Effect of bioprosthetic leaflet anisotropy on stent dynamics of Transcatheter Aortic Valve Replacement devices
Source abstract
Abstract
The assessment of stent fatigue in Transcatheter Aortic Valve Replacement (TAVR) systems is critical for the design of next-generation devices, both in vitro and in vivo. The mechanical properties of the bioprosthetic heart valves (BHVs) have a significant impact on the fatigue life of the metallic stent and thus must be taken into consideration when evaluating new TAVR device designs. This study aims to investigate the relationship between BHV anisotropic behaviour and the asymmetric deflections of the stent frame observed during in vitro testing. An explicit dynamics finite element model of the nitinol stent with attached bioprosthetic valve leaflets was developed to evaluate the deflections of the TAVR device under haemodynamic loading. Our results demonstrate that pericardium behaviour plays a dominant role in determining stent frame deflection. The anisotropic behaviour of the leaflets, resulting from collagen fibre orientation, affects the extent of deflection encountered by each commissure of the frame. This leads to asymmetric variation in frame deflection that can influence the overall fatigue life of the nitinol stent. This study highlights the importance of considering both the flexible nature of the metallic stent as well as the leaflet anisotropic behaviour in the design and fatigue assessment of TAVR systems.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Cardiac Valve Diseases and Treatments · Coronary Interventions and Diagnostics · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (51)
Show source-provided references
1. Mechanical testing of pericardium for manufacturing prosthetic heart valves10.1093/icvts/ivv282 ↗
2. Simulation of transcatheter aortic valve implantation: A patient-specific finite element approach10.1080/10255842.2012.746676 ↗
3. Shape-memory alloys: macromodelling and numerical simulations of the superelastic behavior10.1016/S0045-7825(96)01232-7 ↗
4. Assessing the impact of including leaflets in the simulation of TAVI deployment into a patient-specific aortic root10.1080/10255842.2015.1058928 ↗
5. The exponential law of endurance tests
6. In silico fatigue optimization of TAVR stent designs with physiological motion in a beating heart model10.1016/j.cmpb.2023.107886 ↗
7. Long-term mechanical fatigue response of porcine bioprosthetic heart valves
8. Fluid-structure interaction of heart valve dynamics in comparison to finite-element analysis10.1515/cdbme-2018-0063 ↗
9. Evaluation of transcatheter heart valve biomaterials: Biomechanical characterization of bovine and porcine pericardium10.1016/j.jmbbm.2017.08.013 ↗
10. The influence of mean strain on the high-cycle fatigue of nitinol with application to medical devices10.1016/j.jmps.2020.104057 ↗
11. Radial force: An underestimated parameter in oversizing transcatheter aortic valve replacement prostheses: In vitro analysis with five commercialized valves10.1097/MAT.0000000000000659 ↗
12. Simulation of planar soft tissues using a structural constitutive model: Finite element implementation and validation10.1016/j.jbiomech.2014.03.014 ↗
13. Finite element analysis of TAVI: Impact of native aortic root computational modeling strategies on simulation outcomes10.1016/j.medengphy.2017.06.045 ↗
14. Hyperelastic modelling of arterial layers with distributed collagen fibre orientations10.1098/rsif.2005.0073 ↗
15. Mass-spring model for simulation of heart valve tissue mechanical behavior10.1007/s10439-011-0278-5 ↗
16. A new constitutive framework for arterial wall mechanics and a comparative study of material models10.1023/A:1010835316564 ↗
17. Dynamic and fluid-structure interaction simulations of bioprosthetic heart valves using parametric design with T-splines and fung-type material models10.1007/s00466-015-1166-x ↗
18. TAVR and SAVR: Current treatment of aortic stenosis
