journal article·2022-10

Micro-CT scan optimisation for mechanical loading of tibia with titanium tibial tray: A digital volume correlation zero strain error analysis

Lauren S. Wearne, Sophie Rapagna, Mark Taylor, Egon Perilli

Publisher page

Abstract

Primary stability of press-fit tibial trays is achieved by introducing an interference fit between bone and implant. The internal cancellous bone strains induced during this process and during loading have yet to be quantified experimentally. Advancements in large-gantry micro-CT imaging and digital volume correlation (DVC) allow quantification of such strains. However, before undertaking such a test, experimental requirements and DVC performance need to be examined, particularly considering the presence of a large orthopaedic implant (tibial tray). The aim of this study was to assess the DVC zero-strain accuracy (mean absolute error: MAER) and precision (standard deviation of error: SDER) on a cadaveric human tibia implanted with a titanium press-fit tray across four plausible scanning configurations, using a cabinet micro-CT system (Nikon XT H 225 ST). These varied in rotation step and resulting scanning time (106 min vs. 66 min), presence or absence of a 2 mm-thick aluminium cylinder for mechanical testing, and X-ray tube voltage (150 kVp vs. 215 kVp). One proximal tibia was implanted and micro-CT scanned (42 μm/pixel), with repeated scanning and specimen repositioning in between. DVC (DaVis, LaVision, direct correlation) was performed on nine cubic volumes of interest (VOIs: 13.4 mm-side) and across the entire proximal tibia. Strain errors were comparable across the four scanning configurations and sufficiently low for assessing bone within its elastic region in VOIs (MAER=223-540 με; SDER=88-261 με) and at organ level (MAER=536 με; SDER=473 με). Whilst the investigated experimental conditions, including a large titanium implant, present added complexity for DVC analysis, scans of sufficient quality can be achieved, reaching a compromise between the DVC requirements and the wanted application. The approach used for choosing the X-ray source settings considering the transmitted X-ray signal intensity and source power, is also discussed.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (49)

