Biomechanical comparison of locking and non-locking patient-specific mandibular reconstruction plate using finite element analysis
Source abstract
Abstract
Patient-specific mandibular reconstruction plate (PSMRP), as one of the patient-specific implants (PSIs), offers a host of benefits to mandibular reconstruction. Due to the limitation of fabricating screw hole threads in the PSMRP, 3D printed PSMRP is applied to the non-locking system directly in the mandibular reconstruction with bone graft regardless of the locking system. Since the conventional manual-bending reconstruction plate (CMBRP) provides better fixation in the locking system, it needs to be validated whether the locking PSMRP performs better than the non-locking PSMRP in the patient-specific mandibular reconstruction. Thereupon, the purpose of this study was to compare the biomechanical behavior between the locking and non-locking PSMRP. Finite element analysis (FEA) was used to conduct the biomechanical comparison between the locking PSMRP and non-locking PSMRP by simulating the momentary incisal clenching through static structural analysis. Mandible was reconstructed through the virtual surgical planning, and subsequently a 3D model of mandibular reconstruction assembly, including reconstructed mandible, PSMRP, and fixation screws, was generated and meshed for the following FEA simulations. In the form of equivalent von Mises stress, equivalent elastic strain, and total deformation, the locking PSMRP demonstrated its higher strengths of preferable safety, desirable flexibility, and anticipated stability compared with the non-locking PSMRP, indicated by much lower maximum stress, lower maximum strain and equivalent displacement. Locking PSMRP/screw system provides a better fixation effect to the patient-specific mandibular reconstruction than the non-locking one as a result of its productive fixation nature. FEA plays a paramount role in pre-validating the design of PSMRP through the biomechanical behavior evaluation in static structural analysis.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Facial Trauma and Fracture Management · Reconstructive Surgery and Microvascular Techniques · Bone fractures and treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (43)
Show source-provided references
1. Al-Ahmari · 2015 · 1
2. A comparative study on the customized design of mandibular reconstruction plates using finite element method10.1177/1687814015593890 ↗
3. Patient-specific implants for maxillofacial defects: challenges and solutions10.1186/s40902-020-00262-7 ↗
4. Finite element analysis of three patterns of internal fixation of fractures of the mandibular condyle10.1016/j.bjoms.2012.08.007 ↗
5. Biomechanical evaluation of different plating methods used in mandibular angle fractures with 3-dimensional finite element analysis: favorable fractures10.1016/j.joms.2017.02.028 ↗
6. Reconstruction plate-related complications in mandibular continuity defects10.1007/s10006-019-00762-5 ↗
7. Low- and high-cycle fatigue resistance of Ti-6Al-4V ELI additively manufactured via selective laser melting: mean stress and defect sensitivity10.1016/j.ijfatigue.2017.10.021 ↗
8. Finite element analysis of the human mandible at 3 different stages of life10.1016/j.tripleo.2010.01.025 ↗
9. The biomechanical aspects of reconstruction for segmental defects of the mandible: a finite element study to assess the optimisation of plate and screw factors10.1016/j.jcms.2013.12.005 ↗
10. Biomechanical evaluation of the screw preload values used in the plate placement for bone fractures10.1177/0954411920964628 ↗
11. Mandibular reconstruction and second generation locking reconstruction plates: outcome of 110 patients10.1016/j.ijom.2009.03.721 ↗
12. Flexural and torsional properties of a glass fiber-reinforced composite diaphyseal bone model with multidirectional fiber orientation10.1016/j.jmbbm.2018.07.027 ↗
13. Use of a 2.0-mm locking plate/screw system for mandibular fracture surgery10.1053/joms.2002.33110 ↗
14. Customized mandibular reconstruction plates improve mechanical performance in a mandibular reconstruction model10.1080/10255842.2016.1240788 ↗
15. Patient-specific orthopaedic implants10.1111/os.12282 ↗
16. Does plate adaptation affect stabilityA biomechanical comparison of locking and nonlocking plates10.1053/joms.2002.35732 ↗
17. Adaquate fixation of plates for stability during mandibular reconstruction10.1016/j.jcms.2006.01.003 ↗
18. Deformation of the human mandible during simulated tooth clenching10.1177/00220345940730010801 ↗
19. Kristina · 2009 · 169
20. Lecci · Intercondylar distance before and after mandible resection · 2013 · 5 · 9
21. Critical assessment of the fatigue performance of additively manufactured Ti–6Al–4V and perspective for future research10.1016/j.ijfatigue.2015.12.003 ↗
22. Biostable glass fibre-reinforced dimethacrylate-based composites as potential candidates for fracture fixation plates in toy-breed dogs
23. Application of finite element analysis in oral and maxillofacial surgery-A literature review10.3390/ma13143063 ↗
24. An investigation of two finite element modeling solutions for biomechanical simulation using a case study of a mandibular bone10.1115/1.4037633 ↗
25. Biomechanical optimization of bone plates used in rigid fixation of mandibular fractures10.1016/j.joms.2008.12.032 ↗
26. Evaluation of the therapeutic effect of Bi-maxillary osteotomy using the stress distribution on the temporomandibular joint when doing anterior teeth occlusion10.1115/1.4047425 ↗
27. Patient specific implants: scope for the future10.1007/s12178-014-9214-2 ↗
28. Computer assisted design and analysis of customized porous plate for mandibular reconstruction10.1016/j.irbm.2017.01.003 ↗
29. Mechanical properties and in vivo performance of load-bearing fiber-reinforced composite intramedullary nails with improved torsional strength10.1016/j.jmbbm.2014.08.020 ↗
30. Finite element analysis of customized reconstruction plates for mandibular continuity defect therapy10.1016/j.jbiomech.2013.11.016 ↗
31. Nicolas Homsi · 2020
32. Complication of osteo reconstruction by utilizing free vascularized fibular bone graft10.1186/s12893-020-00875-9 ↗
33. The feasibility of a custom-made endoprosthesis in mandibular reconstruction: implant design and finite element analysis10.1016/j.jcms.2015.10.004 ↗
34. Patient specific glass fiber reinforced composite versus titanium plate: a comparative biomechanical analysis under cyclic dynamic loading10.1016/j.jmbbm.2018.12.014 ↗
35. Failure analysis of an in-vivo fractured patient-specific Ti6Al4V mandible reconstruction plate fabricated by selective laser melting10.1016/j.engfailanal.2021.105353 ↗
36. Fracture of mandibular reconstruction plates used after tumor resection10.1053/joms.2002.29817 ↗
37. Standard 3D titanium miniplate versus locking 3D miniplate in fracture of mandible: a prospective comparative study10.1007/s12663-015-0817-y ↗
38. Resistance and stress finite element analysis of different types of fixation for mandibular orthognathic surgery10.1590/0103-6440201600336 ↗
39. Szypryt · Trauma, The use and abuse of locking plates · 2009 · 23 · 281
40. Biomechanics of mandibular reconstruction: a review10.1016/j.ijom.2009.11.003 ↗
41. R.C.W. Wong, J.S.P. Loh, I. Islam, The Role of Finite Element Analysis in Studying Potential Failure of Mandibular Reconstruction Methods, Perusal of the Finite Element Method2016.10.5772/64890 ↗
42. Patient-specific mandibular reconstruction plates increase accuracy and long-term stability in immediate alloplastic reconstruction of segmental mandibular defects10.1007/s12663-019-01323-9 ↗
43. Zezo · 2019