Understanding the differences in wear testing method standards for total knee replacement

A. Abdelgaied, J. Fisher, L.M. Jennings

Abstract

Preclinical evaluation of the wear of total knee replacements (TKR) is usually undertaken using International Standards Organization (ISO) test methods. Two international standards for the preclinical wear simulation of TKRs have been developed; using either force or displacement control. In addition, based on previously published measured kinematics of healthy subjects, a gait cycle (displacement control) was also developed at the University of Leeds, which pre-dates the ISO displacement control standard. Furthermore, different test methods have adopted different approaches to defining the centres of rotation and polarity (direction of application) of motions. However, the effects of using these different control regimes and input conditions on the kinematics, contact mechanics, and wear of any one TKR have not been fully investigated previously. The current study investigated the kinematics, contact mechanics, and wear performance of a TKR when running under ISO force and displacement control test methods as well as the Leeds gait cycle inputs using experimental and computational simulation methods, with the aim of understanding the mechanical and tribological outcomes predicted by the different test method standard conditions. Three ISO wear testing standards were investigated using a mid-size Sigma curved TKR (DePuy, UK), with moderately cross-linked UHMWPE curved inserts; ISO-14243-3-2004, ISO-14243-3-2014 and ISO-14243-1-2009. In addition, the Leeds displacement control gait cycle was also investigated. According to the computational simulation predictions, reversing the anterior-posterior (AP) displacement and tibial rotation polarities in the displacement control ISO-2014 standard compared to the ISO-2004 standard resulted in high stress, of more than 65 MPa, at the posterior edge of the inserts with more than 10% increase in wear rate for this TKR design. Although Leeds gait input kinematics produced femoral rollback, it did not result in high stress edge loading on the posterior lip of the insert. This was attributed to different test input kinematics and different centres of rotation of the femoral component adopted in the displacement control standard ISO-2014 and Leeds gait test methods. The predicted AP displacement and tibial rotation from the force control ISO-2009 had different polarities and magnitudes to the corresponding displacement control profiles. In addition, the predicted wear rate, from the computational model, under the force control ISO-2009 standard was more than double that predicted under displacement control ISO standards due to the increased AP displacement and tibial rotation motions predicted under the force control standard. These major differences, in the mechanics and wear, between different test methods imply that each standard must therefore be used with its own predicate control results from a device with proven clinical history and results across different standards should never be compared, as the choice of test method standard may well be dependent on the design solution for the knee. Clinically, the kinematics in the population are extremely variable, which results in highly variable wear rates. While a standard method is necessary, on its own it is not adequate and needs to be supported by tests under a portfolio of representative conditions with different kinematic conditions, different soft tissue constraints, as well as with different alignments, so that the variability and range of wear rates expected clinically might be determined. This study enables further progress towards the definition of such a portfolio of representative conditions, by deepening the understanding of the relationships between currently used input conditions and the resulting mechanical and wear outputs.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Total Knee Arthroplasty Outcomes · Orthopaedic implants and arthroplasty · Tribology and Wear Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (30)

Show source-provided references
1. The effect of insert conformity and material on total knee replacement wear10.1177/0954411913513251 ↗
2. A comprehensive combined experimental and computational framework for pre-clinical wear simulation of total knee replacements10.1016/j.jmbbm.2017.11.022 ↗
3. A comparison between electromechanical and pneumatic-controlled knee simulators for the investigation of wear of total knee replacements10.1177/0954411917696519 ↗
4. Computational wear prediction of artificial knee joints based on a new wear law and formulation10.1016/j.jbiomech.2011.01.027 ↗
5. Abdelgaied · 2022
6. Dose effects of cross-linking polyethylene for total knee arthroplasty on wear performance and mechanical properties10.1002/jbm.b.30835 ↗
7. Comparison of wear in a total knee replacement under different kinematic conditions
8. The influence of simulator input conditions on the wear of total knee replacements: an experimental and computational study10.1177/0954411916645134 ↗
9. 2009 knee society presidential guest lecture: polyethylene wear in total knees10.1007/s11999-009-1033-1 ↗
10. Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses10.1016/j.jbiomech.2009.05.010 ↗
11. 2009
12. 2002
13. 2014
14. 2004
15. The influence of femoral condylar lift-off on the wear of artificial knee joints10.1243/09544119JEIM215 ↗
16. Johnson · 2000 · 46 · 30
17. The effect of surgical alignment and soft tissue conditions on the kinematics and wear of a fixed bearing total knee replacement10.1016/j.jmbbm.2019.103386 ↗
18. Representing the effect of variation in soft tissue constraints in experimental simulation of total knee replacements10.1016/j.jmbbm.2018.07.011 ↗
19. Three-dimensional kinematics of the human knee during walking10.1016/0021-9290(92)90254-X ↗
20. The influence of design, materials and kinematics on the in vitro wear of total knee replacements10.1016/j.jbiomech.2004.02.015 ↗
21. The mechanics of the knee joint in relation to normal walking10.1016/0021-9290(70)90050-3 ↗
22. 2020
23. 2014
24. Force actions transmitted by joints in the human body10.1098/rspb.1976.0004 ↗
25. The effect of walking speed on the force actions transmitted at the hip and knee joints
26. Forces transmitted at the hip and knee joint of normal and disabled persons during a range of activities
27. A computer model with surface friction for the prediction of total knee kinematics10.1016/S0021-9290(96)00114-5 ↗
28. In vitro response of the natural cadaver knee to the loading profiles specified in a standard for knee implant wear testing10.1016/j.jbiomech.2010.03.042 ↗
29. Knee implant kinematics are task-dependent10.1098/rsif.2018.0678 ↗
30. A knee simulating machine for performance evaluation of total knee replacements10.1016/S0021-9290(96)00118-2 ↗