Show source-provided references
1. 2020
2. Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue10.1016/S0021-9290(03)00257-4 ↗
3. Glenoid bone strain after anatomical total shoulder arthroplasty: in vitro measurements with micro-CT and digital volume correlation10.1016/j.medengphy.2020.09.009 ↗
4. The influence of tibial component fixation techniques on resorption of supporting bone stock after total knee replacement10.1016/j.jbiomech.2010.11.026 ↗
5. Strain shielding in proximal tibia of stemmed knee prosthesis: experimental study10.1016/j.jbiomech.2007.10.006 ↗
6. Croom · 2020 · 1
7. About the inevitable compromise between spatial resolution and accuracy of strain measurement for bone tissue: a 3D zero-strain study10.1016/j.jbiomech.2014.07.019 ↗
8. Consequences of an interference fit on the fixation of porous-coated tibial components in total knee replacement10.2106/00004623-199274020-00009 ↗
9. Biomechanics and strain mapping in bone as related to immediately-loaded dental implants10.1016/j.jbiomech.2015.05.014 ↗
10. Laboratory x-ray micro-computed tomography: a user guideline for biological samples10.1093/gigascience/gix027 ↗
11. The application of digital volume correlation (DVC) to study the microstructural behaviour of trabecular bone during compression10.1016/j.jmbbm.2013.09.014 ↗
12. Extracting accurate strain measurements in bone mechanics: a critical review of current methods10.1016/j.jmbbm.2015.06.006 ↗
13. Finite element analysis of tibial implants—effect of fixation design and friction model10.1080/10255840008915264 ↗
14. Hubbell · 1996
15. Digital volume correlation for study of the mechanics of whole bones10.1016/j.piutam.2012.05.013 ↗
16. Deviations from optimal alignment in TKA: is there a biomechanical difference between femoral or tibial component alignment?10.1016/j.arth.2015.07.038 ↗
17. Future young patient demand for primary and revision joint replacement: national projections from 2010 to 203010.1007/s11999-009-0834-6 ↗
18. Performance of QCT-derived scapula finite element models in predicting local displacements using digital volume correlation10.1016/j.jmbbm.2019.05.021 ↗
19. Characterization of the bone-metal implant interface by digital volume correlation of in-situ loading using neutron tomography10.1016/j.jmbbm.2017.07.001 ↗
20. Bone damage evolution around integrated metal screws using X-ray tomography—in situ pullout and digital volume correlation10.3389/fbioe.2020.00934 ↗
21. Micro-computed tomography for small animal imaging: technological details10.1016/j.pnsc.2008.01.002 ↗
22. Accuracy and precision of digital volume correlation in quantifying displacements and strains in trabecular bone10.1016/j.jbiomech.2007.04.019 ↗
23. Time-elapsed synchrotron-light microstructural imaging of femoral neck fracture10.1016/j.jmbbm.2018.05.016 ↗
24. Measurement of internal implantation strains in analogue bone using DVC10.3390/ma13184050 ↗
25. Mechanical effects of postmortem changes, preservation, and allograft bone treatments
26. First clinical study of a novel complete metal-free ceramic total knee replacement system10.1186/s13018-016-0352-7 ↗
27. 2016 · 1
28. 2018
29. Digital volume correlation can be used to estimate local strains in natural and augmented vertebrae: an organ-level study10.1016/j.jbiomech.2016.10.018 ↗
30. Effect of SR-microCT radiation on the mechanical integrity of trabecular bone using in situ mechanical testing and digital volume correlation10.1016/j.jmbbm.2018.08.012 ↗
31. Structural parameters and mechanical strength of cancellous bone in the femoral head in osteoarthritis do not depend on age10.1016/j.bone.2007.07.014 ↗
32. Failure strength of human vertebrae: prediction using bone mineral density measured by DXA and bone volume by micro-CT10.1016/j.bone.2012.03.002 ↗
33. Micro-CT Examination of human bone: from biopsies towards the entire organ
34. The importance of tibial alignment: finite element analysis of tibial malalignment10.1054/arth.2000.17941 ↗
35. A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration10.1016/j.jbiomech.2003.12.004 ↗
36. Cemented or cementless total knee arthroplasty?: comparative results of 200 cases at a minimum follow-up of 11 years10.1051/sicotj/2017046 ↗
37. Quantification of human bone microarchitecture damage in press-fit femoral knee implantation using HR-pQCT and digital volume correlation10.1016/j.jmbbm.2019.04.054 ↗
38. X-ray computed tomography in life sciences10.1186/s12915-020-0753-2 ↗
39. Application of the digital volume correlation technique for the measurement of displacement and strain fields in bone: a literature review10.1016/j.jbiomech.2014.01.001 ↗
40. Spatial resolution characterization of a X-ray microCT system10.1016/j.apradiso.2014.08.014 ↗
41. Failure of partial cementation to achieve implant stability and bone ingrowth: a long-term roentgen stereophotogrammetric study of tibial components10.1002/jor.1100170304 ↗
42. Stock · 2009
43. Digital volume correlation and micro-CT: an in-vitro technique for measuring full-field interface micromotion around polyethylene implants10.1016/j.jbiomech.2015.05.024 ↗
44. Influence of loading and activity on the primary stability of cementless tibial trays10.1002/jor.22056 ↗
45. Four decades of finite element analysis of orthopaedic devices: where are we now and what are the opportunities?10.1016/j.jbiomech.2014.12.019 ↗
46. Finite element analysis of the implanted proximal tibia: a relationship between the initial cancellous bone stresses and implant migration10.1016/S0021-9290(98)00022-0 ↗
47. Strain uncertainties from two digital volume correlation approaches in prophylactically augmented vertebrae: local analysis on bone and cement-bone microstructures10.1016/j.jmbbm.2016.12.006 ↗
48. 3D full-field strain in bone-implant and bone-tooth constructs and their morphological influential factors10.1016/j.jmbbm.2020.103858 ↗
49. 3D full-field biomechanical testing of a glenoid before and after implant placement10.1016/j.eml.2019.100614 ↗