19. Stress-strain behavior of the passive basilar artery in normotension and hypertension10.1016/j.jbiomech.2006.11.007 ↗
20. Fatigue analysis of novel aortic valve stent using finite element analysis10.59324/ejmhr.2023.1(2).04 ↗
21. Dynamic simulation pericardial bioprosthetic heart valve function10.1115/1.2244578 ↗
22. Dynamic simulation of bioprosthetic heart valves using a stress resultant shell model10.1007/s10439-007-9409-4 ↗
23. Numerical simulation of leaflet flexure in bioprosthetic valves mounted on rigid and expansile stents10.1016/0021-9290(93)90055-J ↗
24. Labs · 2023
25. Finite element analysis for fatigue behaviour of a self-expanding nitinol peripheral stent under physiological biomechanical conditions10.1016/j.compbiomed.2018.11.019 ↗
26. Simulated thin pericardial bioprosthetic valve leaflet deformation under static pressure-only loading conditions: implications for percutaneous valves10.1007/s10439-010-0009-3 ↗
27. Fatigue of nitinol: The state-of-the-art and ongoing challenges10.1016/j.jmbbm.2015.06.010 ↗
28. Thermal stress and low cycle fatigue10.1115/1.3625225 ↗
29. Fatigue damage of collagenous tissues: experiment, modeling and simulation studies10.1615/JLongTermEffMedImplants.2015011749 ↗
30. Life expectancy after surgical aortic valve replacement10.1016/j.jacc.2021.09.861 ↗
31. Effects of cyclic flexural fatigue on porcine bioprosthetic heart valve heterograft biomaterials10.1002/jbm.a.32659 ↗
32. Fatigue analysis of shape memory alloys: energy approach10.1088/0964-1726/14/5/017 ↗
33. Durability of nitinol cardiovascular devices10.1007/s40830-022-00370-5 ↗
34. Modeling of transcatheter aortic valve replacement: Patient specific vs general approaches based on finite element analysis10.1016/j.compbiomed.2015.12.001 ↗
35. The incidence of arterial stent fractures with exclusion of coronary, aortic, and non-arterial settings10.1016/j.ejvs.2008.05.005 ↗
36. Quantification of uncertainty in computational fluid dynamics10.1146/annurev.fluid.29.1.123 ↗
37. Principles of TAVR valve design, modelling, and testing10.1080/17434440.2018.1536427 ↗
38. Orthotropic mechanical properties of chemically treated bovine pericardium10.1114/1.135 ↗
39. Collagen fiber architecture of bovine pericardium10.1097/00002480-199407000-00075 ↗
40. Heart valve function: A biomechanical perspective10.1098/rstb.2007.2122 ↗
41. SIMULIA · 2021
42. Biomechanical behavior of bioprosthetic heart valve heterograft tissues: Characterization, simulation, and performance10.1007/s13239-016-0276-8 ↗
43. Evaluation of transcatheter heart valve biomaterials: Computational modeling using bovine and porcine pericardium10.1016/j.jmbbm.2019.05.020 ↗
44. Simulated bioprosthetic heart valve deformation under quasi-static loading10.1115/1.2049337 ↗
45. Collagen fibre orientation and dispersion govern ultimate tensile strength, stiffness and the fatigue performance of bovine pericardium10.1016/j.jmbbm.2018.09.038 ↗
46. Investigation into early stage fatigue-damage accumulation in glutaraldehyde-fixed bovine pericardium using a novel equibiaxial bulge inflation system10.1016/j.jmbbm.2021.104588 ↗
47. Bovine pericardium of high fibre dispersion has high fatigue life and increased collagen content; potentially an untapped source of heart valve leaflet tissue10.1007/s10439-020-02644-4 ↗
48. Fluid-structure interaction model of a percutaneous aortic valve: Comparison with an in vitro test and feasibility study in a patient-specific case10.1007/s10439-015-1429-x ↗
49. Stent and leaflet stresses in a 26-mm first-generation balloon-expandable transcatheter aortic valve10.1016/j.jtcvs.2016.12.016 ↗
50. Simulating the time evolving geometry, mechanical properties, and fibrous structure of bioprosthetic heart valve leaflets under cyclic loading10.1016/j.jmbbm.2021.104745 ↗
51. A material modeling approach for the effective response of planar soft tissues for efficient computational simulations10.1016/j.jmbbm.2018.09.016 ↗