journal article·2021-07Open access

Analytical methods for braided stents design and comparison with FEA

Alissa Zaccaria, Giancarlo Pennati, Lorenza Petrini

Abstract

Braiding technology is nowadays commonly adopted to build stent-like devices. Indeed, these endoprostheses, thanks to their typical great flexibility and kinking resistance, find several applications in mini-invasive treatments, involving but not limiting to the cardiovascular field. The design process usually involves many efforts and long trial and error processes before identifying the best combination of manufacturing parameters. This paper aims to provide analytical tools to support the design and optimization phases: the developed equations, based on few geometrical parameters commonly used for describing braided stents and material stiffness, are easily implementable in a worksheet and allow predicting the radial rigidity of braided stents, also involving complex features such as multiple twists and looped ends, and the diameter variation range. Finite element simulations, previously validated with respect to experimental tests, were used as a comparator to prove the reliability of the analytical results. The illustrated tools can assess the impact of each selected parameter modification and are intended to guide the optimal selection of geometrical and mechanical stent proprieties to obtain the desired radial rigidity, deliverability (minimum diameter), and, if forming processes are planned to modify the shape of the stent, the required diameter variations (maximum and minimum diameters).
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Coronary Interventions and Diagnostics · Cardiac Valve Diseases and Treatments · Corrosion Behavior and Inhibition

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (33)

Show source-provided references
1. Supera self-expanding stents for endovascular treatment of femoropopliteal disease: a review of the clinical evidence
2. Design optimisation of coronary artery stent systems10.1007/s10439-015-1373-9 ↗
3. Treatment of middle cerebral artery aneurysms with flow- diverter stents: a systematic review and meta-Analysis10.3174/ajnr.A5388 ↗
4. Mechanical properties and performances of contemporary drug-eluting stent: focus on the metallic backbone10.1080/17434440.2019.1573142 ↗
5. Carotid artery stenting: an update10.1093/eurheartj/ehu446 ↗
6. Latest advancements in ureteral stent technology10.21037/tau.2019.08.16 ↗
7. Novel mitral valve technologies-transcatheter mitral valve implantation: a systematic review10.21037/acs.2018.11.01 ↗
8. Treatment of tracheobronchial injuries: a contemporary review10.1016/j.chest.2018.07.018 ↗
9. A study of the geometrical and mechanical properties of a self‐expanding metallic stent—theory and experiment10.1002/jab.770040111 ↗
10. Stents: biomechanics, biomaterials, and insights from computational modeling10.1007/s10439-017-1806-8 ↗
11. Comparison of computational modelling techniques for braided stent analysis10.1080/10255842.2019.1663414 ↗
12. Current developments in endovascular therapy of peripheral vascular disease10.21037/jtd.2019.12.130 ↗
13. Design methodology of a balloon expandable polymeric stent10.35248/2475-7586.19.4.139 ↗
14. Computer modeling of deployment and mechanical expansion of neurovascular flow diverter in patient-specific intracranial aneurysms10.1016/j.jbiomech.2012.06.013 ↗
15. Amplatzer patent foramen ovale occluder: safety and efficacy10.1080/17434440.2019.1581060 ↗
16. Endovascular treatment of various aortic pathologies: review of the latest data and technologies10.1055/s-0038-1645881 ↗
17. Transcatheter aortic valve replacement (TAVR/TAVI , percutaneous replacement) contraindications
18. An experimental evaluation of the mechanics of bare and polymer-covered self-expanding wire braided stents10.1016/j.jmbbm.2019.103549 ↗
19. A finite element investigation on design parameters of bare and polymer-covered self-expanding wire braided stents10.1016/j.jmbbm.2020.104305 ↗
20. Modeling stented coronary arteries: where we are, where to go10.1007/s10439-012-0681-6 ↗
21. Surveying the 2019 venous stent landscape
22. Percutaneous devices for left atrial appendage occlusion: a contemporary review10.4330/wjc.v11.i2.57 ↗
23. A literature review of the numerical analysis of abdominal aortic aneurysms treated with endovascular stent grafts10.1155/2012/820389 ↗
24. Coronary stents: history, design, and construction10.3390/jcm7060126 ↗
25. The Lotus valve system: an in-depth review of the technology10.1007/s11886-019-1234-5 ↗
26. Looped ends versus open ends braided stent: a comparison of the mechanical behaviour using analytical and numerical methods10.1016/j.jmbbm.2017.08.025 ↗
27. Echocardiographic guidance of AMPLATZER amulet left atrial appendage occlusion device placement10.1177/1089253218758463 ↗
28. Esophageal stenting in clinical practice: an overview10.1007/s11938-018-0181-3 ↗
29. Long and short of optimal stent design
30. Endovascular interventional cardiology: 2015 in review10.1111/joic.12273 ↗
31. Recent review of transcatheter closure of atrial septal defect10.1016/j.kjms.2018.05.001 ↗
32. Finite element simulations of the ID venous system to treat venous compression Disorders : from model validation to realistic implant prediction