Full-field strain measurements on medical devices using digital image correlation: Considerations and practical examples
Source abstract
Abstract
Access to direct stress or strain measurements on implantable medical devices could transform device design, durability assessment, and computational model development and validation activities. However, examples of acquiring such measurements are lacking. Here, we consider the use of optical digital image correlation (DIC) for characterizing full-field strains on medical devices and medical device surrogate specimens. Several common and unique speckle pattern application approaches are investigated including aerosol painting, application of transfer or decal papers, low- and high-pressure airbrushing using traditional and ultra-low reflectance paints, deposition of monodisperse graphite powder, and custom microstamping. Among the methods considered, comparative speckling studies show high-pressure airbrushing generates the best patterns at the millimeter-scale and monodisperse graphite powder or microstamping generate the best patterns at the sub-millimeter scale. Full-field DIC strain measurements additionally reveal Lüders-like phase transformation bands on nitinol wire, nitinol Z-specimens, and nitinol stent rings under bending loads. Overall, the study demonstrates the feasibility of direct strain measurement on real medical devices at fields of view ranging from millimeters to hundreds of microns and can be used to inform the planning and execution of future DIC studies on medical devices.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Optical measurement and interference techniques · Advanced Measurement and Metrology Techniques · Surface Roughness and Optical Measurements
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (87)
Show source-provided references
1. High resolution three-dimensional strain mapping of bioprosthetic heart valves using digital image correlation10.1016/j.jbiomech.2018.05.020 ↗
2. Experimental validation of an ITAP numerical model and the effect of implant stem stiffness on bone strain energy10.1007/s10439-020-02456-6 ↗
3. The effect of implant position on bone strain following lateral unicompartmental knee arthroplasty: a biomechanical model using digital image correlation10.1302/2046-3758.68.BJR-2017-0067.R1 ↗
4. Full-field microscale strain measurements of a nitinol medical device using digital image correlation10.1016/j.jmbbm.2020.104221 ↗
5. Digital volume correlation: three-dimensional strain mapping using X-ray tomography10.1007/BF02323555 ↗
6. Micro- and nanoscale deformation measurement of surface and internal planes via digital image correlation10.1007/s11340-006-0531-2 ↗
7. Increasing accuracy and precision of digital image correlation through pattern optimization10.1016/j.optlaseng.2016.11.005 ↗
8. Present and future approaches to lifetime prediction of superelastic nitinol10.1016/j.tafmec.2017.04.001 ↗
9. Micro speckle stamping: high contrast, no basecoat, repeatable, well-adhered
10. Microstamping for improved speckle patterns to enable digital image correlation10.1017/S1431927615003050 ↗
11. Impact of speckle pattern parameters on DIC strain resolution calculated from in-situ SEM experiments. Fracture, fatigue, failure, and damage evolution
12. A full-field DIC analysis of the mechanical-deformation behavior of polyethylene terephthalate glycol (PET-G) aligners10.1016/j.jmbbm.2022.105391 ↗
13. Full-field in vitro measurements and in silico predictions of strain shielding in the implanted femur after total hip arthroplasty10.1177/0954411915591617 ↗
14. A method to transfer speckle patterns for digital image correlation10.1088/0957-0233/26/9/095201 ↗
15. Cheng · 2019
16. Speckle pattern quality assessment for digital image correlation10.1016/j.optlaseng.2013.03.014 ↗
17. Plastic strain mapping with sub-micron resolution using digital image correlation10.1007/s11340-012-9685-2 ↗
18. The influence of acetabular cup material on pelvis cortex surface strains, measured using digital image correlation10.1016/j.jbiomech.2011.11.042 ↗
