Carbon-fibre-reinforced PEEK: An alternative material for flexion bushings of rotating hinged knee joints?

Stefan Schroeder, Steffen Braun, Ulrike Mueller, Matthias Vogel, Robert Sonntag, Sebastian Jaeger, Jan Philippe Kretzer

Publisher page

Abstract

Background For prosthetic knee joints of the hinged type, typically polyethylene (PE) flexion bushings are used between axis and femoral component to prevent metallic wear. Nevertheless, PE-wear can lead to periprosthetic osteolysis followed by aseptic loosening of the implant. Based on high creep and wear resistance carbon-fibre-reinforced polyether ether ketones (CFR-PEEKs) could provide an alternative material to ultra-high-molecular-weight polyethylene (UHMWPE) for this bearing type. Methods Flexion bushings of four different materials were investigated (CFR-PEEK pitch fibres, CFR-PEEK PAN fibres, virgin PEEK without carbon fibres and UHMWPE) using a bushing tester. For determination of the polymeric and metallic wear, gravimetric measurements and particle analyses were performed. Results The polymeric wear rates of CFR-PEEK PAN (6.657 ± 0.714 mg/106 cycles) and CFR-PEEK pitch (32.085 ± 2.748 mg/106 cycles) were significantly higher compared to the polymeric wear rates of virgin PEEK (0.764 ± 0.283 mg/106 cycles) or UHMWPE (-0.015 ± 0.011 mg/106 cycles) (p < .001). In addition, the metallic wear rates were significantly higher when using bushings made of CFR-PEEK PAN (3.373 ± 0.214 mg/106 cycles) and CFR-PEEK pitch (3.604 ± 0.355 mg/106 cycles), than when using bushings made of virgin PEEK (0.177 ± 0.049 mg/106 cycles) or UHMWPE (0.031 ± 0.011 mg/106 cycles) (p < .001). The mean particle size of the two CFR-PEEK groups was significantly smaller compared to the virgin PEEK and UHMWPE group (p < .001). Conclusion According to these results, the clinical use of CFR-PEEK in combination with CoCr could lead to higher wear rates and this should be avoided. Virgin PEEK may still be considered as an alternative to UHMWPE to use it for flexion bushings in a hinged knee joint.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Total Knee Arthroplasty Outcomes · Orthopaedic implants and arthroplasty · Tribology and Wear Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (43)

