Finite element analysis of a customized implant in PMMA coupled with the cranial bone
Source abstract
Abstract
This computational study investigates the effect of the Von Misses stresses and deformations distribution generated by coupling a customized cranial implant with its fixation system for anchoring in the cranial bone of a specific patient. Three simulations were carried out under static loads, in different areas of the implant and during the rest-activity; and another three simulations were considered preset maximum intracranial pressures. Anatomical models were obtained by computed tomography. The design of the device to be implanted was carried out by applying reverse engineering processes, from the corresponding computer-aided design (CAD) model of the bone structure of interest. Likewise, the anchoring system was modeled in detail. Loads were applied at three points on the custom implant. The stress distribution on the artificial plate and the implant-natural bone interface was analyzed. The distribution of the stresses caused by the internal load states on the plate and the anchoring system was also studied. The neurocranial reconstruction with the customized polymethylmethacrylate (PMMA)-based implant and the finite element analysis demonstrated that the fixation and coupling system of the bone-implant interface guarantees adequate protection for the internal structures of the restored area. In addition, the custom-designed and placed implant will not cause non-physiological harm to the patient. Nor will failures occur in the anchoring system.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Facial Trauma and Fracture Management · Spinal Fractures and Fixation Techniques · Traumatic Brain Injury and Neurovascular Disturbances
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (44)
Show source-provided references
1. Craniofacial bone reconstruction with bioactive fiber-reinforced composite implant10.1002/hed.23370 ↗
2. Evaluation of additive manufacturing technologies for dimensional and geometric accuracy10.1504/IJMPT.2019.097665 ↗
3. Design, finite element analysis (FEA), and fabrication of custom titanium alloy cranial implant using electron beam melting additive manufacturing
4. Engineering design and manufacturing of custom craniofacial implants
5. From medical imaging data to 3D printed anatomical models10.1371/journal.pone.0178540 ↗
6. Mechanical evaluation of the large cranial implant using finite elements method
7. Monitoring and interpretation of intracranial pressure10.1136/jnnp.2003.033126 ↗
8. Mechanical characterization and numerical simulation of polyether–ether–ketone (PEEK) cranial implants10.1016/j.jmbbm.2011.05.039 ↗
9. 3D slicer as an image computing platform for the quantitative imaging network10.1016/j.mri.2012.05.001 ↗
10. Fracture of components with V-shaped notches10.1016/S0013-7944(03)00131-0 ↗
11. Patient-specific surgical implants made of 3D printed PEEK: material, technology, and scope of surgical application10.1155/2018/4520636 ↗
12. Low-cost 3-D-printer–assisted personalized cranioplasty treatment: A case series of 14 consecutive patients10.1016/j.wneu.2023.04.098 ↗
13. Materials and techniques used in cranioplasty fixation: A review10.1016/j.msec.2016.04.101 ↗
14. 3D powder printed calcium phosphate implants for reconstruction of cranial and maxillofacial defects10.1016/j.jcms.2010.01.009 ↗
15. Multi-objective optimization of a parallel manipulator for the design of a prosthetic arm using genetic algorithms10.1590/1679-78254044 ↗
16. Mandibular defect reconstruction with the help of mirror imaging coupled with laser stereolithographic modeling technique10.1016/S0929-6646(09)60247-3 ↗
17. Cranioplasty with customized titanium and PEEK implants in a mechanical stress model10.1089/neu.2011.1794 ↗
18. Comparison between two design methods implants, based on reverse engineering, design and engineering technologies, BIOCAD/CAD/CAE
19. Biomechanical performance of cranial implants with different thicknesses and material properties: A finite element study10.1016/j.compbiomed.2019.04.016 ↗
20. Description and application of a comprehensive methodology for custom implant design and surgical planning10.1016/j.inat.2022.101585 ↗
21. Surgical planning and finite element analysis for the neurocraneal protection in cranioplasty with PMMA: A case study10.1016/j.heliyon.2022.e10706 ↗
22. The mechanical properties of cranial bone: the effect of loading rate and cranial sampling position10.1016/j.jbiomech.2009.05.030 ↗
23. Mechanical biocompatibilities of titanium alloys for biomedical applications10.1016/j.jmbbm.2007.07.001 ↗
24. Analysis of temporomandibular joint prosthesis using finite element method and a patient specific design10.5267/j.esm.2018.10.001 ↗
25. Cranioplasty10.1016/j.nec.2016.11.008 ↗
26. Outcomes of cranioplasty with synthetic materials and autologous bone grafts10.1016/j.wneu.2015.01.014 ↗
27. Paediatric cranial defect reconstruction using bioactive fibre-reinforced composite implant: early outcomes10.1007/s00701-015-2363-2 ↗
28. A glass fiber-reinforced composite–bioactive glass cranioplasty implant: a case study of an early development stage implant removed due to a late infection10.1016/j.jmbbm.2015.10.030 ↗
29. Structural and mechanical implications of PMMA implant shape and interface geometry in cranioplasty–A finite element study10.1016/j.jcms.2015.10.014 ↗
30. Surgical planning of sacral nerve stimulation procedure in presence of sacral anomalies by using personalized polymeric prototypes obtained with additive manufacturing techniques10.3390/polym12030581 ↗
31. Effects of poly-ether-ether ketone (PEEK) veneer thickness on the reciprocating friction and wear behavior of PEEK/Ti6Al4V structures in artificial saliva10.1016/j.wear.2016.09.009 ↗
32. Trabecular bone fracture healing simulation with finite element analysis and fuzzy logic10.1016/j.jbiomech.2004.10.019 ↗
33. Composites and hybrid materials used for implants and bone reconstruction: a state of the art10.12988/ces.2022.91974 ↗
34. Experimental analysis of the bending resistance of the long cortical bovine bone under controlled conditions
35. A new multipartite plate system for anterior cervical spine surgery; finite element analysis10.1080/02688697.2017.1418290 ↗
36. Simulated calculation of the flexural strength of long cortical bovine bone10.12988/ces.2020.91618 ↗
37. Application of a comprehensive methodology for the development of personalized bone implants: exemplification through three cases10.1088/1742-6596/2516/1/012004 ↗
38. Design and manufacture of a customised temporomandibular prosthesis
39. A finite element model technique to determine the mechanical response of a lumbar spine segment under complex loads
40. Reconstruction of large cranial defects with poly-methyl-methacrylate (PMMA) using a rapid prototyping model and a new technique for intraoperative implant modeling10.1016/j.pjnns.2017.02.007 ↗
41. Bioactive glass-containing cranial implants: an overview10.1007/s10853-017-0888-x ↗
42. Tribological behavior of Ti-6Al-4V against cortical bone in different biolubricants10.1016/j.jmbbm.2018.10.031 ↗
43. Physical properties of the human head: mass, center of gravity and moment of inertia10.1016/j.jbiomech.2009.03.029 ↗
44. Accurate reconstruction of discontinuous mandible using a reverse engineering/computer-aided design/rapid prototyping technique: a preliminary clinical study10.1016/j.joms.2009.09.033 ↗