafting: preliminary report10.1148/radiology.147.1.6828741 ↗
14. Simulations of self-expanding braided stent using macroscopic model of NiTi shape memory alloys covering R-phase10.1007/s11665-014-0966-z ↗
15. A study of the geometrical and mechanical properties of self-expanding metallic stent-theory and experiment10.1002/jab.770040111 ↗
16. Mechanical modeling of self-expandable stent fabricated using braiding technology10.1016/j.jbiomech.2008.08.005 ↗
17. Simulation of mechanical behavior of temperature-responsive braided stents made of shape memory polyurethanes10.1016/j.jbiomech.2009.10.032 ↗
18. MatWeb, 2017. Elgiloy® Co-Cr-Ni Alloy, Wire, 40% Cold Reduction, Heat Treated. 〈http://www.matweb.com/search/datasheet.aspx?Matguid=8b7b228cb341407a834c46743e843ca3&n=1〉. (Accessed 1 February 2017).
19. Numerical investigations of the mechanical properties of a braided non-vascular stent design using finite element method
20. Peirlinck, M., 2013. Design of biodegradable esophageal stents (PhD Thesis), University of Ghent, Ghent, Belgium.
21. Poncet, P.P., 2000. Nitinol Medical Device Design Considerations 1–12. 〈http://memry.com/sites/default/files/documents/Nitinol_Medical_Device_Considerations_SMST00.pdf〉. (Accessed 2 November 2015).
22. How a dentist's name became a synonym for a life-saving device: the story of Dr. Charles Stent
23. Self-expanding endovascular prosthesis: an experimental study10.1148/radiology.164.3.3303120 ↗
24. A view of vascular stents10.1161/01.CIR.79.2.445 ↗
25. Thermomechanical model for NiTi-based shape memory alloys including R-phase and material anisotropy under multi-axial loadings10.1016/j.ijplas.2012.06.008 ↗
26. A survey of stent designs10.1080/136457002760273340 ↗
27. Self-expanding nitinol stents: material and design considerations10.1007/s00330-003-2022-5 ↗
28. Verhegghe, B., 2013. pyFormex Documentation.
29. Wahl · 1944
30. Prosthesis comprising an expansible or contractile tubular body
31. Wallsten, H.I., Imbert, C., 1991. Self-expanding Prosthesis, U.S. Pat. 5,061, 275.
32. Mechanical response of a metallic aortic stent – Part I: pressure-diameter relationship10.1115/1.1782650 ↗
33. Fabrication and evaluation of polymer-based esophageal stents for benign esophagus stricture insertion10.1039/C5RA23763G ↗
34. Models of stents – comparison and applications10.33549/physiolres.931309 ↗
35. Structural analysis of a two-dimensional braided fabric10.1080/00405009708658528 ↗
36. Mechanical performance of cup-spherical-shaped and straight form braided esophageal stent10.11591/telkomnika.v11i11.3512 ↗
37. The impact of wire stent fabrication technique on the performance of stent placement10.1115/1.4005788 ↗