Experimental study of the mechanical behavior of an explanted mesh: The influence of healing
Source abstract
Abstract
To better understand the in vivo mechanical behavior of synthetic mesh implants, we designed a specific experimental protocol for the mechanical characterization of explanted mesh under uniaxial tension. The implantation of a mesh leads to the development of scar tissue and the formation of a new composite made of native tissue, a mesh implant and scar tissues. This study focused on three points: determining the minimum representative size of mesh implants required for mechanical test samples, highlighting the influence of healing, and defining the healing time required to ensure stabilized mechanical properties. First, we determined the minimum representative size of mesh implants for the mechanical characterization with a study on a synthetic composite made of mesh and an elastomeric matrix mimicking the biological tissues. The size of the samples tested was gradually decreased. The downsizing process was stopped, when the mechanical properties of the composite were not preserved under uniaxial tension. It led to a sample representative size 3cm long and 2cm wide between the grips. Then an animal study was conducted on Wistar rats divided into eight groups. One group was set as control, consisting of the healthy abdominal wall. The other seven groups underwent surgery as follows: one placebo (i.e., without mesh placement), and six with a mesh installation on the abdominal wall and healing time. The rats were sacrificed after different healing times ranging from 1 to 5 months. We observed the influence of healing and healing time on the mechanical response under uniaxial tension of the new composite formed by scar, native tissue, and textile. It seems that 2 months are required to ensure the stabilization of the mechanical properties of the implanted mesh. We were not able to tell the control group (native abdominal wall) from the placebo group (native and scar tissue). This protocol was tested on two different prostheses after 3 months of healing. With this protocol, we were able to differentiate one mesh from another after host integration.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Pelvic floor disorders treatments · Hernia repair and management · Pelvic and Acetabular Injuries
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (46)
Show source-provided references
1. Abed, H., Rahn, D.D., Lowenstein, L., Balk, E.M., Clemons, J.L., Rogers, R.G., 2011. Incidence and management of graft erosion, wound granulation, and dyspareunia following vaginal prolapse repair with graft materials: a systematic review.
2. Amid, P.K., 1997. Classification of biomaterials and their related complications in abdominal wall hernia surgery, pp. 15–21.10.1007/BF02426382 ↗
3. Integration of biomaterials implanted into abdominal wall10.1016/0142-9612(95)98855-8 ↗
4. Integration of biomaterials implanted into abdominal wall10.1016/0142-9612(95)98855-8 ↗
5. Healing process induced by three composite prostheses in the repair of abdominal wall defects10.1002/jbm.10123 ↗
6. Integration of biomaterials implanted into abdominal wall10.1016/0142-9612(95)98855-8 ↗
7. Use of vaginal mesh for pelvic organ prolapse repair
8. Study of the biomechanical properties of synthetic mesh implanted in vivo10.1016/j.ejogrb.2007.02.023 ↗
9. Tissue integration and tolerance to meshes used in gynecologic surgery10.1016/j.ejogrb.2005.07.029 ↗
10. Mechanical properties of pelvic soft tissue of young women and impact of aging10.1007/s00192-014-2439-1 ↗
11. Corr, D.T., Gallant-behm, C.L., Shrive, N.G., Hart, D.A., 2009. Biomechanical behavior of scar tissue and uninjured skin in a porcine model.10.1111/j.1524-475X.2009.00463.x ↗
12. Synthetic meshes for transvaginal surgical cure of genital prolapse10.1016/S0368-2315(06)76416-X ↗
13. Characterisation of clinical and newly fabricated meshes for pelvic organ prolapse repair10.1016/j.jmbbm.2013.04.002 ↗
14. A new method to measure elastic properties of plastic-viscoelastic connective tissue10.1016/S1350-4533(98)00026-5 ↗
