journal article·2025-08Open access

The utility of additively manufactured β-Ti latticed hip implants in reducing femoral stress shielding: A finite element study

Peter Ibrahim, Piyapat Jameekornkul, Ajit Panesar, Moataz M. Attallah

Abstract

Total or partial hip replacement procedures have become one of the most frequently performed surgical procedures worldwide. However, a significant and increasing percentage of these patients return for revision surgeries due to stress shielding phenomena and the following implications such as aseptic implant loosening. Various strategies are under investigation to reduce the mechanical properties gap between the implant and the femur tissues. In this study, different development aspects have been merged to improve the design of hip implants and minimise the stress shielding. A low elastic modulus β-Ti TNTZO alloy has been used as the implant material. An actively optimised lattice structure has been embedded into the implant to tune its mechanical properties according to the bone properties of the patient. The study showed that using TNTZO reduced the average stress shielding of the implant by 31 % compared to commercial CoCr implants. Moreover, with two-step TNTZO lattice optimisation, the stress shielding was fully eliminated in all the femur sections, which accelerates the healing process of the bone and extends its long-term life. The study presented a scope for further optimisation to address other objectives, such as implant stiffness and eliminating stress concentration zones.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Titanium Alloys Microstructure and Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (84)

Show source-provided references
1. Achakri
2. Porous lattice structure of femoral stem for total hip arthroplasty
3. Mechanical and fatigue behavior of cellular structure Ti-6Al-4V alloy femoral stems: a finite element analysis
4. Hybrid biomechanical design of dental implants: integrating solid and gyroid triply periodic minimal surface lattice architectures for optimized stress distribution10.3390/jfb16020054 ↗
5. Finite element study of functionally graded porous femoral stems incorporating body-centered cubic structure10.1111/aor.13444 ↗
6. Reduction of the surface roughness of additively manufactured metallic parts by enhanced electrolytic smoothening
7. The effect of post-sintering treatments on the fatigue and biological behavior of Ti-6Al-4V ELI parts made by selective laser melting10.1016/j.jmbbm.2017.03.024 ↗
8. Hip joint loading during walking and running, measured in two patients10.1016/0021-9290(93)90058-M ↗
9. The design of additively manufactured lattices to increase the functionality of medical implants10.1016/j.msec.2018.10.052 ↗
10. The design of additively manufactured lattices to increase the functionality of medical implants10.1016/j.msec.2018.10.052 ↗
11. Reduced stress shielding with limited micromotions using a carbon fibre composite biomimetic hip stem: a finite element model10.1177/0954411911412465 ↗
12. Review—electropolishing of additive manufactured metal parts10.1149/1945-7111/ac6450 ↗
13. Trabecular bone microarchitecture: a review10.1016/j.morpho.2008.10.003 ↗
14. Evaluation of orthogonal mechanical properties and density of human trabecular bone from the major metaphyseal regions with materials testing and computed tomography10.1002/jor.1100090507 ↗
15. Production of customized hip stem prostheses - a comparison between conventional machining and electron beam melting (EBM)10.1108/RPJ-07-2011-0067 ↗
16. Optimization of artificial hip joint parameters10.1002/mawe.200900430 ↗
17. Fatigue performance evaluation of selective laser melted Ti-6Al-4V10.1016/j.msea.2014.01.041 ↗
18. Stiffness optimization design for TPMS architected cellular materials
19. Mechanical properties and biocompatibility of various cobalt chromium dental alloys
20. Finite element modelling of the femur bone in SCI during FES finite element modelling of the femur bone of a subject suffering from motor neuron lesion subjected to electrical stimulation
21. Effects of materials of cementless femoral stem on the functional adaptation of bone10.1016/S1672-6529(11)60098-X ↗
