Fatigue life assessment of cardiovascular balloon-expandable stents: A two-scale plasticity–damage model approach
Source abstract
Abstract
Cardiovascular disease has become a major global health care problem in the present decade. To tackle this problem, the use of cardiovascular stents has been considered a promising and effective approach. Numerical simulations to evaluate the in vivo behavior of stents are becoming more and more important to assess potential failures. As the material failure of a stent device has been often associated with fatigue issues, as a result of the high number of cyclic loads these devices are subjected to in vivo, numerical approaches for fatigue life assessment of stents has gained special interest in the engineering community. Numerical fatigue predictions can be used to modify the design and prevent failure, without making and testing numerous physical devices, thus preventing from undesired fatigue failures. This work presents a fatigue life numerical method for the analysis of cardiovascular balloon-expandable stainless steel stents. The method is based on a two-scale continuum damage mechanics model in which both plasticity and damage mechanisms are assumed to take place at a scale smaller than the scale of the representative volume element. The fatigue failure criterion is based on the Soderberg relation. The method is applied to the fatigue life assessment of both PalmazShatz and Cypher stent designs. Validation of the method is performed through comparison of the obtained numerical results with some experimental results available for the PalmazShatz stent design. The present study gives also possible directions for future research developments in the framework of the numerical fatigue life assessment of real balloon-expandable stents.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Fatigue and fracture mechanics · Hydrogen embrittlement and corrosion behaviors in metals · Corrosion Behavior and Inhibition
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (55)
Show source-provided references
1. Enhanced fatigue resistance in 316l austenitic stainless steel due to low-temperature paraequilibrium carburization10.1016/j.actamat.2007.06.025 ↗
2. AHA., 2010. Circulation 121, e46–e215.
3. Fatigue-life computational analysis for the self-expanding endovascular nitinol stents
4. Finite element analysis of a stenotic artery revascularization through stent insertion10.1080/10255840108908007 ↗
5. Stent expansion: a combination of delivery balloon underexpansion and acute stent recoil reduces predicted stent diameter irrespective of reference vessel size10.1136/hrt.2006.107052 ↗
6. Barbier, G., Desmorat, R., Sermage, J.-P., du Tertre, A., Courtin, S., Dehouve, J., Tchou-Kien, D., 2008. Mean stress effect by incremental two scale damage model. In: LCF6—6th International Conference on Low Cycle Fatigue, Berlin, Germany.
7. Realistic finite element-based stent design: the impact of balloon folding10.1016/j.jbiomech.2007.08.014 ↗
8. Continuum damage mechanics: part I—general concepts10.1115/1.3173661 ↗
9. Continuum damage mechanics: part II—damage growth, crack initiation, and crack growth10.1115/1.3173662 ↗
10. Damage and fatigue: continuum damage mechanics modelling for fatigue of materials and structures
11. Desmorat, R., Lemaitre, J., 2001. Continuous damage: a two-scale model for quasi-brittle and fatigue damage. In: Handbook of Materials Behavior Models, Deformations of Materials, vol. 1. Academic Press, pp. 525–535.10.1016/B978-012443341-0/50058-2 ↗
12. Two scale damage model and related numerical issues for thermo-mechanical high cycle fatigue10.1016/j.euromechsol.2007.01.002 ↗
13. The determination of the elastic field of an ellipsoidal inclusion and related problems
14. Coupling between mesoplasticity and damage in high-cycle fatigue10.1177/1056789506067935 ↗
15. Competition between mesoplasticity and damage under HCF—elasticity/damage shakedown concept10.1016/j.ijfatigue.2006.11.001 ↗
16. On the effects of different strategies in modelling balloon-expandable stenting by means of finite element method10.1016/j.jbiomech.2008.01.027 ↗
17. Glenn, R., Lee, J., 1997. Accelerated pulsite fatigue testing of Ni–Ti coronary stents. In: Pelton, A., Hodgson, D., Russell, S., Duerig, T. (Eds.), SMST-97: Proceedings of the Second International Conference on Shape Memory and Superelastic Technologies. SMST Society, Pacific Grove, California, CA, USA.
