The occurrence of squeaking under wear testing standards for ceramic on ceramic total hip replacements
Source abstract
Abstract
Ceramic on ceramic total hip replacement clinical reports may on occasion note a noise or squeaking. There is much debate on whether this is an actual concern, but some medical centres want to avoid any possible negative impact on the patients' wellbeing due to the noise generated. The aim of this study was to determine what sound frequencies can be picked up from hip testing standards for ceramic on ceramic under different lubrication conditions. The ISO-14242-1 (35° cup angle) and ISO-14242-4 (55° cup angle with a 4 mm translational mismatch) standards were used for testing with dry, water and serum lubrication conditions up to 10000 cycles. No sound was detected for water and serum conditions under standard walking (ISO-14242-1) testing. An audible noise with a frequency range of 0.4-0.8 kHz was picked up within 600 cycles under water and edge loading (ISO-14242-4) conditions. All dry testing produced a high pitch squeak when the frequency was higher than 2 kHz. One sample under dry edge loading conditions had an audible noise of 0.8 kHz, considered not as squeaking, as it was not high pitch. Dry testing for both, standard walking (ISO-14242-1) and edge loading (ISO-14242-4) conditions, which resulted in a high pitch noise, had a frequency range of 2-8 kHz and 5-9 kHz respectively. One sample tested with edge loading and serum produced a faint squeak noise after 6000 cycles with a frequency of 7 kHz. Edge loading due to ISO-14242-4 conditions had an increased torque which may be playing a role in increased friction leading to noise. Edge loading conditions were more prone to the generation of audible noise and squeaking whilst under lubricated conditions.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Hip and Femur Fractures
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (40)
Show source-provided references
1. Ceramic liner fractures presenting as squeaking after primary total hip arthroplasty10.2106/JBJS.M.00737 ↗
2. Is ceramic-on-ceramic squeaking phenomenon reproducible in vitro? A long-term simulator study under severe conditions10.1002/jbm.b.31398 ↗
3. Wear of novel ceramic-on-ceramic bearings under adverse and clinically relevant hip simulator conditions
4. Is patient satisfaction related to patient reported sounds from ceramic on ceramic total hip arthroplasty? A study of 265 hips10.1007/s00590-019-02426-2 ↗
5. Squeaking and other noises in patients with ceramic-on-ceramic total hip arthroplasty
6. High friction moments in large hard-on-hard hip replacement bearings in conditions of poor lubrication10.1002/jor.22255 ↗
7. Long-term results of Delta ceramic-on-ceramic total hip arthroplasty10.1016/j.artd.2021.11.006 ↗
8. Friction of total hip replacements with different bearings and loading conditions10.1002/jbm.b.30691 ↗
9. Squeaking hip arthroplasties: a tribological phenomenon10.1016/j.arth.2012.01.023 ↗
10. Assessment of squeaking in total hip replacements10.1053/j.sart.2011.10.002 ↗
11. Retrieval analysis of squeaking ceramic implants: are there related specific features?
12. The 2009 Frank Stinchfield Award: “Hip squeaking”: a biomechanical study of ceramic-on-ceramic bearing surfaces10.1007/s11999-009-0911-x ↗
13. A proposed mechanism for squeaking of ceramic-on-ceramic hips10.1016/j.wear.2010.08.006 ↗
14. Incidental discovery of an undisplaced ceramic liner fracture at total hip arthroplasty revision for squeaking
15. Squeaking in total hip replacement: no cause for concern10.3928/01477447-20080901-11 ↗
16. The influence of stem design on critical squeaking friction with ceramic bearings10.1002/jor.22413 ↗
17. The influence of component design, bearing clearance and axial load on the squeaking characteristics of ceramic hip articulations10.1016/j.jbiomech.2010.12.012 ↗
18. 2014
19. 2018
20. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings10.1007/s11999-008-0177-8 ↗
21. Squeaking sound in total hip arthroplasty using ceramic-on-ceramic bearing surfaces10.1007/s00776-010-0005-3 ↗
22. Advances in zirconia toughened alumina biomaterials for total joint replacement10.1016/j.jmbbm.2013.03.022 ↗
23. Nanoparticle-containing hyaluronate solution for improved lubrication of orthopedic ceramics10.3390/polym14173485 ↗
24. Effect of microseparation on contact mechanics in metal-on-metal hip replacements—a finite element analysis10.1002/jbm.b.33313 ↗
25. The effect of lubrication on the friction and wear of Biolox®delta10.1016/j.actbio.2011.12.037 ↗
26. Incidence of “Squeaking” after ceramic-on-ceramic total hip arthroplasty
27. Does Biolox Delta ceramic reduce the rate of component fractures in total hip replacement?10.1016/j.otsr.2014.05.010 ↗
28. The incidence of noise generation arising from the large-diameter Delta Motion ceramic total hip bearing10.1302/0301-620X.95B2.30450 ↗
29. An in vitro simulation model to assess the severity of edge loading and wear, due to variations in component positioning in hip joint replacements
30. The natural history of ceramic-on-ceramic prosthetic hip squeak and its impact on patients10.1007/s00590-012-1142-5 ↗
31. An estimation of the incidence of squeaking and revision surgery for squeaking in ceramic-on-ceramic total hip replacement: a meta-analysis and report from the australian , orthopaedic association national joint registry
32. A squeaky reputation: the problem may be design-dependent
33. A numerical model to reproduce squeaking of ceramic-on-ceramic total hip arthroplasty. Influence of design and material
34. Three-dimensional modeling of in vitro hip kinematics under micro-separation regime for ceramic on ceramic total hip prosthesis: an analysis of vibration and noise10.1016/j.jbiomech.2009.08.031 ↗
35. Sarrazin · 2022
36. Long-term wear of ceramic matrix composite materials for hip prostheses under severe swing phase microseparation10.1002/jbm.b.10035 ↗
37. The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings10.1016/j.arth.2007.05.038 ↗
38. Early diagnosis of ceramic liner fracture: guidelines based on a twelve-year clinical experience
39. Risk factors for ceramic liner fracture after total hip arthroplasty10.5301/HIP.2012.10339 ↗
40. Hip squeaking after ceramic-on-ceramic total hip arthroplasty10.4103/0366-6999.186654 ↗