22. Potentiality of the ‘gum Metal’ titanium-based alloy for biomedical applications10.1016/j.msec.2014.08.003 ↗
23. Estimation of mass apparent density and Young's modulus of femoral neck-head region10.1080/03091902.2020.1799093 ↗
24. On the design evolution of hip implants: a review
25. Tailoring selective laser melting process for titanium drug-delivering implants with releasing micro-channels
26. Fabrication, structure and mechanical and ultrasonic properties of medical Ti6Al4V alloys part I: microstructure and mechanical properties of Ti6Al4V alloys suitable for ultrasonic scalpel
27. Improved design of cementless hip stems using two-dimensional functionally graded materials10.1002/jbm.b.30509 ↗
28. Determination of muscle loading at the hip joint for use in pre-clinical testing10.1016/j.jbiomech.2004.05.022 ↗
29. A CBCT atlas of buccal cortical bone thickness in interradicular spaces10.2319/082214-593.1 ↗
30. Assessment of trapped powder removal and inspection strategies for powder bed fusion techniques10.1007/s00170-020-04930-w ↗
31. Laser powder bed fusion of a β titanium alloy: microstructural development, post-processing, and mechanical behaviour10.1016/j.msea.2024.146617 ↗
32. Hybrid manufacturing and performance evaluation of β Ti-alloy stents
33. ISO-7206-4, “Implants for Surgery Partial and Total Hip Joint Prostheses — Part 4: Determination of Endurance Properties and Performance of Stemmed Femoral Components”.
34. Selective laser melting of ti-6al-4v: the impact of post-processing on the tensile, fatigue and biological properties for medical implant applications10.3390/ma13122813 ↗
35. Development, characterisation, and modelling of processability of nitinol stents using laser powder bed fusion10.1016/j.jallcom.2022.164681 ↗
36. An experimental and numerical investigation of compressive response of designed schwarz primitive triply periodic minimal surface with non-uniform shell thickness10.1016/j.eml.2020.100671 ↗
37. Analysis of hounsfield unit of human bones for strength evaluation10.1016/j.mspro.2014.07.065 ↗
38. Personalized stem length optimization in hip replacement: a microscopic perspective on bone—implant interaction10.3390/bioengineering11111074 ↗
39. Finite element analysis of orthopedic hip implant with functionally graded bioinspired lattice structures10.3390/biomimetics5030044 ↗
40. Pre-launch’ finite element analysis of a short-stem total hip arthroplasty system consisting of two implant types10.1016/j.clinbiomech.2018.11.002 ↗
41. Improved fatigue properties with maintaining low Young's modulus achieved in biomedical beta-type titanium alloy by oxygen addition10.1016/j.msea.2017.07.078 ↗
42. Capability of auxetic femoral stems to reduce stress shielding after total hip arthroplasty10.1016/j.jot.2022.11.001 ↗
43. Titanium alloys in total joint replacement-a materials science perspective10.1016/S0142-9612(97)00146-4 ↗
44. Influence of defects, surface roughness and HIP on the fatigue strength of Ti-6Al-4V manufactured by additive manufacturing10.1016/j.ijfatigue.2018.07.020 ↗
45. Finite element modelling and characterization of 3D cellular microstructures for the design of a cementless biomimetic porous hip stem10.1016/j.matdes.2018.04.002 ↗
46. A comprehensive analysis of bio-inspired design of femoral stem on primary and secondary stabilities using mechanoregulatory algorithm10.1007/s10237-020-01334-3 ↗
47. A novel design, analysis and 3D printing of Ti-6Al-4V alloy bio-inspired porous femoral stem10.1007/s10856-020-06420-7 ↗
48. Recent advances on biofunctionalization of metallic substrate using ceramic coating: how far are we from clinically stable implant?10.1016/j.jtice.2021.01.013 ↗
49. Morgan · 2001
50. Bone mechanical properties in healthy and diseased states10.1146/annurev-bioeng-062117-121139 ↗
51. The pathology of failed McKee-Farrar implants: correlation with modern metal-on-metal-implant failure10.1007/s10856-017-5882-y ↗
52. Stress shielding and bone resorption of press-fit polyether–ether–ketone (PEEK) hip prosthesis: a sawbone model study10.3390/polym14214600 ↗