18. Goodman · 1899
19. Modeling of size dependent failure in cardiovascular stent struts under tension and bending10.1007/s10439-007-9326-6 ↗
20. Fatigue-life assessment and validation techniques for metallic vascular implants10.1016/j.biomaterials.2009.10.028 ↗
21. On the finite element modelling of balloon-expandable stents10.1016/j.jmbbm.2007.07.002 ↗
22. Time of the rupture process under creep conditions
23. Ladevéze, P., Lemaitre, J., 1984. Damage effective stress in quasi unilateral conditions. In: The 16th International Congress of Theoretical and Applied Mechanics. Lyngby, Denmark.
24. Cardiovascular stent design and vessel stresses: a finite element analysis10.1016/j.jbiomech.2004.07.022 ↗
25. Erratum to: “Cardiovascular stent design and vessel stresses: a finite element analysis”10.1016/j.jbiomech.2006.03.001 ↗
26. A continuous damage mechanics model for ductile fracture10.1115/1.3225775 ↗
27. Lemaitre · 1992
28. Lemaitre · 1990
29. Lemaitre · 2005
30. Damage 90: a post processor for crack initiation10.1016/0045-7825(94)90060-4 ↗
31. A two scale damage concept applied to fatigue10.1023/A:1018641414428 ↗
32. Anisotropic damage law of evolution10.1016/S0997-7538(00)00161-3 ↗
33. Fatigue life analysis and experimental verification of coronary stent10.1007/s00380-009-1203-9 ↗
34. Fatigue and life prediction for cobalt–chromium stents: a fracture mechanics analysis10.1016/j.biomaterials.2005.10.012 ↗
35. Analysis of the mechanical performance of a cardiovascular stent design based on micromechanical modelling10.1016/j.commatsci.2004.05.001 ↗
36. Computational examination of the effect of material inhomogeneity on the necking of stent struts under tensile loading10.1115/1.2722776 ↗
37. Mechanical behavior of coronary stents investigated through the finite element method10.1016/S0021-9290(02)00033-7 ↗
38. A non-local theory devoted to high cycle multiaxial fatigue
39. A novel simulation strategy for stent insertion and deployment in curved coronary bifurcations: comparison of three drug-eluting stents10.1007/s10439-009-9836-5 ↗
40. The stress-strain behavior of coronary stent struts is size dependent10.1114/1.1569268 ↗
41. A continuum mechanics model for mechanical fatigue analysis10.1016/j.commatsci.2004.08.001 ↗
42. Continuum approach to high-cycle fatigue modelling10.1016/j.ijfatigue.2007.08.009 ↗
43. Fatigue and durability of nitinol stents10.1016/j.jmbbm.2007.08.001 ↗
44. Finite element analysis and fatigue of stents10.1080/136457002760273377 ↗
45. Analysis of prolapse in cardiovascular stents: a constitutive equation for vascular tissue and finite element modelling10.1115/1.1613674 ↗
46. Rabotnov · 1969
47. Coronary stent strut size dependent stress–strain response investigated using micromechanical finite element models10.1023/B:ABME.0000012740.47963.9e ↗
48. Schijve · 2009
49. Factor of safety and working stress10.1115/1.4057307 ↗
50. Suresh · 1998
51. Plastic strain in metals
52. Analysis of the transient expansion behaviour and design optimization of coronary stents by finite element method10.1016/j.jbiomech.2004.11.003 ↗
53. Fatigue and endurance of coronary stents10.1002/mawe.200800409 ↗
54. Fatigue and fracture in materials used for micro-scale biomedical components
55. A combined experimental and computational study of deformation in grains of biomedical grade 316lvm stainless steel10.1016/j.actamat.2006.06.021 ↗