Variability in reference point microindentation and recommendations for testing cortical bone: Maximum load, sample orientation, mode of use, sample preparation and measurement spacing
Source abstract
Abstract
Reference Point Indentation (RPI) is a novel microindentation tool that has emerging clinical potential for the assessment of fracture risk as well as use as a laboratory tool for straight-forward mechanical characterisation of bone. Despite increasing use of the tool, little research is available to advise the set-up of testing protocols or optimisation of testing parameters. Here we consider five such parameters: maximum load, sample orientation, mode of use, sample preparation and measurement spacing, to investigate how they affect the Indentation Distance Increase (IDI), the most published measurement parameter associated with the RPI device. The RPI tool was applied to bovine bone; indenting in the proximal midshaft of five femora and human bone; indenting five femoral heads and five femoral neck samples. Based on the findings of these studies we recommend the following as the best practice. (1) Repeat measurements should be utilised to reduce the coefficient of variation (e.g. 8-15 repeats to achieve a 5-10% error, however the 3-5 measurements used here gives a 15-20% error). (2) IDI is dependent on maximum load (r=0.45 on the periosteal surface and r=0.94 on the machined surface, p<0.05), mode of use (i.e. comparing the device held freehand compared to fixed in its stand, p=0.04) and surface preparation (p=0.004) so these should be kept consistent throughout testing. Though sample orientation appears to have minimal influence on IDI (p>0.05), care should also be taken in combining measurements from different orientations. (3) The coefficient of variation is higher (p=0.04) when holding the device freehand, so it should ideally be kept supported in its stand. (4) Removing the periosteum (p=0.04) and machining the surface of the bone (p=0.08) reduces the coefficient of variation, so should be performed where practical. (5) There is a hyperbolic relationship between thickness and IDI (p<0.001) with a sample thickness 10 fold greater than the maximum indentation depth recommended, to ensure a representative measurement. (6) Measurement spacing does not appear to influence the IDI (p>0.05), so it can be as low as 500 µm. By following these recommendations, RPI users can minimise the potential confounding effects associated with the variables investigated here and reduce the coefficient of variation, hence achieving more consistent testing. This optimisation of the technique enhances both the clinical and laboratory potential of the tool.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Bone health and osteoporosis research · Orthopedic Surgery and Rehabilitation
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (37)
Show source-provided references
1. Interspecies differences in bone composition, density, and quality: Potential implications for in vivo bone research10.1210/endo.139.2.5751 ↗
2. In vivo reference point indentation reveals positive effects of raloxifene on mechanical properties following 6 months of treatment in skeletally mature beagle dogs10.1016/j.bone.2013.07.009 ↗
3. Bone quality: where do we go from here?10.1007/s00198-003-1489-x ↗
4. A new device for performing reference point indentation without a reference probe10.1063/1.3693085 ↗
5. Alterations to the en-bloc basic fuchsin staining protocol for the demonstration of microdamage produced in-vivo10.1016/S8756-3282(95)00241-3 ↗
6. Aging of bone tissue - mechanical-properties10.2106/00004623-197658010-00015 ↗
7. Microindentation for in vivo measurement of bone tissue mechanical properties in humans10.1002/jbmr.73 ↗
8. In Vivo Assessment of Bone Quality in Postmenopausal Women With Type 2 Diabetes10.1002/jbmr.2106 ↗
9. Gallant · Pyridoxamine and Sorbinil Do Not Prevent Bone Mechanical and Tissue Deterioration in a Rat Model of Type 2 Diabetes. in ASBMR · 2011
10. Reference-point indentation correlates with bone toughness assessed using whole-bone traditional mechanical testing10.1016/j.bone.2012.12.015 ↗
11. Insights into reference point indentation involving human cortical bone: sensitivity to tissue anisotropy and mechanical behavior10.1016/j.jmbbm.2014.05.016 ↗
12. Microindentation for In Vivo Measurement of Bone Tissue Material Properties in Atypical Femoral Fracture Patients and Controls10.1002/jbmr.1731 ↗
13. Nanoscale changes in collagen are reflected in physical and mechanical properties of bone at the microscale in diabetic rats. Bone
14. The bone diagnostic instrument II: indentation distance increase10.1063/1.2937199 ↗
15. Bone diagnostic instrument10.1063/1.2221506 ↗
16. A biomechanical perspective on bone quality10.1016/j.bone.2006.06.001 ↗
17. Differentiating human bone from animal bone: A review of histological methods10.1111/j.1556-4029.2006.00368.x ↗
18. The Effects of Freezing on the Mechanical Properties of Bone10.2174/1876525401204010014 ↗
19. The true toughness of human cortical bone measured with realistically short cracks10.1038/nmat2221 ↗
20. Fluorescence-aided detection of microdamage in compact bone10.1046/j.1469-7580.1998.19320179.x ↗
21. Variability and anisotropy of mechanical behavior of cortical bone in tension and compression10.1016/j.jmbbm.2013.02.021 ↗
22. Bending and fracture of compact circumferential and osteonal lamellar bone of the baboon tibia10.1023/A:1008989719560 ↗
23. A comparative study of young and mature bovine cortical bone10.1016/j.actbio.2012.08.040 ↗
24. Nano-structural, compositional and micro-architectural signs of cortical bone fragility at the superolateral femoral neck in elderly hip fracture patients vs. healthy aged controls10.1016/j.exger.2014.03.001 ↗
25. Histological Comparison of Long-Bone Cortex between 11-Year-Old Giant Cow with Dermal Dysplasia and the Child Cow Aged 8.5 Years10.1292/jvms.10-0325 ↗
26. Role of microstructure in the aging-related deterioration of the toughness of human cortical bone10.1016/j.msec.2005.08.021 ↗
27. The bone diagnostic instrument III: Testing mouse femora10.1063/1.3147383 ↗
28. Applications of a New Handheld Reference Point Indentation Instrument Measuring Bone Material Strength
29. Reference point indentation study of age-related changes in porcine femoral cortical bone10.1016/j.jbiomech.2013.04.003 ↗
30. Fracture incidence and association with bone mineral density in elderly men and women: the Rotterdam Study10.1016/j.bone.2003.10.001 ↗
31. Towards a standardized reference point indentation testing procedure10.1016/j.jmbbm.2014.01.012 ↗
32. Caspase-2 Maintains Bone Homeostasis by Inducing Apoptosis of Oxidatively-Damaged Osteoclasts10.1371/journal.pone.0093696 ↗
33. Bone mineral density thresholds for pharmacological intervention to prevent fractures10.1001/archinte.164.10.1108 ↗
34. Modeling and remodeling in a developing artiodactyl calcaneus: A model for evaluating Frost׳s Mechanostat hypothesis and its corollaries10.1002/ar.1094 ↗
35. The Effect of NaF In Vitro on the Mechanical and Material Properties of Trabecular and Cortical Bone10.1002/adma.200801204 ↗
36. Lamellar bone: Structure-function relations10.1006/jsbi.1999.4107 ↗
37. Fracture toughness of manatee rib and bovine femur using a chevron-notched beam test10.1016/j.jbiomech.2005.02.016 ↗