Validation of the computational model of a coronary stent: a fundamental step towards in silico trials
Source abstract
Abstract
The proof of the reliability of a numerical model is becoming of paramount importance in the era of in silico clinical trials. When dealing with a coronary stenting procedure, the virtual scenario should be able to replicate the real device, passing through the different stages of the procedure, which has to maintain the atherosclerotic vessel opened. Nevertheless, most of the published studies adopted commercially resembling geometries and generic material parameters, without a specific validation of the employed numerical models. In this work, a workflow for the generation and validation of the computational model of a coronary stent was proposed. Possible sources of variability in the results, such as the inter-batches variability in the material properties and the choice of proper simulation strategies, were accounted for and discussed. Then, a group of in vitro tests, representative of the device intended use was used as a comparator to validate the model. The free expansion simulation, which is the most used simulation in the literature, was shown to be only partially useful for stent model validation purposes. On the other hand, the choice of proper additional experiments, as the suggested uniaxial tensile tests on the stent and deployment tests into a deformable tube, could provide further suitable information to prove the efficacy of the numerical approach.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Coronary Interventions and Diagnostics · Elasticity and Material Modeling · Cardiac Valve Diseases and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (23)
Show source-provided references
1. ASME · 2018
2. ASME · 2006
3. BSI · 2012
4. Numerical analysis of crimping and inflation process of balloon-expandable coronary stent using implicit solution10.1002/cnm.2890 ↗
5. Biomechanical impact of wrong positioning of a dedicated stent for coronary bifurcations: a virtual bench testing study
6. Numerical simulation of stent angioplasty with predilation: an investigation into lesion constitutive representation and calcification influence10.1007/s10439-017-1851-3 ↗
7. A finite element strategy to investigate the free expansion behaviour of a biodegradable polymeric stent10.1016/j.jbiomech.2015.03.024 ↗
8. Modelling of the provisional side-branch stenting approach for the treatment of atherosclerotic coronary bifurcations: effects of stent positioning10.1007/s10237-010-0196-8 ↗
9. On the importance of modeling balloon folding, pleating, and stent crimping: an FE study comparing experimental inflation tests10.1002/cnm.3249 ↗
10. Cardiovascular stent design and vessel stresses: a finite element analysis10.1016/j.jbiomech.2004.07.022 ↗
11. Comparison of percutaneous coronary intervention with bare-metal and drug-eluting stents for cardiac allograft vasculopathy10.1016/j.jcin.2008.10.001 ↗
12. Finite element analysis of balloon-expandable coronary stent deployment: influence of angioplasty balloon configuratio10.1002/cnm.2557 ↗
13. Computational structural modelling of coronary stent deployment: a review10.1080/10255841003766845 ↗
14. Simultaneous kissing stents to treat unprotected left main stem coronary artery bifurcation disease; stent expansion, vessel injury, hemodynamics, tissue healing, restenosis, and repeat revascularization
15. Assessing computational model credibility using a risk-based framework: application to hemolysis in centrifugal blood pumps10.1097/MAT.0000000000000996 ↗
16. A novel simulation strategy for stent insertion and deployment in curved coronary bifurcations: comparison of three drug-eluting stents10.1007/s10439-009-9836-5 ↗
17. A platinum-chromium steel for cardiovascular stents10.1016/j.biomaterials.2010.01.146 ↗
18. Research into biodegradable polymeric stents: a review of experimental and modelling work10.20517/2574-1209.2018.13 ↗
19. Effects of material, coating, design and plaque composition on stent deployment inside a stenotic artery - finite element simulation10.1016/j.msec.2014.05.057 ↗
20. Position paper: credibility of in silico trial technologies - a theoretical framing
21. DES or BMS in acute myocardial infarction?10.1093/eurheartj/ehm474 ↗
22. Experimentally validated simulation of coronary stents considering different dogboning ratios and asymmetric stent positioning10.1371/journal.pone.0224026 ↗
23. On the importance of modeling stent procedure for predicting arterial mechanics10.1115/1.4023094 ↗