53. A novel hybrid design and modelling of a customised graded Ti-6Al-4V porous hip implant to reduce stress-shielding: an experimental and numerical analysis
54. Development of low rigidity β-type titanium alloy for biomedical applications10.2320/matertrans.43.2970 ↗
55. Patellofemoral cartilage stresses are Most sensitive to variations in vastus medialis muscle forces10.1080/10255842.2018.1544629 ↗
56. Effect of density and unit cell size grading on the stiffness and energy absorption of short fibre-reinforced functionally graded lattice structures
57. Ratner · 2004
58. Relations of mechanical properties to density and CT numbers in human bone10.1016/1350-4533(95)97314-F ↗
59. Problem of stress shielding and improvement to the hip implant designs A review10.3923/jms.2007.460.467 ↗
60. Hip implants VII: finite element analysis and optimization of cross-sections10.1016/j.matdes.2007.09.002 ↗
61. Effect of anatomical variability on stress-shielding induced by short calcar-guided stems: automated finite element analysis of 90 femora10.1002/jor.24240 ↗
62. When the total hip replacement fails: a review on the stress-shielding effect10.3390/pr10030612 ↗
63. Optimal material selection for total hip implant: a finite element case study10.1007/s13369-019-04088-y ↗
64. Analysis of stress shielding reduction in bone fracture fixation implant using functionally graded materials10.1016/j.compstruct.2023.117262 ↗
65. Projections and epidemiology of primary hip and knee arthroplasty in medicare patients to 2040-206010.2106/JBJS.OA.22.00112 ↗
66. Good stability but high periprosthetic bone mineral loss and late-occurring periprosthetic fractures with use of unce-mented tapered femoral stems in patients with a femoral neck fracture A 5-year follow-up of 31 patients using RSA and DXA
67. Mechanical properties comparison of strut-based and triply periodic minimal surface lattice structures produced by electron beam melting
68. Advancement in total hip implant: a comprehensive review of mechanics and performance parameters across diverse novelties10.1039/D3TB01469J ↗
69. Stress transfer, stress shielding, and micro-movement analysis in the implanted femoral bone region with various hip implant stem materials: a finite element approach
70. Effects of unit cell size on the mechanical performance of additive manufactured lattice structures
71. Physiologically based boundary conditions in finite element modelling10.1016/j.jbiomech.2006.10.038 ↗
72. Laser powder bed fusion of Ti-rich TiNi lattice structures: process optimisation, geometrical integrity, and phase transformations10.1016/j.ijmachtools.2019.04.002 ↗
73. Additive manufacturing of bio-inspired multi-scale hierarchically strengthened lattice structures
74. Hip joint prosthesis using SiC CMC and Ti-6Al-4V materials
75. Selective laser melting of high-strength, low-modulus Ti–35Nb–7Zr–5Ta alloy
76. Hip implant design with three-dimensional porous architecture of optimized graded density10.1115/1.4041208 ↗
77. On the design and properties of porous femoral stems with adjustable stiffness gradient10.1016/j.medengphy.2020.05.003 ↗
78. The impact of reducing the Femoral stem length in total hip arthroplasty during gait10.1007/s00402-021-03852-w ↗
79. Effects of electropolishing on mechanical properties and bio-corrosion of Ti6Al4V fabricated by electron beam melting additive manufacturing10.3390/ma12091466 ↗
80. Numerical investigation of the effect of porous titanium femoral prosthesis on bone remodeling10.1016/j.matdes.2010.12.042 ↗
81. Using strain energy-based prediction of effective elastic properties in topology optimization of material microstructures10.1007/s10409-006-0045-2 ↗
82. The effects of length of femoral stem on aseptic loosening following cemented distal femoral endoprosthetic replacement in tumour surgery10.1007/s00264-020-04629-3 ↗
83. Optimal design of functionally graded lattice structures using hencky bar-grid model and topology optimization10.1007/s00158-022-03368-w ↗
84. Evaluation of orthogonal mechanical properties and density of human trabecular bone from the major metaphyseal regions with materials testing and computed tomography10.1002/jor.1100090507 ↗