Quantifying the immediate post-implantation strain field of cadaveric tibiae implanted with cementless tibial trays: A time-elapsed micro-CT and digital volume correlation analysis during stair descent
Source abstract
Abstract
Primary stability, the mechanical fixation between implant and bone prior to osseointegration, is crucial for the long-term success of cementless tibial trays. However, little is known about the mechanical interplay between the implant and bone internally, as experimental studies quantifying internal strain are limited. This study employed digital volume correlation (DVC) to quantify the immediate post-implantation strain field of five cadaveric tibiae implanted with a commercially available cementless titanium tibial tray (Attune, DePuy Synthes). The tibiae were subjected to a five-step loading sequence (0-2.5 bodyweight, BW) replicating stair descent, with concomitant time-elapsed micro-CT imaging. With progressive loads, increased compression of trabecular bone was quantified, with the highest strains directly under the posterior region of the tibial tray implant, dissipating with increasing distance from the bone-implant interface. After load removal of the last load step (2.5BW), residual strains were observed in all of the five tibiae, with residual strains confined within 3.14 mm from the bone-implant interface. The residual strain is reflective of the observed initial migration of cementless tibial trays reported in clinical studies. The presence of strains above the yield strain of bone accepted in literature suggests that inelastic properties should be included within finite element models of the initial mechanical environment. This study provides a means to experimentally quantify the internal strain distribution of human tibia with cementless trays, increasing the understanding of the mechanical interaction between bone and implant.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Bone fractures and treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (63)
Show source-provided references
1. Low preoperative BMD is related to high migration of tibia components in uncemented TKA—92 patients in a combined DEXA and RSA study with 2-year follow-up10.1016/j.arth.2017.02.032 ↗
2. 2020
3. 2020
4. 2020
5. A randomised controlled trial of cemented versus cementless press-fit condylar total knee replacement: 15-year survival analysis10.1302/0301-620X.89B12.19363 ↗
6. Dynamic finite element knee simulation for evaluation of knee replacement mechanics10.1016/j.jbiomech.2011.11.052 ↗
7. Digital volume correlation: three-dimensional strain mapping using x-ray tomography10.1007/BF02323555 ↗
8. Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue10.1016/S0021-9290(03)00257-4 ↗
9. An experimental study to investigate biomechanical aspects of the initial stability of press-fit implants10.1016/j.jmbbm.2014.11.014 ↗
10. The effects of bone resection depth and malalignment on strain in the proximal tibia after total knee arthroplasty10.1016/j.arth.2009.01.021 ↗
11. The influence of bone damage on press-fit mechanics10.1016/j.jbiomech.2014.01.029 ↗
12. Glenoid bone strain after anatomical total shoulder arthroplasty: in vitro measurements with micro-CT and digital volume correlation10.1016/j.medengphy.2020.09.009 ↗
13. Comparison of cemented and uncemented fixation in total knee arthroplasty10.3928/01477447-20130426-10 ↗
14. The influence of tibial component fixation techniques on resorption of supporting bone stock after total knee replacement10.1016/j.jbiomech.2010.11.026 ↗
15. Analysis of bone–prosthesis interface micromotion for cementless tibial prosthesis fixation and the influence of loading conditions10.1016/j.jbiomech.2009.12.006 ↗
16. Finite element and experimental cortex strains of the intact and implanted tibia10.1115/1.2768382 ↗
17. Strain shielding in proximal tibia of stemmed knee prosthesis: experimental study10.1016/j.jbiomech.2007.10.006 ↗
18. Structural behaviour and strain distribution of the long bones of the human lower limbs10.1016/j.jbiomech.2009.11.022 ↗
19. About the inevitable compromise between spatial resolution and accuracy of strain measurement for bone tissue: a 3D zero-strain study10.1016/j.jbiomech.2014.07.019 ↗
20. Application of digital volume correlation to study the efficacy of prophylactic vertebral augmentation10.1016/j.clinbiomech.2016.07.010 ↗
21. Influence of trabecular bone quality and implantation direction on press-fit mechanics10.1002/jor.23257 ↗
22. Consequences of an interference fit on the fixation of porous-coated tibial components in total knee replacement10.2106/00004623-199274020-00009 ↗
23. Questionnaire on the perceptions of patients about total knee replacement10.1302/0301-620X.80B1.0800063 ↗
24. Deben · 2020
25. How flat is the tibial osteotomy in total knee arthroplasty?10.1016/j.arth.2019.10.007 ↗
26. Biomechanics and strain mapping in bone as related to immediately-loaded dental implants10.1016/j.jbiomech.2015.05.014 ↗
27. Mechanics of post‐cam engagement during simulated dynamic activity10.1002/jor.22366 ↗
28. Experimental validation of a finite element model of a human cadaveric tibia10.1115/1.2913335 ↗