19. A review of speckle pattern fabrication and assessment for digital image correlation10.1007/s11340-017-0283-1 ↗
20. Time-resolved in situ synchrotron-microCT: 4D deformation of bone and bone analogues using digital volume correlation10.1016/j.actbio.2021.06.014 ↗
21. Polymerization shrinkage of resin-based composites for dental restorations: a digital volume correlation study10.1016/j.dental.2019.08.116 ↗
22. Experimental validation of finite element models of intact and implanted composite hemipelvises using digital image correlation10.1115/1.4007173 ↗
23. The application of digital volume correlation (DVC) to study the microstructural behaviour of trabecular bone during compression10.1016/j.jmbbm.2013.09.014 ↗
24. 3D deformation and strain measurement of an intact eardrum using digital image correlation10.1088/1742-6596/1149/1/012026 ↗
25. Microscale repeatability of the shape-memory effect in fine NiTi wires10.1007/s40830-016-0081-0 ↗
26. Fatigue to fracture: an informative, fast, and reliable approach for assessing medical implant durability
27. Load path change on superelastic NiTi alloys: in situ synchrotron XRD and SEM DIC10.1016/j.actamat.2017.11.035 ↗
28. Fluorescent stereo microscopy for 3D surface profilometry and deformation mapping10.1364/OE.21.011808 ↗
29. Digital volume correlation for study of the mechanics of whole bones10.1016/j.piutam.2012.05.013 ↗
30. Development and in vitro validation of a simplified numerical model for the design of a biomimetic femoral stem10.1016/j.jmbbm.2017.10.019 ↗
31. Jones · 2018
32. Experimental investigation of strain rate dependence of nanocrystalline Pt films10.1007/s11340-008-9212-7 ↗
33. Small-scale patterning methods for digital image correlation under scanning electron microscopy10.1088/0957-0233/22/12/125501 ↗
34. Self-assembled nanoparticle surface patterning for improved digital image correlation in a scanning electron microscope10.1007/s11340-013-9734-5 ↗
35. Digital image correlation of coated and uncoated Religa Heart_Ext ventricular assist device
36. Quality assessment of speckle patterns for digital image correlation10.1016/j.optlaseng.2005.10.004 ↗
37. Analysis of speckle patterns for deformation measurements by digital image correlation
38. Grain size effects on NiTi shape memory alloy fatigue crack growth10.1557/jmr.2017.395 ↗
39. Optimum paint sequence for speckle patterns in digital image correlation10.1007/s40799-017-0192-3 ↗
40. A practical approach to optimizing the preparation of speckle patterns for digital-image correlation10.1088/0957-0233/25/10/107001 ↗
41. Anisotropic crack propagation and deformation in dentin observed by four-dimensional X-ray nano-computed tomography10.1016/j.actbio.2019.06.042 ↗
42. The effect of rotating platform TKA on strain distribution and torque transmission on the proximal tibia10.1016/j.arth.2013.08.024 ↗
43. Thermo-mechanical toner transfer for high-quality digital image correlation speckle patterns10.1016/j.optlaseng.2015.06.009 ↗
44. Demonstration of an integrated-heating load frame for quantitatively assessing microscale tensile properties of copper and Zircaloy-210.1016/j.jnucmat.2024.155339 ↗
45. Advancing regulatory science with computational modeling for medical devices at the FDA’s Office of Science and Engineering Laboratories10.3389/fmed.2018.00241 ↗
46. How active are patients undergoing total joint arthroplasty?: a systematic review10.1007/s11999-009-1135-9 ↗
47. The use of digital image correlation in the biomechanical area: a review10.1080/23335432.2015.1117395 ↗
48. Three-dimensional local measurements of bone strain and displacement: comparison of three digital volume correlation approaches10.1115/1.4030174 ↗
49. Mean intensity gradient: an effective global parameter for quality assessment of the speckle patterns used in digital image correlation10.1016/j.optlaseng.2009.08.010 ↗
50. A flexible and accurate digital volume correlation method applicable to high-resolution volumetric images10.1088/1361-6501/aa7e8a ↗
51. A probabilistic approach with built-in uncertainty quantification for the calibration of a superelastic constitutive model from full-field strain data10.1016/j.commatsci.2021.110357 ↗
