In vitro validation of a novel mechanical model for testing the anchorage capacity of pedicle screws using physiological load application

Christian Liebsch, Julia Zimmermann, Nicolas Graf, Christoph Schilling, Hans-Joachim Wilke, Annette Kienle

Publisher page

Abstract

Biomechanical in vitro tests analysing screw loosening often include high standard deviations caused by high variabilities in bone mineral density and pedicle geometry, whereas standardized mechanical models made of PU foam often do not integrate anatomical or physiological boundary conditions. The purpose of this study was to develop a most realistic mechanical model for the standardized and reproducible testing of pedicle screws regarding the resistance against screw loosening and the holding force as well as to validate this model by in vitro experiments. The novel mechanical testing model represents all anatomical structures of a human vertebra and is consisting of PU foam to simulate cancellous bone, as well as a novel pedicle model made of short carbon fibre filled epoxy. Six monoaxial cannulated pedicle screws (Ø6.5 × 45mm) were tested using the mechanical testing model as well as human vertebra specimens by applying complex physiological cyclic loading (shear, tension, and bending; 5Hz testing frequency; sinusoidal pulsating forces) in a dynamic materials testing machine with stepwise increasing load after each 50.000 cycles (100.0N shear force + 20.0N per step, 51.0N tension force + 10.2N per step, 4.2Nm bending moment + 0.8Nm per step) until screw loosening was detected. The pedicle screw head was fixed on a firmly clamped rod while the load was applied in the vertebral body. For the in vitro experiments, six human lumbar vertebrae (L1-3, BMD 75.4 ± 4.0mg/cc HA, pedicle width 9.8 ± 0.6mm) were tested after implanting pedicle screws under X-ray control. Relative motions of pedicle screw, specimen fixture, and rod fixture were detected using an optical motion tracking system. Translational motions of the mechanical testing model experiments in the point of load introduction (0.9-2.2mm at 240N shear force) were reproducible within the variation range of the in vitro experiments (0.6-3.5mm at 240N shear force). Screw loosening occurred continuously in each case between 140N and 280N, while abrupt failures of the specimen were observed only in vitro. In the mechanical testing model, no translational motion was detected in the screw entry point, while in vitro, translational motions of up to 2.5mm in inferior direction were found, leading to a slight shift of the centre of rotation towards the screw tip. Translational motions of the screw tip of about 5mm in superior direction were observed both in vitro and in the mechanical testing model, while they were continuous in the mechanical testing model and rapidly increasing after screw loosening initiation in vitro. The overall pedicle screw loosening characteristics were qualitatively and quantitatively similar between the mechanical testing model and the human vertebral specimens as long as there was no translation of the screw at the screw entrance point. Therefore, the novel mechanical testing model represents a promising method for the standardized testing of pedicle screws regarding screw loosening for cases where the screw rotates around a point close to the screw entry point.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Spinal Fractures and Fixation Techniques · Spine and Intervertebral Disc Pathology · Medical Imaging and Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (57)