29. The effects of varus tibial alignment on proximal tibial surface strain in total knee arthroplasty: the posteromedial hot spot10.1054/arth.2002.35796 ↗
30. Micromotion and migration of cementless tibial trays under functional loading conditions10.1016/j.arth.2020.07.017 ↗
31. Finite element analysis of tibial implants—effect of fixation design and friction model10.1080/10255840008915264 ↗
32. Differences between the tensile and compressive strengths of bovine tibial trabecular bone depend on modulus10.1016/0021-9290(94)90054-X ↗
33. An investigation of the inelastic behaviour of trabecular bone during the press-fit implantation of a tibial component in total knee arthroplasty10.1016/j.medengphy.2013.05.007 ↗
34. Malrotation in total knee arthroplasty: effect on tibial cortex strain captured by laser-based strain acquisition10.1016/j.clinbiomech.2006.01.011 ↗
35. The limit of tolerable micromotion for implant osseointegration: a systematic review10.1038/s41598-021-90142-5 ↗
36. A randomized controlled trial comparing two-year postoperative femoral and tibial migration of a new and an established cementless rotating platform total knee arthroplasty10.1302/0301-620X.105B2.BJJ-2022-0414.R1 ↗
37. Future young patient demand for primary and revision joint replacement: National projections from 2010 to 203010.1007/s11999-009-0834-6 ↗
38. Performance of QCT-derived scapula finite element models in predicting local displacements using digital volume correlation10.1016/j.jmbbm.2019.05.021 ↗
39. Characterization of the bone-metal implant interface by digital volume correlation of in-situ loading using neutron tomography10.1016/j.jmbbm.2017.07.001 ↗
40. Bone damage evolution around integrated metal screws using x-ray tomography—in situ pullout and digital volume correlation10.3389/fbioe.2020.00934 ↗
41. Accuracy and precision of digital volume correlation in quantifying displacements and strains in trabecular bone10.1016/j.jbiomech.2007.04.019 ↗
42. Damage tolerance and toughness of elderly human femora10.1016/j.actbio.2021.01.011 ↗
43. Time-elapsed synchrotron-light microstructural imaging of femoral neck fracture10.1016/j.jmbbm.2018.05.016 ↗
44. Digital volume correlation can be used to estimate local strains in natural and augmented vertebrae: an organ-level study10.1016/j.jbiomech.2016.10.018 ↗
45. A combined RSA and FE study of the implanted proximal tibia: correlation of the post-operative mechanical environment with implant migration10.1016/j.jbiomech.2003.12.004 ↗
46. Mechanical performance of cementless total knee replacements: it is not all about the maximum loads10.1002/jor.24194 ↗
47. Undersizing the tibial baseplate in cementless total knee arthroplasty has only a small impact on bone-implant interaction: a finite element biomechanical study10.1016/j.arth.2022.10.032 ↗
48. Quantification of human bone microarchitecture damage in press-fit femoral knee implantation using HR-pQCT and digital volume correlation10.1016/j.jmbbm.2019.04.054 ↗
49. Application of the digital volume correlation technique for the measurement of displacement and strain fields in bone: a literature review10.1016/j.jbiomech.2014.01.001 ↗
50. The new knee society knee scoring system10.1007/s11999-011-2135-0 ↗
51. A comparison in proximal tibial strain between metal-backed and all-polyethylene anatomic graduated component total knee arthroplasty tibial components10.1016/j.arth.2009.06.018 ↗
52. Stair ambulation biomechanics following total knee arthroplasty: a systematic review10.1016/j.arth.2014.03.040 ↗
53. Influence of loading and activity on the primary stability of cementless tibial trays10.1002/jor.22056 ↗
54. Fatigue failure of cancellous bone: a possible cause of implant migration and loosening10.1302/0301-620X.79B2.0790181 ↗
55. Finite element analysis of the implanted proximal tibia: a relationship between the initial cancellous bone stresses and implant migration10.1016/S0021-9290(98)00022-0 ↗
56. Surface flatness after bone cutting: a cadaver study of tibial condyles10.3109/17453679108993084 ↗
57. Surface characteristics following tibial preparation during total knee arthroplasty10.1016/0883-5403(94)90138-4 ↗
58. Strain uncertainties from two digital volume correlation approaches in prophylactically augmented vertebrae: local analysis on bone and cement-bone microstructures10.1016/j.jmbbm.2016.12.006 ↗
59. Micro-CT scan optimisation for mechanical loading of tibia with titanium tibial tray: a digital volume correlation zero strain error analysis10.1016/j.jmbbm.2022.105336 ↗
60. Impact of surgical alignment, tray material, PCL condition, and patient anatomy on tibial strains after TKA10.1016/j.medengphy.2021.01.001 ↗
61. Validation and sensitivity of model-predicted proximal tibial displacement and tray micromotion in cementless total knee arthroplasty under physiological loading conditions10.1016/j.jmbbm.2020.103793 ↗
62. 3D full-field strain in bone-implant and bone-tooth constructs and their morphological influential factors10.1016/j.jmbbm.2020.103858 ↗
63. 3D full-field biomechanical testing of a glenoid before and after implant placement10.1016/j.eml.2019.100614 ↗