Comparison of tendon attachment to 3D printed Ti6Al4V implant versus Trevira® implant: A paired experimental animal study
Source abstract
Abstract
Background Soft-tissue attachment is crucial for the success of megaprosthesis surgery and improvement in current treatment is needed. The aim of this study was to compare the biomechanical and histomorphometric properties of soft-tissue attachment between 3D printed Ti6Al4V implants featuring a 630 μm microporous structure and commercially available Trevira® implants with a 200 μm porous structure in a non-loadbearing ovine model. Methods Ten skeletally mature ewes underwent surgical implantation with both implants. After 4-weeks, mechanical pull-out testing assessed the attachment strength, while histomorphometric analysis evaluated fibroblast cell profile density, multinucleated giant cell profile density, microvessel length and volume density. Results 3D printed Ti6Al4V implants demonstrated a 129% greater attachment strength compared to Trevira® implants (p = 0.003). In the Trevira® group, a 35% increase in fibroblast profile density (p < 0.001) and a 98% increase in multinucleated giant cell profile density (p < 0.001) were observed, with no significant difference in microvessel length density between the groups. However, the Ti6Al4V group exhibited a 50% higher microvessel volume density (p < 0.001) compared to the Trevira® group. Conclusion 3D printed Ti6Al4V implants with a 630 μm microporous structure demonstrated superior attachment strength, enhanced neovascularization, and reduced foreign body reaction compared to the Trevira® implants. These findings suggest that 3D printed Ti6Al4V implants may enhance soft-tissue attachment in megaprosthesis surgeries.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Tendon Structure and Treatment · Shoulder Injury and Treatment · Reconstructive Surgery and Microvascular Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (28)
Show source-provided references
1. Comparison of trabecular metal cups and titanium fiber-mesh cups in primary hip arthroplasty10.3109/17453674.2011.572251 ↗
2. Reconstruction of the extensor mechanism after proximal tibia endoprosthetic replacement10.1054/arth.2001.25502 ↗
3. Gait and electromyographic analysis of patients recovering after limb-saving surgery10.1016/S0268-0033(00)00021-8 ↗
4. Effect of storage and preconditioning of healing rat Achilles tendon on structural and mechanical properties10.1038/s41598-020-80299-w ↗
5. Modular prosthetic replacement of the proximal femur after resection of a bone tumour a long-term follow-up10.1302/0301-620X.83B8.0831156 ↗
6. Soft tissue adhesion patterns over Trevira tube on modular endoprosthesis for malignant bone tumours: an in vitro study
7. The effect of pore size on tissue ingrowth and neovascularization in porous bioceramics of controlled architecture in vivo10.1088/1748-6041/6/1/015007 ↗
8. Soft tissue reconstruction of megaprostheses using a trevira tube10.1097/00003086-200112000-00030 ↗
9. 3D-printed porous Ti6Al4V scaffolds for long bone repair in animal models: a systematic review10.1186/s13018-022-02960-6 ↗
10. Some new, simple and efficient stereological methods and their use in pathological research and diagnosis10.1111/j.1699-0463.1988.tb05320.x ↗
11. Effect of pore size of porous-structured titanium implants on tendon ingrowth10.1155/2022/2801229 ↗
12. Tantalum acetabular cups provide secure fixation in THA after pelvic irradiation at minimum 5-year followup10.1007/s11999-012-2382-8 ↗
13. The effects of multiple freeze–thaw cycles on the biomechanical properties of the human bone‐patellar tendon‐bone allograft10.1002/jor.21373 ↗
14. Measurement of stress-strain relationship and stress relaxation in various synthetic ligaments10.1007/BF01552654 ↗
15. Comparative analysis of bone ingrowth in 3D-printed titanium lattice structures with different patterns10.3390/ma16103861 ↗
16. 10.05 - selective laser sintering/melting
17. First experiences of a hospital-based 3D printing facility – an analytical observational study10.1186/s12913-023-10511-w ↗
18. New developments of Ti-based alloys for biomedical applications10.3390/ma7031709 ↗
19. Mechanical characterisation and numerical modelling of TPMS-based gyroid and diamond Ti6Al4V scaffolds for bone implants: an integrated approach for translational consideration10.3390/bioengineering9100504 ↗
20. Effect of eccentric exercise on the healing process of injured patellar tendon in rats10.1007/s00776-008-1242-6 ↗
21. The ARRIVE guidelines 2.0: updated guidelines for reporting animal research
22. How stereological analysis of vascular morphology can quantify the blood volume fraction as a marker for tumor vasculature: comparison with magnetic resonance imaging10.1038/jcbfm.2011.151 ↗
23. Pull-out bond strength of titanium post cemented with novel fast-setting calcium silicate cement
24. Gait and function in patients with a femoral endoprosthesis after tumor resection10.1080/00016470216319 ↗
25. Load and failure behavior of human muscle samples in the context of proximal femur replacement
26. Macrophages, foreign body giant Cells and their response to implantable biomaterials10.3390/ma8095269 ↗
27. Effects of postmortem storage by freezing on ligament tensile behavior10.1016/0021-9290(86)90016-3 ↗
28. Zheng · 2020