journal article·2015-09Open access

Static and dynamic fatigue behavior of topology designed and conventional 3D printed bioresorbable PCL cervical interbody fusion devices

Ashleen R. Knutsen, Sean L. Borkowski, Edward Ebramzadeh, Colleen L. Flanagan, Scott J. Hollister, Sophia N. Sangiorgio

Abstract

Recently, as an alternative to metal spinal fusion cages, 3D printed bioresorbable materials have been explored; however, the static and fatigue properties of these novel cages are not well known. Unfortunately, current ASTM testing standards used to determine these properties were designed prior to the advent of bioresorbable materials for cages. Therefore, the applicability of these standards for bioresorbable materials is unknown. In this study, an image-based topology and a conventional 3D printed bioresorbable poly(ε)-caprolactone (PCL) cervical cage design were tested in compression, compression-shear, and torsion, to establish their static and fatigue properties. Difficulties were in fact identified in establishing failure criteria and in particular determining compressive failure load. Given these limitations, under static loads, both designs withstood loads of over 650 N in compression, 395 N in compression-shear, and 0.25 Nm in torsion, prior to yielding. Under dynamic testing, both designs withstood 5 million (5M) cycles of compression at 125% of their respective yield forces. Geometry significantly affected both the static and fatigue properties of the cages. The measured compressive yield loads fall within the reported physiological ranges; consequently, these PCL bioresorbable cages would likely require supplemental fixation. Most importantly, supplemental testing methods may be necessary beyond the current ASTM standards, to provide more accurate and reliable results, ultimately improving preclinical evaluation of these devices.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials · Spine and Intervertebral Disc Pathology

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (37)

Show source-provided references
1. ASTM F2077-11, 2011. Test Methods for Intervertebral Body Fusion Devices.
2. Development of a low-cost polymethylmethacrylate stand-alone cervical cage: technical note10.1055/s-0034-1368686 ↗
3. Anterior cervical discectomy and fusion: comparison of titanium and polyetheretherketone cages10.1186/1471-2474-13-172 ↗
4. Comparison of titanium and polyetheretherketone (PEEK) cages in the surgical treatment of multilevel cervical spondylotic myelopathy: a prospective, randomized, control study with over 7-year follow-up10.1007/s00586-013-2772-y ↗
5. Mechanical characterization of biodegradable implants10.1016/0267-6605(92)90083-6 ↗
6. U.S. Food and Drug Administration · 2007
7. Numerical modeling of bone tissue adaptation--a hierarchical approach for bone apparent density and trabecular structure10.1016/j.jbiomech.2009.01.020 ↗
8. Time-dependent failure of amorphous polylactides in static loading conditions10.1007/s10856-009-3851-9 ↗
9. Mechanical and microstructural properties of polycaprolactone scaffolds with one-dimensional, two-dimensional, and three-dimensional orthogonally oriented porous architectures produced by selective laser sintering10.1016/j.actbio.2010.02.002 ↗
10. Biomechanical behavior of a novel composite intervertebral body fusion device
11. Biomechanical comparison of cervical interbody cage versus structural bone graft10.1016/S1529-9430(03)00029-9 ↗
12. A prospective randomized multicenter clinical evaluation of an anterior cervical fusion cage10.1097/00007632-200010150-00017 ↗
13. Healthcare Cost and Utilization Project (HCUP), 2011. HCUP Databases: Agency for Healthcare Research and Quality.
14. Porosity in polyaryletheretherketone
15. In vitro biomechanical investigation of the stability and stress-shielding effect of lumbar interbody fusion devices
16. Biomechanical comparison of cervical spine interbody fusion cages10.1097/00007632-200109010-00007 ↗
17. Topology optimization of three dimensional tissue engineering scaffold architectures for prescribed bulk modulus and diffusivity10.1007/s00158-010-0508-8 ↗
18. Porous biodegradable lumbar interbody fusion cage design and fabrication using integrated global–local topology optimization with laser sintering10.1115/1.4025102 ↗
19. Biodegradable polymers in bone tissue engineering10.3390/ma2030833 ↗
20. Fabrication, mechanical and in vivo performance of polycaprolactone/tricalcium phosphate composite scaffolds10.1016/j.actbio.2012.05.018 ↗
21. Anterior cervical reconstruction using titanium cages with anterior plating10.1097/00007632-199908010-00016 ↗
22. Safety and effectiveness of bone allografts in anterior cervical discectomy and fusion surgery10.1097/BRS.0b013e3181ff37eb ↗
23. Anterior cervical discectomy and fusion with implantable titanium cage: initial impressions, patient outcomes and comparison to fusion with allograft10.1016/j.spinee.2003.05.001 ↗
24. Analysis and measurement of neck loads10.1002/jor.1100060514 ↗
25. Selective laser sintering process optimization for layered manufacturing of CAPA 6501 polycaprolactone bone tissue engineering scaffolds10.1115/1.2162589 ↗
26. Strength of the cervical spine in compression and bending10.1097/BRS.0b013e318074c40b ↗
27. Mechanical evaluation of novel spinal interbody fusion cages made of bioactive, resorbable composites10.1016/S0142-9612(03)00155-8 ↗
28. Sterilization and strength of 70/30 polylactide cages: e-beam versus ethylene oxide10.1097/01.brs.0000259057.94986.3b ↗
29. Time-dependent mechanical strength of 70/30 Poly(L, DL-lactide): shedding light on the premature failure of degradable spinal cages10.1097/BRS.0b013e31815e39df ↗
30. Bioresorbable polylactide interbody implants in an ovine anterior cervical discectomy and fusion model: three-year results10.1097/BRS.0b013e3181695716 ↗
31. Fabrication of poly(propylene fumarate)-based orthopaedic implants by photo-crosslinking through transparent silicone molds10.1016/S0142-9612(03)00364-8 ↗
32. Failure of human cervical endplates: a cadaveric experimental model10.1097/01.BRS.0000084881.11695.50 ↗
33. A prospective randomized comparison between the cloward procedure and a carbon fiber cage in the cervical spine: a clinical and radiologic study10.1097/00007632-200208150-00003 ↗
34. Spine update lumbar interbody cages10.1097/00007632-199803010-00020 ↗
35. White · 1990
36. Bone tissue engineering using polycaprolactone scaffolds fabricated via selective laser sintering10.1016/j.biomaterials.2004.11.057 ↗
37. Effect of elastic modulus on biomechanical properties of lumbar interbody fusion cage