Finite element analysis of a low modulus Ti-20Zr-3Mo-3Sn alloy designed to reduce the stress shielding effect of a hip prosthesis
Source abstract
Abstract
After total hip arthroplasty, the stress shielding effect can occur due to the difference of stiffness between the metallic alloy of the stems and the host bone, which may cause a proximal bone loss. To overcome this problem, a low-modulus metastable β Ti-20Zr-3Mo-3Sn alloy composition has recently been designed to be potentially used for the cementless femoral hip stems. After having verified experimentally that the β alloy has a low modulus of around 50 GPa, a finite element analysis was performed on a Ti-20Zr-3Mo-3Sn alloy hip prosthesis model to evaluate the influence of a reduced modulus on stress shielding and stress fields in both stem and bone compared with the medical grade Ti-6Al-4V alloy whose elastic modulus reached 110 GPa. Our results show that the Ti-20Zr-3Mo-3Sn stem with low elastic modulus can effectively reduce the total stress shielding by 45.5% compared to the common Ti-6Al-4V prosthesis. Moreover, it is highlighted that the material elasticity affects the stress distribution in the implant, especially near the bone-stem interfaces.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Advanced materials and composites
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (49)
Show source-provided references
1. Finite element analysis of porous titanium alloy hip stem to evaluate the biomechanical performance during walking and stair climbing10.1007/s42235-019-0122-4 ↗
2. Fully porous 3D printed titanium femoral stem to reduce stress-shielding following total hip arthroplasty: fully porous 3D printed titanium femoral stem10.1002/jor.23445 ↗
3. TiNbSn stems with gradient changes of Young’s modulus and stiffness reduce stress shielding compared to the standard fit-and-fill stems10.1186/s40001-023-01199-z ↗
4. An approach for time-dependent bone modeling and remodeling-theoretical development: time-dependent modeling and remodeling10.1002/jor.1100080506 ↗
5. Finite element analysis of hip stem designs10.1016/j.matdes.2006.12.014 ↗
6. Bone remodeling around cementless hip implants: South
7. A patterned microtexture to reduce friction and increase longevity of prosthetic hip joints10.1016/j.wear.2014.04.001 ↗
8. Additive manufacturing structural redesign of hip prostheses for stress-shielding reduction and improved functionality and safety10.1016/j.mechmat.2021.104173 ↗
9. Strain distribution in the proximal human femoral metaphysis10.1243/09544119JEIM497 ↗
10. Analysis of cementless implants using interface nonlinear friction - experimental and finite element studies10.1016/S0021-9290(96)00110-8 ↗
11. Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results10.1302/0301-620X.69B1.3818732 ↗
12. A contact model with ingrowth control for bone remodelling around cementless stems10.1016/S0021-9290(01)00204-4 ↗
13. Topological optimization in hip prosthesis design10.1007/s10237-009-0183-0 ↗
14. Influence of texture and transformation strain on the superelastic performance of a new Ti–20Zr–3Mo–3Sn alloy10.1016/j.msea.2019.138075 ↗
15. Periprosthetic bone remodelling of a collum femoris preserving cementless titanium femoral hip replacement10.1080/10255840601003577 ↗
16. Experimental validation of finite element model for proximal composite femur using optical measurements10.1016/j.jmbbm.2013.02.006 ↗
17. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening
18. Fabrication of a high-performance hip prosthetic stem using β Ti–33.6Nb–4Sn10.1016/j.jmbbm.2013.11.002 ↗
19. Micromotion and friction evaluation of a novel surface architecture for improved primary fixation of cementless orthopaedic implants10.1016/j.jmbbm.2013.01.017 ↗
20. Titanium as the material of choice for cementless femoral components in total hip arthroplasty
21. Mathematical relationships between bone density and mechanical properties: a literature review10.1016/j.clinbiomech.2007.08.024 ↗
22. Effects of mechanical forces on maintenance and adaptation of form in trabecular bone10.1038/35015116 ↗
23. The relationship between stress shielding and bone resorption around total hip stems and the effects of flexible materials10.1097/00003086-199201000-00014 ↗
24. In vivo skeletal responses to porous-surfaced implants subjected to small induced motions10.2106/00004623-199705000-00010 ↗
25. Development and in vitro validation of a simplified numerical model for the design of a biomimetic femoral stem10.1016/j.jmbbm.2017.10.019 ↗
26. Analysis of a femoral hip prosthesis designed to reduce stress shielding10.1016/S0021-9290(00)00110-X ↗
27. Evaluation of a femoral stem with reduced stiffness. A randomized study with use of radiostereometry and bone densitometry10.2106/00004623-200209000-00020 ↗
28. Friction and stem stiffness affect dynamic interface motion in total hip replacement10.1002/jor.1100140108 ↗
29. The theory of early loosening of hip prostheses10.3928/0147-7447-19971201-12 ↗
30. Bone mechanical properties in healthy and diseased states10.1146/annurev-bioeng-062117-121139 ↗
31. LibHip: an open-access hip joint model repository suitable for finite element method simulation10.1016/j.cmpb.2022.107140 ↗
32. A novel hybrid design and modelling of a customised graded Ti-6Al-4V porous hip implant to reduce stress-shielding: an experimental and numerical analysis10.3389/fbioe.2023.1092361 ↗
33. Observations on the effect of movement on bone ingrowth into porous-surfaced implants10.1097/00003086-198607000-00023 ↗
34. The influence of contact ratio and its location on the primary stability of cementless total hip arthroplasty: a finite element analysis10.1016/j.jbiomech.2016.02.031 ↗
35. Young’s modulus of trabecular and cortical bone material: ultrasonic and microtensile measurements10.1016/0021-9290(93)90042-D ↗
36. Long-term bone remodelling around ‘legendary’ cementless femoral stems10.1302/2058-5241.3.170024 ↗
37. Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants10.1002/jbm.820270205 ↗
38. Biomechanical optimization of elastic modulus distribution in porous femoral stem for artificial hip joints10.1007/s42235-018-0057-1 ↗
39. The influence of clinical factors on periprosthetic bone remodeling10.1097/00003086-199601000-00034 ↗
40. Primary cementless total hip arthroplasty using a modular femoral component: a minimum 6-year follow-up10.1054/arth.2001.29140 ↗
41. Improving peri-prosthetic bone adaptation around cementless hip stems: a clinical and finite element study10.1016/j.medengphy.2013.12.006 ↗
42. A review of current fixation use and registry outcomes in total hip arthroplasty: the uncemented paradox10.1007/s11999-013-2941-7 ↗
43. The mechanism of bone remodeling and resorption around press-fitted THA stems10.1016/0021-9290(93)90001-U ↗
44. Primary stability of an anatomical cementless hip stem: a statistical analysis10.1016/j.jbiomech.2005.03.024 ↗
45. Effects of material properties of femoral hip components on bone remodeling10.1002/jor.1100100614 ↗
46. Sensitivity of periprosthetic stress-shielding to load and the bone density–modulus relationship in subject-specific finite element models10.1016/S0021-9290(00)00036-1 ↗
47. In-vitro biomechanical evaluation of stress shielding and initial stability of a low-modulus hip stem made of β type Ti-33.6Nb-4Sn alloy10.1016/j.medengphy.2014.09.002 ↗
48. Improving stress shielding following total hip arthroplasty by using a femoral stem made of β type Ti-33.6Nb-4Sn with a Young’s modulus gradation10.1016/j.jbiomech.2017.08.017 ↗
49. Interfacial frictional behavior: cancellous bone, cortical bone, and a novel porous tantalum biomaterial10.1142/S0218957799000269 ↗