52. Phase transformation volume amplitude as a low-cycle fatigue indicator in nickel–titanium shape memory alloys10.1016/j.scriptamat.2019.12.014 ↗
53. Development of the Z specimen for tensile-tensile, tensile-compression, compression-compression wire testing10.1007/s11665-010-9766-2 ↗
54. Speckle patterns for DIC in challenging scenarios: rapid application and impact endurance10.1088/1361-6501/abaae8 ↗
55. Full‐field strain mapping by optical correlation of micrographs acquired during deformation10.1111/j.1365-2818.2005.01461.x ↗
56. Tension, compression, and bending of superelastic shape memory alloy tubes10.1016/j.jmps.2012.12.012 ↗
57. Tips and tricks for characterizing shape memory wire part 5: full-field strain measurement by digital image correlation10.1111/j.1747-1567.2011.00717.x ↗
58. Axial-torsion behavior of superelastic tubes: part I, proportional isothermal experiments10.1016/j.ijsolstr.2020.03.018 ↗
59. All about speckles: aliasing10.1111/ext.12111 ↗
60. All about speckles: speckle size measurement10.1111/ext.12110 ↗
61. All about speckles: contrast10.1111/ext.12126 ↗
62. All about speckles: edge sharpness10.1111/ext.12139 ↗
63. All about speckles: speckle density10.1111/ext.12161 ↗
64. DIC: a revolution in experimental mechanics10.1111/ext.12173 ↗
65. Virtual strain gage size study10.1111/ext.12172 ↗
66. Selectively electron-transparent microstamping toward concurrent digital image correlation and high-angular resolution electron backscatter diffraction (EBSD) analysis10.1017/S1431927617012703 ↗
67. Assessment of the film-free water decal method for speckle pattern application in digital image correlation10.3390/s24175657 ↗
68. Characterization and validation of fatigue strains for superelastic nitinol using digital image correlation
69. Dimensional variability characterization of additively manufactured lattice coupons10.1186/s41205-022-00141-z ↗
70. Refractive three-dimensional reconstruction for underwater stereo digital image correlation10.1364/OE.421708 ↗
71. Experimental validation of 3D printed patient-specific implants using digital image correlation and finite element analysis10.1016/j.compbiomed.2014.06.002 ↗
72. Strain field measurements on mouse carotid arteries using microscopic three‐dimensional digital image correlation10.1002/jbm.a.31268 ↗
73. Volumetric digital image correlation (VDIC)
74. Sutton · 2009
75. Comparative analysis of the biomechanical behavior of two different design metaphyseal-fitting short stems using digital image correlation10.1186/s12938-020-00806-y ↗
76. Mapping the strain distribution on the proximal femur with titanium and flexible-stemmed implants using digital image correlation10.1302/0301-620X.92B8.23553 ↗
77. Evaluating the bending response of two osseointegrated transfemoral implant systems using 3D digital image correlation10.1115/1.4003871 ↗
78. A digital image correlation analysis on the influence of crown material in implant-supported prostheses on bone strain distribution10.1016/j.jpor.2011.05.003 ↗
79. Digital image correlation analysis of the load transfer by implant-supported restorations10.1016/j.jbiomech.2011.02.015 ↗
80. Validation of finite element models for strain analysis of implant-supported prostheses using digital image correlation10.1016/j.dental.2013.04.010 ↗
81. Non-clinical engineering tests and recommended labeling for intravascular stents and associated delivery systems
82. Technical considerations for non-clinical assessment of medical devices containing nitinol
83. Assessing the credibility of computational modeling and simulation in medical device submissions
84. Development of a versatile speckle pattern of nano-sized polymer particles for high-resolution SEM-DIC10.1016/j.polymertesting.2023.108134 ↗
85. Rotary Bend fatigue of nitinol to one billion cycles10.1007/s40830-022-00409-7 ↗
86. Investigation of particle fracture during fatigue of aluminum 202410.1016/j.ijfatigue.2018.02.001 ↗
87. Fabrication of speckle patterns by focused ion beam deposition and its application to micro-scale residual stress measurement10.1088/0957-0233/26/9/095601 ↗