journal article·2023-08Open access

Metal debris release is commonly seen from explanted total ankle replacements

Shona Haston, David Langton, David Townshend, Rohan Bhalekar, Thomas Joyce

Abstract

This study aimed to characterise the damage mechanisms present on the metal components used in various contemporary total ankle replacements. Twenty-seven explanted total ankle replacements comprising 8 different designs (3 fixed bearing and 5 mobile bearing) were analysed using various explant analysis techniques. Pitting and scratching were the most commonly observed wear features. Microscopic analysis revealed metallic pitting on 52% of tibial components and 95% of talar components. Pitting was identified on more cobalt-chromium than titanium alloy tibial components (63% versus 0%). Non-contact profilometry confirmed the presence of pitting, with significant (p < 0.05) differences in the measured average surface roughness values of pitted and unpitted areas for tibial and talar components. There was macroscopically visible sliding plane scratching, indicating the presence of hard third body particles, on 78% of talar components. Changes to the non-articulating surfaces coatings in terms of coating loss and/or changes in reflectivity was identified visually on 80% of metal components. Scanning electron microscopy with energy dispersive X-ray spectroscopy identified metallic embedded debris in 19% of polyethylene inserts. This explant study demonstrates the release of metal debris from both the metallic tibial and talar component articulating surfaces and non-articulating surface coatings of various contemporary total ankle replacements. Metal particulate debris release from total ankle replacements may be more common than previously recognised. Metal debris should be considered in further study into the aetiology of failed total ankle arthroplasty.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Surgery and Rehabilitation · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (26)

Show source-provided references
1. Wear behaviour in total ankle replacement: a comparison between an in vitro simulation and retrieved prostheses10.1016/j.clinbiomech.2009.06.006 ↗
2. Demographics and outcome of ankle arthroplasty: supplementary report
3. Wear study of total ankle replacement explants by microstructural analysis10.1016/j.jmbbm.2015.12.035 ↗
4. Analysis of failed ankle arthroplasty components10.1177/1071100718802589 ↗
5. Histologic study of periprosthetic osteolytic lesions after AES total ankle replacement. A 22 case series10.1016/j.otsr.2013.07.009 ↗
6. The AES total ankle arthroplasty analysis of failures and survivorship at ten years10.1007/s00264-017-3605-0 ↗
7. Furnes · 2021
8. History of total ankle replacement in north America
9. Greenwald
10. Polyethylene insert damage in unicondylar knee replacement: a comparison of in vivo function and in vitro simulation10.1243/09544119JEIM699 ↗
11. Damage patterns in polyethylene fixed bearings of retrieved total ankle replacements10.1016/j.fas.2020.06.012 ↗
12. Severe periprosthetic osteolytic lesions after the Ankle Evolutive System total ankle replacement10.1302/0301-620X.91B7.22434 ↗
13. The influence of hla genotype on the development of metal hypersensitivity following joint replacement
14. Biomechanics of the ankle and foot
15. McKie · 2022
16. Pathogenesis, evaluation, and management of osteolysis following total ankle arthroplasty10.1177/1071100720978426 ↗
17. Why does titanium alloy wear cobalt chrome alloy despite lower bulk hardness: a nanoindentation study?10.1016/j.tsf.2013.06.020 ↗
18. Titanium nitride ceramic film against polyethylene. A 48 million cycle wear test
19. Cobalt-based orthopaedic alloys: relationship between forming route, microstructure and tribological performance10.1016/j.msec.2012.03.012 ↗
20. Survival of primary ankle replacements: data from global joint registries10.1186/s13047-022-00539-2 ↗
21. Reed · 2022
22. Histological analysis of early osteolysis in total ankle arthroplasty10.1177/1071100716682333 ↗
23. Isolation and characterisation of wear debris surrounding failed total ankle replacements10.1016/j.actbio.2023.01.051 ↗
24. Better implant survival with modern ankle prosthetic designs: 1,226 total ankle prostheses followed for up to 20 years in the Swedish Ankle Registry10.1080/17453674.2019.1709312 ↗
25. Etiopathogenesis of osteolytic cysts associated with total ankle arthroplasty, a histological study10.1016/j.fas.2015.02.004 ↗
26. Analysis of retrieved Aglity™ total ankle arthroplasty systems10.3113/FAI.2009.0815 ↗