Show source-provided references
1. ASTM F1717-15 · 2015
2. ASTM F1798-13 · 2013
3. ASTM F2077-14 · 2014
4. ASTM F543-13e1 · 2013
5. A morphometric study of human lumbar and selected thoracic vertebrae10.1097/00007632-198705000-00010 ↗
6. Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines10.1016/j.spinee.2013.03.010 ↗
7. Polymethylmethacrylate augmentation of cannulated pedicle screws for fixation in osteoporotic spines and comparison of its clinical results and biomechanical characteristics with the needle injection method10.1097/BSD.0b013e318246ae8a ↗
8. Increasing bending strength and pullout strength in conical pedicle screws: biomechanical tests and finite element analyses
9. Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws10.1016/j.medengphy.2014.05.003 ↗
10. Biomechanical analysis of pedicle screws in osteoporotic bone with bioactive cement augmentation using simulated in vivo multicomponent loading10.1097/BRS.0b013e3181d449ec ↗
11. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients10.6061/clinics/2015(02)08 ↗
12. Morphometric study of lumbar vertebrae's pedicle10.1590/S1413-78522007000400001 ↗
13. Instrumentation-related complications of multilevel fusions for adult spinal deformity patients over age 65: surgical considerations and treatment options in patients with poor bone quality10.1097/01.brs.0000236893.65878.39 ↗
14. The" Fixatuer Interne" as a versatile implant for spine surgery10.1097/00007632-198711000-00009 ↗
15. The mechanical properties of bone in osteoporosis10.1302/0301-620X.63B2.7217148 ↗
16. Pedicle screw loosening: a clinically relevant complication?10.1007/s00586-015-3768-6 ↗
17. Surgical treatment of degenerative and traumatic spinal diseases with expandable screws in patients with osteoporosis: 2-year follow-up clinical study10.3171/2016.3.SPINE151294 ↗
18. A 10-year follow-up evaluation of lumbar spine fusion With pedicle screw fixation10.1097/01.BRS.0000067112.15753.AD ↗
19. The shear strength of trabecular bone from the femur, and some factors affecting the shear strength of the cement-bone interface10.1007/BF00381636 ↗
20. Increase of pullout strength of spinal pedicle screws with conical core: biomechanical tests and finite element analyses10.1016/j.orthres.2004.11.002 ↗
21. Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine10.1016/j.spinee.2017.01.009 ↗
22. Differences between the tensile and compressive strengths of bovine tibial trabecular bone depend on modulus10.1016/0021-9290(94)90054-X ↗
23. Comparative analysis of effect of density, insertion angle and reinsertion on pull-out strength of single and two pedicle screw constructs using synthetic bone model10.4184/asj.2016.10.3.414 ↗
24. Influence of the screw augmentation technique and a diameter increase on pedicle screw fixation in the osteoporotic spine: pullout versus fatigue testing10.1007/s00586-014-3476-7 ↗
25. Caudo-Cephalad loading of pedicle screws: mechanisms of loosening and methods of augmentaion10.1097/00007632-199312000-00012 ↗
26. Complications associated with thoracic pedicle screws in spinal deformity10.1007/s00586-010-1316-y ↗
27. Analysis of anatomic morphometry of the pedicles and the safe zone for through-pedicle procedures in the thoracic and lumbar spine10.1007/s00586-006-0245-2 ↗
28. Fusion of the lumbar and sacral spine by internal fixation with screw plates10.1097/00003086-198602000-00004 ↗
29. Complications after transpedicular stabilization of the spine: a survivorship analysis of 163 cases10.1097/00007632-199412150-00007 ↗
30. Comparison of teriparatide and bisphosphonate treatment to reduce pedicle screw loosening after lumbar spinal fusion surgery in postmenopausal women with osteoporosis from a bone quality perspective10.1097/BRS.0b013e31828826dd ↗
31. Posterior lumbar interbody fusion: a retrospective study of complications after facet joint excision and pedicle screw fixation in 148 cases10.3109/17453679908997819 ↗
32. Can insertional torque predict screw loosening and related failures? An in vivo study of pedicle screw fixation augmenting posterior lumbar interbody fusion10.1097/00007632-200004010-00015 ↗
33. Human lumbar vertebrae: quantitative three-dimensional anatomy10.1097/00007632-199203000-00010 ↗
34. Complications of transpedicular lumbosacral fixation for non-traumatic disorders10.1302/0301-620X.79B2.0790183 ↗
35. The elastic modulus for bone10.1016/0021-9290(74)90018-9 ↗
36. A spinal fixation device for in vivo load measurement10.1016/0021-9290(94)90268-2 ↗
37. Monitoring in vivo implant loads with a telemeterized internal spinal fixation device10.1097/00007632-200012010-00004 ↗
38. Internal fixation of the lumbar spine with pedicle screw plating10.1097/00003086-198602000-00003 ↗
39. Longitudinal shear properties of human compact bone and its constituents, and the associated failure mechanisms10.1007/BF00566240 ↗
40. The significance of radiolucent zones surrounding pedicle screws10.1302/0301-620X.86B3.14323 ↗
41. Insertion torque is not a good predictor of pedicle screw loosening after spinal instrumentation: a prospective study in 8 patients10.1186/1754-9493-4-14 ↗
42. Biomechanical evaluation of a posterior non-fusion instrumentation of the lumbar spine10.1007/s00586-011-2121-y ↗
43. Vertebral body and posterior element morphology: the normal spine in middle life10.1097/00007632-198810000-00002 ↗
44. Segmental spine plates with pedicle screw fixation: a new internal fixation device for disorders of the lumbar and thoracolumbar spine10.1097/00003086-198602000-00006 ↗
45. Pedicle screw motion in the osteoporotic spine After augmentation With laminar hooks, sublaminar wires, or calcium phosphate cement: a comparative analysis10.1097/01.BRS.0000134569.63542.49 ↗
46. Effect of various factors on pull out strength of pedicle screw in normal and osteoporotic cancellous bone models10.1016/j.medengphy.2016.11.012 ↗
47. Quantitative bone strength measurements in senile osteoporosis10.1093/geronj/16.2.120 ↗
48. Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants10.1007/s005860050045 ↗
49. A pedicle screw system and a lamina hook system provide similar primary and long-term stability: a biomechanical in vitro study with quasi-static and dynamic loading conditions10.1007/s00586-016-4679-x ↗
50. A comparative study on screw loosening in osteoporotic lumbar spine fusion between expandable and conventional pedicle screws10.1007/s00402-011-1439-6 ↗
51. The comparison of pullout strengths of various pedicle screw designs on synthetic foams and ovine vertebrae
52. Geometrical dimensions of the lower lumbar vertebrae–analysis of data from digitised CT images10.1007/s005860000140 ↗
53. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine10.1097/00003086-198602000-00012 ↗
54. Analysis of the morphometric characteristics of the thoracic and lumbar pedicles10.1097/00007632-198703000-00012 ↗
55. Sawbones® Europe AB, 2016. Biomechanical Test Materials. 〈www.sawbones.com/UserFiles/Docs/biomechanical_catalog.pdf〉 Last (Accessed 22 February 2017).
56. Bergmann, G., 2008. Charité Universitaetsmedizin Berlin OrthoLoad. 〈http://www.OrthoLoad.com〉 Last (Accessed 22 February 2017).
57. Nowak, B., 2013. The proposition of a numerical model of the experimental set-up for testing bone-screw interface with use of the foam vertebra model. In: Proceedings of the 18th International Symposium on Computational Biomechanics, pp. 57–58. 〈https://oparu.uni-ulm.de/xmlui/bitstream/handle/123456789/3208/vts_8937_13402.pdf?Sequence=1&isAllowed=y〉. Last (Accessed 22 February 2017).