15. Deterioration in biomechanical properties of the vagina following implantation of a high-stiffness prolapse mesh10.1111/1471-0528.12077 ↗
16. Characterizing the ex vivo textile and structural properties of synthetic prolapse mesh products10.1007/s00192-012-1901-1 ↗
17. Molecular behaviour of elastomeric materials under large deformation10.1016/0032-3861(92)90726-D ↗
18. Vagina, abdominal skin, and aponeurosis10.1007/s00192-010-1237-7 ↗
19. Pelvic organ prolapse in the womens health initiative10.1067/mob.2002.123819 ↗
20. Long-term anisotropic mechanical response of surgical meshes used to repair abdominal wall defects10.1016/j.jmbbm.2011.09.005 ↗
21. ISIS4D, 2015. Isis4d – In situ Innovative Set-ups Under X-ray Microtomography. URL 〈http://isis4d.univ-lille1.fr/〉
22. Biomechanical properties of prolapsed or non-prolapsed vaginal tissue10.1007/s00192-010-1208-z ↗
23. Tensile properties of commonly used prolapse meshes10.1007/s00192-008-0781-x ↗
24. Impact of polymer pore size on the interface scar formation in a rat model10.1006/jsre.2002.6358 ↗
25. Modified mesh for hernia repair that is adapted to the physiology of the abdominal wall10.1080/110241598750005138 ↗
26. Functional and morphological evaluation of different polypropylene-mesh modifications for abdominal wall repair10.1016/S0142-9612(98)00115-X ↗
27. Biomechanical and histological evaluation of abdominal wall compliance with intraperitoneal onlay mesh implants in rabbits10.1016/j.medengphy.2011.09.022 ↗
28. Biomechanical properties of raw meshes used in pelvic floor reconstruction10.1007/s00192-008-0711-y ↗
29. Pore size and pore shape-but not mesh density-alter the mechanical strength of tissue ingrowth and host tissue response to synthetic mesh materials in a porcine model of ventral hernia repair10.1016/j.jmbbm.2014.11.011 ↗
30. Melman, L., Jenkins, E. D., Hamilton, N.A., Bender, L.C., Brodt, M.D., Deeken, C. R., Greco, S.C., Frisella, M.M., Matthews, B.D., 2011. Histologic and biomechanical evaluation of a novel macroporous polytetra, pp. 423–431.10.1007/s10029-011-0787-z ↗
31. Biomaterial supports for colonic wall defect healing10.1016/0142-9612(96)87283-8 ↗
32. Lightweight meshes10.1007/s00404-013-3093-4 ↗
33. Experimental comparison of abdominal wall repair using different methods of enhancement by small intestinal submucosa graft10.1007/s00192-008-0793-6 ↗
34. Shrinkage and biomechanical evaluation of lightweight synthetics in a rabbit model for primary fascial repair10.1007/s00192-011-1440-1 ↗
35. Pascual, G., Pe, B., 2014. Postimplantation host tissue response and biodegradation of biologic versus polymer meshes implanted in an intraperitoneal position, pp. 559–569.10.1007/s00464-013-3205-z ↗
36. Experimental study and constitutive modeling of the viscoelastic mechanical properties of the human prolapsed vaginal tissue10.1007/s10237-009-0157-2 ↗
37. Biomechanical properties of synthetic and biologic graft materials following long-term implantation in the rabbit abdomen and vagina
38. R Core Team, 2013. R: A Language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria. URL 〈http://www.r-project.org/〉
39. A non-biological model system to simulate the in vivo mechanical behavior of prosthetic meshes10.1016/j.jmbbm.2013.01.029 ↗
40. Röhrnbauer, B., Mazza, E., 2014. Uniaxial and biaxial mechanical characterization of a prosthetic mesh at different length scales. J. Mech. Behav. Biomed. Mater. 29 (January), 7-1910.1016/j.jmbbm.2013.07.021 ↗
41. Biomechanical properties of vaginal tissue10.1007/s00192-007-0533-3 ↗
42. Biomechanical properties of vaginal tissue. Part 110.1016/j.juro.2007.03.040 ↗
43. Biomechanical properties of human pelvic organs
44. Pelvic Organ Support Study (POSST)10.1016/j.ajog.2004.10.602 ↗
45. U.S. Food and Drug Administration, 2011. FDA Safety Communication: Update on Serious Complications Associated with Transvaginal Placement of Surgicalmesh for Pelvic Organ Prolapse. URL 〈www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm262435.htm〉.
46. Some forms of the strain energy function for rubber10.5254/1.3538343 ↗