Stented valve dynamic behavior induced by polyester fiber leaflet material in transcatheter aortic valve devices
Source abstract
Abstract
Objective This study aims at assessing the global dynamic behavior, elastic deformability, closing energy and turbulence of rigid versus deformable stented (RS vs DS) valve systems with deformable and rigid textile materials (DT vs RT) through studying the stent-valve interaction compared to a bioprosthetic material in transcatheter aortic valves (TAV). Methods Three 19 mm stented textile TAV designs (RS-DT, RS-RT and DS-RT) with different stent and leaflet properties were tested and compared with a control bioprosthetic TAV (RS-DB) in a left heart simulator flow loop under physiological pressure and flow. Particle Image Velocimetry and high speed imaging were performed. Pressure gradients (PG), leakage fractions (LF), Pinwheeling indices (PI), closing energy (E) and Reynolds shear stresses (RSS) were calculated. Results (a) PGs and LFs were 11.86 ± 0.51 mmHg, 11.70 ± 0.34%; 8.84 ± 0.40 mmHg, 29.80 ± 0.76%; 11.59 ± 0.12 mmHg, 14.23 ± 1.64%; and 7.05 ± 0.09 mmHg, 12.08 ± 0.45% % for RS-DB, RS-DT, RS-RT and DS-RT respectively. (b) PIs were 15.79 ± 2.34%, 4.36 ± 0.84%, 2.47 ± 0.51% and 2.03 ± 0.33% for RS-DB, RS-DT, RS-RT and DS-RT respectively. (c) E is lowest for DS-RT (0.0010 ± 0.0002 J) followed by RS-RT (0.0017 ± 0.0002 J), RS-DB (0.0023 ± 0.0004 J) and highest with RS-DT (0.0036 ± 0.0007 J). (d) At peak systole lowest RSS was obtained with RS-DT (87.82 ± 0.58 Pa) and highest with DS-RT (122.98 ± 1.87 Pa). Conclusion PGs, LFs, PIs and E were improved with DS-RT compared to other textile TAVs and RS-DB. Despite achieving more RSS than the rest of TAVs, DS-RT still falls within the same range of RSS produced by the other 2 valves and control exceeding the threshold for platelet activation.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Cardiac Valve Diseases and Treatments · Soft Robotics and Applications · Elasticity and Material Modeling
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (36)
Show source-provided references
1. 'A general two-equation Macroscopic turbulence model for incompressible flow in porous media'10.1016/S0017-9310(96)00370-5 ↗
2. 'On the mechanics of Transcatheter aortic valve replacement'10.1007/s10439-016-1759-3 ↗
3. 'Fluid Mechanics of Artificial Heart Valves'10.1111/j.1440-1681.2008.05099.x ↗
4. Dasi, Lakshmi Prasad, Morshed, Khandakar Niaz, Forleo, Marcio, 2013. Phenomenology of Hemolysis in Turbulent Flows. In: Proceedings of the ASME 2013 Summer Bioengineering Conference (American Society of Mechanical Engineers, pp. V01AT05A020-V01AT05A20.
5. 'valve-in-valve outcome: design impact of a pre-existing bioprosthesis on the hydrodynamics of an edwards Sapien Xt valve'10.1093/ejcts/ezw317 ↗
6. Estimation of shear stress-related blood damage in heart valve prostheses--in vitro comparison of 25 aortic valves10.1177/039139889001300507 ↗
7. Estimation of shear stress-related blood damage in heart valve prostheses-in vitro comparison of 25 aortic valves10.1177/039139889001300507 ↗
8. An in vitro evaluation of the impact of eccentric deployment on transcatheter aortic valve hemodynamics10.1007/s10439-014-1008-6 ↗
9. Total ellipse of the heart valve: the impact of eccentric stent distortion on the regional dynamic deformation of pericardial tissue leaflets of a transcatheter aortic valve replacement'10.1098/rsif.2015.0737 ↗
10. Turbulence downstream from the Ionescu-Shiley bioprosthesis in steady and pulsatile flow10.1007/BF02447332 ↗
11. Effect of severe bioprosthetic valve tissue ingrowth and inflow calcification on valve-in-valve performance'10.1016/j.jbiomech.2018.04.039 ↗
12. Implantation depth and rotational orientation effect on valve-in-valve hemodynamics and sinus flow10.1016/j.athoracsur.2018.01.070 ↗
13. Impact of patient morphologies on sinus flow stasis in transcatheter aortic valve replacement: an in-vitro study'
14. An in-vitro evaluation of turbulence after transcatheter aortic valve implantation'10.1016/j.jtcvs.2018.05.042 ↗
15. Aortic sinus flow stasis likely in valve-in-valve transcatheter aortic valve implantation10.1016/j.jtcvs.2017.03.053 ↗
16. A relationship between reynolds stresses and viscous dissipation: implications to red cell damage'10.1007/BF02368297 ↗
17. Transcatheter fiber heart valve: effect of crimping on material performances'10.1002/jbm.b.33330 ↗
18. New trends in experimental turbulence research'10.1146/annurev.fl.07.010175.001515 ↗
19. Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress10.1002/cnm.2814 ↗
20. Turbulence characteristics downstream of bileaflet aortic valve prostheses'10.1115/1.429643 ↗
21. Effect of aortic annulus size and prosthesis oversizing on the hemodynamics and leaflet bending stress of transcatheter valves: an in-vitro study'
22. Simulation of long-term fatigue damage in Bioprosthetic heart valves: effects of leaflet and stent elastic properties'10.1007/s10237-013-0532-x ↗
23. Theory to predict shear stress on cells in turbulent blood flow'10.1371/journal.pone.0105357 ↗
24. Flow simulations over porous media–comparisons with experiments10.1016/j.compfluid.2017.03.002 ↗
25. 'two-dimensional color-mapping of turbulent shear stress distribution downstream of two aortic Bioprosthetic valves in vitro'10.1016/0021-9290(92)90262-Y ↗
26. 'An engineering review of transcatheter aortic valve technologies'10.1007/s13239-010-0008-4 ↗
27. 'Shear-induced activation of platelets'10.1016/0021-9290(79)90150-7 ↗
28. A matter of thrombosis: different thrombus-Like formations in balloon-expandable transcatheter aortic valve prostheses10.1093/ejcts/ezy033 ↗
29. Transcatheter aortic valve implantation: current and future approaches10.1038/nrcardio.2011.164 ↗
30. Velocity and turbulence measurements Past mitrial valve prostheses in a model left ventricle10.1016/0021-9290(91)90288-X ↗
31. Bioprosthetic heart valves: impact of implantation on biomaterials'10.5402/2013/728791 ↗
32. Fiber heart valve prosthesis: influence of the fabric construction parameters on the valve fatigue performances'10.1016/j.jmbbm.2014.08.015 ↗
33. Heart valves from polyester fibers: a preliminary 6-month in vivo study'
34. Flow in prosthetic heart valves: state-of-the-art and future directions'10.1007/s10439-005-8759-z ↗
35. Heart valves from polyester fibers Vs. biological tissue: comparative study in vitro'10.1007/s10439-016-1632-4 ↗
36. Highly resolved pulsatile flows through prosthetic heart valves using the entropic lattice-Boltzmann method'10.1017/jfm.2014.393 ↗