Effect of vacuum-treatment on deformation properties of PMMA bone cement

Fatima Zivic, Miroslav Babic, Nenad Grujovic, Slobodan Mitrovic, Gregory Favaro, Mihaela Caunii

Publisher page

Abstract

Deformation behavior of polymethyl methacrylate (PMMA) bone cement is explored using microindentation. Two types of PMMA bone cement were prepared. Vacuum treated samples were subjected to the degassing of the material under vacuum of 270 mbar for 35 s, followed by the second degassing under vacuum of 255 mbar for 35 s. Air-cured samples were left in ambient air to cool down and harden. All samples were left to age for 6 months before the test. The samples were then subjected to the indentation fatigue test mode, using sharp Vickers indenter. First, loading segment rise time was varied in order to establish time-dependent behavior of the samples. Experimental data showed that viscous part of the deformation can be neglected under the observed test conditions. The second series of microindentation tests were realized with variation of number of cycles and indentation hardness and modulus were obtained. Approximate hardness was also calculated using analysis of residual impression area. Porosity characteristics were analyzed using CellC software. Scanning electron microscopy (SEM) analysis showed that air-cured bone cement exhibited significant number of large voids made of aggregated PMMA beads accompanied by particles of the radiopaque agent, while vacuum treated samples had homogeneous structure. Air-cured samples exhibited variable hardness and elasticity modulus throughout the material. They also had lower hardness values (approximately 65-100 MPa) than the vacuum treated cement (approximately 170 MPa). Porosity of 5.1% was obtained for vacuum treated cement and 16.8% for air-cured cement. Extensive plastic deformation, microcracks and craze whitening were produced during indentation of air-cured bone cement, whereas vacuum treated cement exhibited no cracks and no plastic deformation.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced materials and composites · Metal and Thin Film Mechanics · Orthopaedic implants and arthroplasty

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (37)

Show source-provided references
1. Fatigue of the cement/bone interface: the surface texture of bone and loosening10.1002/jbm.b.30364 ↗
2. Indentation cracking of composite matrix materials10.1177/00220345940730080901 ↗
3. Nanomechanical characterisation of solid surfaces and thin films10.1179/095066003225010227 ↗
4. Contact damage of poly(methylmethacrylate) during complex microdisplacements10.1016/S0043-1648(99)00366-X ↗
5. Fracture characteristics of acrylic bone cement–bone composites10.1243/09544110360729063 ↗
6. Fatigue of PMMA bone cement, fracture of nano and engineering materials and structures
7. Static and dynamic failure of PMMA spheres10.1016/j.powtec.2003.09.008 ↗
8. Indentation across size scales and disciplines: recent developments in experimentation and modeling10.1016/j.actamat.2006.08.044 ↗
9. Fatigue in porous PMMA: the effect of stress concentrations10.1016/j.ijfatigue.2007.08.022 ↗
10. Materials for joint replacement
11. Biomaterials
12. Factors affecting the mechanical and viscoelastic properties of acrylic bone cement10.1023/A:1016150403665 ↗
13. Time and mixing technique-dependent changes in bone cement smartset§HV
14. Fatigue testing and performance of acrylic bone–cement materials: state-of-the-art review10.1002/jbm.b.10018 ↗
15. Viscoelastic properties of injectable bone cements for orthopaedic applications: state-of-the-art review10.1002/jbm.b.31835 ↗
16. Critical comparison of two methods for the determination of nanomechanical properties of a material: application to synthetic and natural biomaterials10.1002/jbm.b.30489 ↗
17. Evaluation of an accelerated ageing medium for acrylic bone cement based on analysis of nanoindentation measurements on laboratory-prepared and retrieved specimens
18. Some failure modes of four clinical bone cements10.1243/0954411011535948 ↗
19. Early loosening of the femoral component at the cement–prosthesis interface after total hip replacement10.2106/00004623-199509000-00004 ↗
20. The mechanism of loosening in cemented hip prostheses determined from long-term results10.1007/BF00434099 ↗
21. Devices for tribotests at micro/nano scale
22. Reliability of PMMA bone cement fixation: fracture and fatigue crack-growth behaviour10.1023/A:1018574109544 ↗
23. Time dependent mechanical properties of bone cement. An in vitro study over one year10.1002/jbm.b.30811 ↗
24. An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments10.1557/JMR.1992.1564 ↗
25. Analytical techniques for indentation of viscoelastic materials10.1080/14786430600740666 ↗
26. Load–displacement behavior during sharp indentation of viscous-elastic-plastic materials10.1557/JMR.2003.0020 ↗
27. A practical guide for analysis of nanoindentation data10.1016/j.jmbbm.2008.10.002 ↗
28. Relevance of instrumented micro-indentation for the assessment of hardness and Young’s modulus of brittle materials10.1016/j.msea.2006.11.109 ↗
29. Bone cements: review of their physiochemical and biochemical properties in percutaneous vertebroplasty
30. Design of artificial kidneys
31. In vivo behavior of acrylic bone cement in total hip arthroplasty10.1016/j.biomaterials.2005.05.103 ↗
32. Microtomography assessment of failure in acrylic bone cement10.1016/j.biomaterials.2005.04.064 ↗
33. Crack initiation processes in acrylic bone cement10.1002/jbm.a.32037 ↗
34. Reverse modeling and solid free-form fabrication of sternum implant10.1007/s13246-010-0029-1 ↗
35. Mode of loading and contact configuration effects in the wear of polymers10.1002/(SICI)1097-4628(19960815)61:7<1217::AID-APP18>3.0.CO;2-U ↗
36. Rheologic fracture of PMMA material at different strain rates10.1007/s11771-007-0278-7 ↗
37. Dental biomaterials