Show source-provided references
1. Australian Orthopaedic Association
2. Metal wear particles: what we know, what we do not know, and why10.1016/j.esas.2009.11.006 ↗
3. Biological response to prosthetic debris10.5312/wjo.v6.i2.172 ↗
4. Mass gain behaviour of tibial polyethylene inserts during soak testing10.1177/2041303310392629 ↗
5. Influence of joint kinematics on polyethylene wear in anatomic shoulder joint arthroplasty10.1016/j.jse.2018.02.063 ↗
6. Influence of contact pressure, cross-shear and counterface material on the wear of PEEK and CFR-PEEK for orthopaedic applications10.1016/j.jmbbm.2016.06.005 ↗
7. PEEK and CFR-PEEK as alternative bearing materials to UHMWPE in a fixed bearing total knee replacement: an experimental wear study10.1016/j.wear.2016.12.010 ↗
8. Biological effects of clinically relevant wear particles from metal-on-metal hip prostheses10.1243/095441105X63291 ↗
9. Polyethylene and metal wear particles: characteristics and biological effects10.1007/s00281-011-0242-3 ↗
10. Patient satisfaction after total knee arthroplasty10.5792/ksrr.2016.28.1.1 ↗
11. Failure of the rotating-hinge knee megaprosthesis10.1016/j.arth.2012.07.030 ↗
12. Fatigue crack propagation behavior of ultrahigh molecular weight polyethylene10.1002/jor.1100020202 ↗
13. The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: experimental and finite-element analysis10.1007/s00586-005-0915-5 ↗
14. The role of hinges in primary total knee replacement10.1302/0301-620X.96B11.34143 ↗
15. Biotribology of alternative bearing materials for unicompartmental knee arthroplasty10.1016/j.actbio.2010.04.003 ↗
16. Biotribology of a new bearing material combination in a rotating hinge knee articulation10.1016/j.actbio.2013.02.030 ↗
17. Implant survival and complication profiles of endoprostheses for treating tumor around the knee in adults: a systematic review of the literature over the past 30 years10.1016/j.arth.2017.10.051 ↗
18. Macrophage reactivity to different polymers demonstrates particle size- and material-specific reactivity: PEEK-OPTIMA((R)) particles versus UHMWPE particles in the submicron, micron, and 10 micron size ranges10.1002/jbm.b.31974 ↗
19. Total knee replacement: an evidence-based analysis
20. Quantification of the effect of cross-shear on the wear of conventional and highly cross-linked UHMWPE10.1016/j.jbiomech.2007.09.005 ↗
21. Rotating hinge prosthesis for complex revision total knee arthroplasty: a review of the literature10.1016/j.jcot.2017.11.020 ↗
22. Compressive creep characteristics of extruded ultrahigh-molecular-weight polyethylene10.1002/(SICI)1097-4636(199802)39:2<261::AID-JBM13>3.0.CO;2-G ↗
23. Repeated early failure of a newly designed hinged knee system10.1016/j.arth.2012.04.029 ↗
24. The biokinetics of inorganic cobalt in the human body10.1016/j.scitotenv.2007.08.054 ↗
25. A biomechanical evaluation of hinged total knee replacement prostheses10.1177/0954411913488506 ↗
26. Specific material effects of wear-particle-induced inflammation and osteolysis at the bone-implant interface: a rat model10.1016/j.jot.2016.06.026 ↗
27. Potential release of in vivo trace metals from metallic medical implants in the human body: from ions to nanoparticles--a systematic analytical review10.1016/j.actbio.2014.02.027 ↗
28. Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device10.1002/jbm.820150503 ↗
29. Comparison of three joint simulator wear debris isolation techniques: acid digestion, base digestion, and enzyme cleavage10.1002/1097-4636(200108)56:2<245::AID-JBM1091>3.0.CO;2-T ↗
30. Polyethylene wear particles induce TLR 2 upregulation in the synovial layer of mice10.1007/s10856-013-5095-y ↗
31. Component wear of total knee prostheses using Ti-6A1-4V, titanium nitride coated Ti-6Al-4V, and cobalt-chromium-molybdenum femoral components10.1002/jbm.820221005 ↗
32. How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?10.1007/s10856-014-5271-8 ↗
33. Association between UHMWPE particle-induced inflammatory osteoclastogenesis and expression of RANKL, VEGF, and Flt-1 in vivo10.1016/j.biomaterials.2006.04.004 ↗
34. Molecular analysis of chromium and cobalt-related toxicity10.1038/srep05729 ↗
35. Analysis of carbon fiber reinforced PEEK hinge mechanism articulation components in a rotating hinge knee design: a comparison of in vitro and retrieval findings10.1155/2016/7032830 ↗
36. The wear properties of CFR-PEEK-OPTIMA articulating against ceramic assessed on a multidirectional pin-on-plate machine10.1243/09544119JEIM224 ↗
37. Characterization of polyethylene wear particle: the impact of methodology10.1016/j.actbio.2013.07.039 ↗
38. Particle analysis of shape factors according to American Society for Testing and Materials
39. Morphological characteristics of total joint arthroplasty-derived ultra-high molecular weight polyethylene (UHMWPE) wear debris that provoke inflammation in a murine model of inflammation10.1002/jbm.a.10368 ↗
40. In vitro macrophage response to polyethylene and polycarbonate-urethane particles10.1002/jbm.a.32529 ↗
41. Polyetheretherketone as a biomaterial for spinal applications10.1016/j.biomaterials.2005.07.011 ↗
42. Inflammatory response against different carbon fiber-reinforced PEEK wear particles compared with UHMWPE in vivo10.1016/j.actbio.2010.06.002 ↗
43. Suitability and limitations of carbon fiber reinforced PEEK composites as bearing surfaces for total joint replacements10.1016/S0043-1648(99)00026-5 ↗