Viscoelastic properties of a spinal posterior dynamic stabilisation device

Bernard M. Lawless, Spencer C. Barnes, Daniel M. Espino, Duncan E.T. Shepherd

Abstract

The purpose of this study was to quantify the frequency dependent viscoelastic properties of two types of spinal posterior dynamic stabilisation devices. In air at 37°C, the viscoelastic properties of six BDyn 1 level, six BDyn 2 level posterior dynamic stabilisation devices (S14 Implants, Pessac, France) and its elastomeric components (polycarbonate urethane and silicone) were measured using Dynamic Mechanical Analysis. The viscoelastic properties were measured over the frequency range 0.01-30Hz. The BDyn devices and its components were viscoelastic throughout the frequency range tested. The mean storage stiffness and mean loss stiffness of the BDyn 1 level device, BDyn 2 level device, silicone component and polycarbonate urethane component all presented a logarithmic relationship with respect to frequency. The storage stiffness of the BDyn 1 level device ranged from 95.56N/mm to 119.29N/mm, while the BDyn 2 level storage stiffness ranged from 39.41N/mm to 42.82N/mm. BDyn 1 level device and BDyn 2 level device loss stiffness ranged from 10.72N/mm to 23.42N/mm and 4.26N/mm to 9.57N/mm, respectively. No resonant frequencies were recorded for the devices or its components. The elastic property of BDyn 1 level device is influenced by the PCU and silicone components, in the physiological frequency range. The viscoelastic properties calculated in this study may be compared to spinal devices and spinal structures.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Spine and Intervertebral Disc Pathology · Musculoskeletal pain and rehabilitation · Scoliosis diagnosis and treatment

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (43)

Show source-provided references
1. ASTM, 2011. ASTM F2346-05: Standard Test Methods for Static and Dynamic Characterization of Spinal Artificial Discs, 10.1520/F2346-05R11
2. ASTM, 2014. ASTM F2193: Standard Specifications and Test Methods for Metallic Angled Orthopedic Fracture Fixation Devices, 10.1520/F0384-12
3. Frequency dependent viscoelastic properties of porcine bladder10.1016/j.jmbbm.2014.11.017 ↗
4. Mechanical characterisation of a viscoelastic disc for lumbar total disc replacement
5. Carfagni, M., Lenzi, E., Pierini, M., 1998. The loss factor as a measure of mechanical damping. In: 1998 IMAC XVI – 16th International Modal Analysis Conference. pp. 580–584.
6. A comparison of the shock-absorbing properties of cervical disc prosthesis bearing materials10.1016/j.esas.2011.01.002 ↗
7. Viscoelastic properties of bovine knee joint articular cartilage: dependency on thickness and loading frequency10.1186/1471-2474-15-205 ↗
8. Viscoelastic properties of bovine articular cartilage attached to subchondral bone at high frequencies10.1186/1471-2474-10-61 ↗
9. Viscoelastic properties of the intervertebral disc and the effect of nucleus pulposus removal10.1177/2041303310393410 ↗
10. Degenerative lumbar spondylolisthesis. Cohort of 670 patients, and proposal of a new classificatio10.1016/j.otsr.2014.07.006 ↗
11. A multi-component fiber-reinforced PHEMA-based Hydrogel/HAPEXTM device for customized intervertebral disc prosthesis10.1177/0885328209360933 ↗
12. Guerin, P., Gille, O., Persohn, S., Campana, S., Vital, J.M., Skalli, W., 2011. Effect of new dynamic stabilization system on the segmental motion and intradiscal pressure: An in vitro biomechanical study. In: ORS 2011 Annual Meeting.
13. Analysis of load-relaxation in compressed segments of lumbar spine10.1016/1350-4533(95)00047-X ↗
14. Systematic review of psychosocial factors at work and private life as risk factors for back pain10.1097/00007632-200008150-00017 ↗
15. The epidemiology of low back pain10.1016/j.berh.2010.10.002 ↗
16. ISO, 2011. BS EN ISO 6721: Plastics — Determination of dynamic mechanical properties.
17. Dynamic stiffness and damping of human intervertebral disc using axial oscillatory displacement under a free mass system10.1007/s00586-003-0569-0 ↗
18. Dynamics of human lumbar intervertebral joints: experimental and finite-element investigations10.1097/00007632-199201000-00014 ↗
19. Classification of posterior dynamic stabilization devices10.3171/foc.2007.22.1.3 ↗
20. Kurtz · 2006
21. Shock absorption in lumbar disc prosthesis: a preliminary mechanical study10.1097/00024720-200308000-00006 ↗
22. Are occupational drivers at an increased risk for developing musculoskeletal disorders?10.1097/00007632-199603150-00010 ↗
23. Frequency dependence of viscoelastic properties of medical grade silicones
24. Effect of axial load on the flexural properties of an elastomeric total disc replacement10.1097/BRS.0b013e31824da3ba ↗
25. Nonlinear dynamics of the human lumbar intervertebral disc10.1016/j.jbiomech.2014.12.006 ↗
26. Menard · 2008
27. Monède-Hocquard, L., Mesnard, M., Ramos, A., Gille, O., 2013. Optimization of a dynamic spinal implant: Selection of a polymer material. In: Proceedings of the XXIV Congress of the International Society of Biomechanics, Brazil, pp. 3–5.
28. Dynamic and quasi-static mechanical testing for characterization of the viscoelastic properties of human uterine tissue10.1016/j.jbiomech.2015.05.013 ↗
29. In vivo measurements of spinal column vibrations10.2106/00004623-198668050-00009 ↗
30. Spinal fusion in the United States10.1097/BRS.0b013e31820cccfb ↗
31. Preclinical and clinical experience with a viscoelastic total disc replacement10.1016/j.esas.2011.08.001 ↗
32. Behaviour of the spine under shock and vibration: a review10.1016/0268-0033(88)90045-9 ↗
33. A reanalysis of spinal motion during vibration10.1080/00140138708969792 ↗
34. Segment-by-segment stabilization for degenerative disc disease: a hybrid technique10.1007/s00586-010-1282-4 ↗
35. Dynamic stabilization devices in the treatment of low back pain10.1016/S0030-5898(03)00087-7 ↗
36. Motion-preserving technologies for degenerative lumbar spine: the past, present and future horizons10.1016/j.esas.2011.05.001 ↗
37. Dynamic compressive properties of human lumbar intervertebral joints: a comparison between fresh and thawed specimens10.1016/0021-9290(88)90148-0 ↗
38. Fatigue of biomaterials: a review10.1016/S0142-1123(00)00052-9 ↗
39. Biomechanical behavior of a biomimetic artificial intervertebral disc10.1097/BRS.0b013e3182326305 ↗
40. Design of next generation total disk replacements10.1016/j.jbiomech.2011.09.017 ↗
41. Frequency and diameter dependent viscoelastic properties of mitral valve chordae tendineae10.1016/j.jmbbm.2013.11.013 ↗
42. Vibration and the human spine10.1097/00007632-198205000-00008 ↗
43. Shock absorbing function study on denucleated intervertebral disc with or without hydrogel injection through static and dynamic biomechanical tests in vitro