Taper corrosion in total hip arthroplasty – How to assess and which design features are crucial?
Source abstract
Abstract
Background In total hip replacement, wear and corrosion arising from modular taper connections have increasingly become a serious clinical concern. Previous studies led to confounding results regarding the role of specific taper design features, likely due to the application of different analytical approaches. Accordingly, this study has two major objectives: first, to evaluate different analytical approaches to evaluate the fretting-corrosion behavior; and secondly to determine the effect of four specific design variables: the taper engagement situation, the stem taper length and surface topography in terms of roughness and contact ratio. Methods An in vitro fretting-corrosion test setup was used including an aggressive solution. Cyclic loading was applied, varying from 300 N to 2500 N at a frequency of 3 Hz. Taper dummies covering different implant designs were tested in seven different test groups. Different quantitative and qualitative analytical test methods such as electrochemical characterization, ion analysis, gravimetry and corrosion scoring were applied in order to quantify the material degradation. Results A stepwise linear regression analysis showed that the taper engagement situation is the predominant factor that predicts the metal ion release from the taper connection, followed by the contact ratio of the taper surface and subsequently the taper length. A distal taper engagement situation, as well as a high contact ratio and a short taper length are the relevant parameters that decrease the metal ion release. Hereby, metal ion analysis turned out to be the most precise and reliable method for determining corrosive driven material loss, followed by gravimetry. Conclusion It was found that the taper engagement length is the major design parameter that influences the total ion release. It further turned out, that the selection of an appropriate analytical approach is essential for the evaluation of the corrosion behavior of taper connections in an experimental setting.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (56)
Show source-provided references
1. Microgrooved surface topography does not influence fretting corrosion of tapers in total hip arthroplasty: classification and retrieval analysis
2. Do stem taper microgrooves influence taper corrosion in total hip arthroplasty? A matched cohort retrieval study10.1016/j.arth.2016.11.018 ↗
3. Does a micro-grooved trunnion stem surface finish improve fixation and reduce fretting wear at the taper junction of total hip replacements? A finite element evaluation10.1016/j.jbiomech.2017.07.027 ↗
4. Biomechanical evaluation of the fatigue performance, the taper corrosion and the metal ion release of a dual taper hip prosthesis under physiological environmental conditions10.1016/j.biotri.2017.07.002 ↗
5. Effect of hydrogen peroxide on titanium surfaces: in situ imaging and step-polarization impedance spectroscopy of commercially pure titanium and titanium, 6-aluminum, 4-vanadium10.1002/jbm.a.10116 ↗
6. Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers10.1016/j.jbiomech.2020.109689 ↗
7. A crevice corrosion model for biomedical trunnion geometries and surfaces feature10.3390/ma14041005 ↗
8. Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings10.1002/jor.22326 ↗
9. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis10.1302/0301-620X.93B5.26309 ↗
10. Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system10.1002/jor.22970 ↗
11. Fretting corrosion accelerates crevice corrosion of modular hip tapers10.1002/jab.770060104 ↗
12. Corrosion and fretting of a modular hip system: a retrieval analysis of 60 rejuvenate stems10.1016/j.arth.2015.03.010 ↗
13. Effect of surface roughness of self-mated alumina on friction and wear in isooctane-lubricated reciprocating sliding contact10.1016/S0043-1648(01)00900-0 ↗
14. Fretting and corrosion in modular shoulder arthroplasty: a retrieval analysis10.1155/2016/1695906 ↗
15. What is the trouble with trunnions?10.1007/s11999-014-3746-z ↗
16. Micromotion at the head-stem taper junction of total hip prostheses is influenced by prosthesis design-, patient- and surgeon-related factors10.1016/j.jbiomech.2019.109424 ↗
17. The influence of Molybdenum on the fretting corrosion behavior of CoCr/TiAlV couples10.1016/j.biotri.2017.01.001 ↗
18. The use of fretting currents to examine the issues of tolerances in modular hips
