journal article·2023-10

Mechanical behaviour of reconstructed defected skull with custom PEEK implant and Titanium fixture plates under dynamic loading conditions using FEM

Prashant Jindal, Jogendra Bharti, Vipin Gupta, S.S. Dhami

Publisher page

Abstract

Skull reconstruction using cranial implants is often required for repairing skull defects caused due to trauma, diseases, or malignancy to protect intracranial structures. For relieving Intracranial Pressure (ICP) surgeons restore cranial defects either using natural bones or fabricated custom cranial implants. With the increase in Traumatic Brain Injuries (TBI) and challenges faced by TBI patients to regain normalcy, it is imperative to analyse the mechanical behaviour of skull-implant assemblies under some Head Injury Criteria (HIC). Medical grade materials including Titanium Alloys (Ti-6Al-4V) and Polyether-ether-ketone (PEEK) are used by fabricating Patient-Specific Implants (PSI) manufactured using 3D imaging, modelling and printing techniques. 3D technologies are preferred over conventional manufacturing methods, as they enable fabrication of custom shapes, sizes and properties for these PSI. For an effective attachment of PSI with a defective skull, a stable joint and plate arrangement as fixture plates is necessary at their interface. These fixtures can have variable numbers, design shapes, materials and location arrangements. This paper presents the Finite Element Method/Analysis (FEM/FEA) study of PSI attached to a defected skull for reconstruction, with linear shaped fixture configuration, when subjected to an external dynamic loading at 5 m/s, strain rate of 10s-1 to 243s-1 and ICP of 15mm Hg from three sides of the skull faces. Three different materials as Neoprene (soft), Concrete (medium rigid) and E-Glass (highly rigid) have been used, in the form of a rectangular thin cuboidal wall structure, at an angle of 45° with the skull face. Four linear shaped fixture plates which were simplest to design, were used to attach the PSI-skull assembly, to ensure that weight of the PSI-fixation assembly on the patient remains minimal, overall assembly has symmetrical fixations and efforts required by a surgeon for fitment of these plates remain minimal. Placement of these fixture plates has been optimized to encompass the complete PSI-skull interface section, due to which the stresses within all the assembly components (PSI, fixture plate and skull) reduced by nearly 2.5 times than the initial design and remained within yielding limits, thereby, averting any failure under heavy external dynamic loading.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Automotive and Human Injury Biomechanics · Traumatic Brain Injury and Neurovascular Disturbances · Facial Trauma and Fracture Management

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (39)

Show source-provided references
1. Failure analysis of a humeral shaft locking compression plate-surface investigation and simulation by finite element method10.3390/ma12071128 ↗
2. Cranioplasty: review of materials and techniques10.4103/0976-3147.83584 ↗
3. Modelling and structural analysis of skull/cranial implant: beyond mid-line deformities
4. Effect of porous orthopaedic implant material and structure on load sharing with simulated bone ingrowth: a finite element analysis comparing titanium and PEEK10.1016/j.jmbbm.2018.01.017 ↗
5. Finite element analysis OF cranial implant
6. Estimating the global incidence of traumatic brain injury10.3171/2017.10.JNS17352 ↗
7. The strain rates of the brain and skull under dynamic loading
8. On the mechanical behaviour of PEEK and HA cranial implants under impact loading10.1016/j.jmbbm.2017.01.012 ↗
9. 3D modelling and printing of craniofacial implant template10.1108/RPJ-12-2017-0257 ↗
10. Polyether ether ketone (PEEK) and its 3D printed implants applications in medical field: an overview10.1016/j.cegh.2019.01.003 ↗
11. The strain rates in the brain, brainstem, dura, and skull under dynamic loadings
12. Finite element analysis of titanium alloy-graphene based mandible plate10.1080/10255842.2018.1555244 ↗
13. Efficiency of nanoparticle reinforcement using finite element analysis of titanium alloy mandible plate
14. Designing cranial fixture shapes and topologies for optimizing PEEK implant thickness in cranioplasty
15. Optimizing cranial implant and fixture design using different materials in cranioplasty
16. Intracranial pressure–based validation and analysis of traumatic brain injury using a new three-dimensional finite element human head model10.1177/0954411919881526 ↗
17. Effect of implant size and shape on implant success rates: a literature review10.1016/j.prosdent.2005.04.018 ↗
18. Automatic skull defect restoration and cranial implant generation for cranioplasty10.1016/j.media.2021.102171 ↗
19. Biomechanical performance of cranial implants with different thicknesses and material properties: a finite element study10.1016/j.compbiomed.2019.04.016 ↗
20. Poly-Ether-Ether-Ketone (PEEK) in orthopaedic practice- A current concept review10.1016/j.jorep.2022.03.013 ↗
21. Comparison between the different types of heterologous materials used in cranioplasty: a systematic review of the literature
22. The mechanical properties of cranial bone: the effect of loading rate and cranial sampling position10.1016/j.jbiomech.2009.05.030 ↗
23. Nahum · 1977
24. A 3D stereotactic atlas of the adult human skull base10.1186/s40708-018-0082-1 ↗
25. Evaluating different shapes of cranial fixation mini-plates using finite element method
26. A review on recent advances in the constitutive modeling of bone tissue10.1007/s11914-020-00631-1 ↗
27. A new device for bone cranial flap fixation: Technical note and surgical remarks. A multicentric experience10.25259/SNI_819_2020 ↗
28. A glass fiber-reinforced composite – bioactive glass cranioplasty implant: a case study of an early development stage implant removed due to a late infection10.1016/j.jmbbm.2015.10.030 ↗
29. Friction properties of the interface between porous-surfaced metals and tibial cancellous bone10.1002/jbm.820241107 ↗
30. Patient specific glass fiber reinforced composite versus titanium plate: a comparative biomechanical analysis under cyclic dynamic loading10.1016/j.jmbbm.2018.12.014 ↗
31. Structural and mechanical implications of PMMA implant shape and interface geometry in cranioplasty – a finite element study10.1016/j.jcms.2015.10.014 ↗
32. Materials used in cranioplasty: a history and analysis10.3171/2014.2.FOCUS13561 ↗
33. Bioactive glass-containing cranial implants: an overview10.1007/s10853-017-0888-x ↗
34. Vander Vorst · 2003
35. Biomechanics of brain and spinal-cord injury: analysis of neuropathologic and neurophysiology experiments10.1080/10286589908915739 ↗
36. Concussion in professional football: brain responses by finite element analysis: Part 910.1227/01.NEU.0000186950.54075.3B ↗
37. Patient specific implants (PSI)
38. A proposed injury threshold for mild traumatic brain injury10.1115/1.1691446 ↗
39. The application of polyetheretherketone (PEEK) implants in cranioplasty10.1016/j.brainresbull.2019.08.010 ↗