19. Quantification of fretting damage10.1016/S0043-1648(96)07306-1 ↗
20. Medical implant corrosion: electrochemistry at metallic biomaterial surfaces
21. Direct in vivo inflammatory cell-induced corrosion of CoCrMo alloy orthopedic implant surfaces10.1002/jbm.a.35165 ↗
22. In vitro corrosion testing of modular hip tapers10.1002/jbm.b.10526 ↗
23. A multicenter retrieval study of the taper interfaces of modular hip prostheses10.1097/00003086-200208000-00018 ↗
24. Biotribology of alternative bearing materials for unicompartmental knee arthroplasty10.1016/j.actbio.2010.04.003 ↗
25. Is increased modularity associated with increased fretting and corrosion damage in metal-on-metal total hip arthroplasty devices?: a retrieval study10.1016/j.arth.2013.05.040 ↗
26. Method of characterizing fretting and corrosion at the various taper connections of retrieved modular components from metal-on-metal total hip arthroplasty
27. The reliability of a scoring system for corrosion and fretting, and its relationship to material loss of tapered, modular junctions of retrieved hip implants10.1016/j.arth.2013.12.003 ↗
28. Influence of stem type on material loss at the metal-on-metal pinnacle taper junction10.1177/0954411914567931 ↗
29. Corrosion of metal orthopaedic implants10.2106/00004623-199802000-00015 ↗
30. Flexural rigidity, taper angle, and contact length Affect fretting of the femoral stem trunnion in total hip arthroplasty10.1016/j.arth.2016.02.079 ↗
31. Validation of ultratrace analysis of Co, Cr, Mo and Ni in whole blood, serum and urine using ICP-SMS10.1039/b821913c ↗
32. Metal release and corrosion effects of modular neck total hip arthroplasty10.1007/s00264-009-0729-x ↗
33. Wear analysis of unicondylar mobile bearing and fixed bearing knee systems: a knee simulator study10.1016/j.actbio.2010.09.031 ↗
34. Determination of low wear rates in metal-on-metal hip joint replacements based on ultra trace element analysis in simulator studies10.1007/s11249-009-9486-7 ↗
35. Do ceramic femoral heads reduce taper fretting corrosion in hip arthroplasty? A retrieval study10.1007/s11999-013-3096-2 ↗
36. Taper junction failure in large-diameter metal-on-metal bearings10.1302/2046-3758.14.2000047 ↗
37. Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements10.1002/jor.22431 ↗
38. Are damage modes related to microstructure and material loss in severely damaged CoCrMo femoral heads?10.1097/CORR.0000000000001819 ↗
39. The influence of design, materials and kinematics on the in vitro wear of total knee replacements10.1016/j.jbiomech.2004.02.015 ↗
40. Effect of impact assembly on the fretting corrosion of modular hip tapers10.1002/jor.20048 ↗
41. Taper junctions in modular hip joint replacements: what affects their stability?10.1016/j.jmbbm.2020.104258 ↗
42. Same same but different? 12/14 stem and head tapers in total hip arthroplasty10.1016/j.arth.2017.04.027 ↗
43. Metrology to quantify wear and creep of polyethylene tibial knee inserts10.1097/01.blo.0000063604.67412.04 ↗
44. Accelerated fretting corrosion testing modular necks for total hip arthroplasty
45. Taper design affects failure of large-head metal-on-metal total hip replacements10.1007/s11999-013-3115-3 ↗
46. Design, material, and seating load effects on in vitro fretting corrosion performance of modular head-neck tapers10.1016/j.arth.2019.01.043 ↗
47. Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topography10.1002/jor.22461 ↗
48. The effect of frictional torque and bending moment on corrosion at the taper interface : an in vitro study10.1302/0301-620X.97B4.34800 ↗
49. Influence of assembly procedure and material combination on the strength of the taper connection at the head–neck junction of modular hip endoprostheses10.1016/j.clinbiomech.2011.08.002 ↗
50. In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip-implants10.1002/jbm.b.31171 ↗
51. Fretting-corrosion in hip taper modular junctions: the influence of topography and pH levels - an in-vitro study10.1016/j.jmbbm.2021.104443 ↗
52. Otto aufranc award: large heads do not increase damage at the head-neck taper of metal-on-polyethylene total hip arthroplasties10.1007/s11999-015-4468-6 ↗
53. Factors related to imprinting corrosion in modular head-neck junctions
54. Special modes of corrosion under physiological and simulated physiological conditions10.1016/j.actbio.2007.12.003 ↗
55. Volk · 2013
56. Variation in taper surface roughness for a single design effects the wear rate in total hip arthroplasty10.1002/jor.23456 ↗